OBJECTIVE: Our objective was to quantify the uncertainty in localization of the trigeminal nerve (TGN) with magnetic resonance imaging (MRI) and computed tomography (CT) and to determine the effect of this uncertainty on gamma-knife dose delivery.
METHODS: An MR/CT test phantom with 9, 0.6-mm diameter, copper rings was devised. The absolute ring positions in stereotactic space were determined by the angiographic module of the LGP software. The standard deviation, sigma, in the difference between the absolute and MR-measured or CT-measured coordinates of the rings was determined. The trigeminal nerve in 52 previously treated patients was contoured and expanded by 1sigma and 2sigma margins to model the uncertainty in the location of the nerve. For gamma-knife treatment, a single isocenter was used and was located at the distal cisternal portion of the trigeminal nerve root. Irradiation methods included a 4-mm collimator, 90 Gy to isocenter and a 4&8-mm collimator, 70 Gy to isocenter. A patient outcome survey that sampled pain relief and morbidity was done.
RESULTS: The MR coordinate sigma was 0.7 mm left-right, 0.8 mm anterior-posterior, and 0.6 mm superior-inferior, and the CT coordinate sigma was 0.4 mm left-right, 0.2 mm anterior-posterior, and 0.2 mm superior-inferior. A 45% higher dose line covered the TGN with the 4&8-mm method. No significant increase in pain reduction or morbidity occurred.
CONCLUSIONS: The uncertainty of target location by MRI is more than twice that found in CT imaging. The 4&8-mm collimator method covers the trigeminal root cross section with a higher isodose line than does the 4-mm method. This higher dose did not significantly reduce pain or increase morbidity.

Entities:  

Mesh:

Year:  2005        PMID: 16029818     DOI: 10.1016/j.ijrobp.2005.01.059

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  7 in total

1.  Image registration strategy of T(1)-weighted and FIESTA MRI sequences in trigeminal neuralgia gamma knife radiosurgery.

Authors:  Tony J C Wang; Ronald Brisman; Zheng Feng Lu; Xiang Li; Steven R Isaacson; Jinesh N Shah; Emi J Yoshida; Tian Liu
Journal:  Stereotact Funct Neurosurg       Date:  2010-06-08       Impact factor: 1.875

2.  Stereotactic radiosurgery of essential trigeminal neuralgia using Leksell Gamma Knife model C with automatic positioning system: technical nuances and evaluation of outcome in 130 patients with at least 2 years follow-up after treatment.

Authors:  Motohiro Hayashi; Mikhail Chernov; Noriko Tamura; Takaomi Taira; Masahiro Izawa; Shoji Yomo; Mariko Nagai; Cheng-Siu Chang; Pavel Ivanov; Manabu Tamura; Yoshihiro Muragaki; Yoshikazu Okada; Hiroshi Iseki; Kintomo Takakura
Journal:  Neurosurg Rev       Date:  2011-06-24       Impact factor: 3.042

3.  Distortion inherent to magnetic resonance imaging can lead to geometric miss in radiosurgery planning.

Authors:  Tyler M Seibert; Nathan S White; Gwe-Ya Kim; Vitali Moiseenko; Carrie R McDonald; Nikdokht Farid; Hauke Bartsch; Joshua Kuperman; Roshan Karunamuni; Deborah Marshall; Dominic Holland; Parag Sanghvi; Daniel R Simpson; Arno J Mundt; Anders M Dale; Jona A Hattangadi-Gluth
Journal:  Pract Radiat Oncol       Date:  2016-06-01

4.  Comparison of trigeminal neuralgia radiosurgery plans using two film detectors for the commissioning of small photon beams.

Authors:  Karina P Esparza-Moreno; Olivia A García-Garduño; Paola Ballesteros-Zebadúa; José M Lárraga-Gutiérrez; Sergio Moreno-Jiménez; Miguel A Celis-Lopez
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

5.  Characterizing geometrical accuracy in clinically optimised 7T and 3T magnetic resonance images for high-precision radiation treatment of brain tumours.

Authors:  Jurgen Peerlings; Inge Compter; Fiere Janssen; Christopher J Wiggins; Alida A Postma; Felix M Mottaghy; Philippe Lambin; Aswin L Hoffmann
Journal:  Phys Imaging Radiat Oncol       Date:  2019-01-03

6.  Target localization accuracy in frame-based stereotactic radiosurgery: Comparison between MR-only and MR/CT co-registration approaches.

Authors:  Eleftherios P Pappas; Ioannis Seimenis; Panagiotis Kouris; Stefanos Theocharis; Kostas I Lampropoulos; Georgios Kollias; Pantelis Karaiskos
Journal:  J Appl Clin Med Phys       Date:  2022-03-14       Impact factor: 2.243

7.  MRI-Related Geometric Distortions in Stereotactic Radiotherapy Treatment Planning: Evaluation and Dosimetric Impact.

Authors:  Eleftherios P Pappas; Mukhtar Alshanqity; Argyris Moutsatsos; Hani Lababidi; Khalid Alsafi; Konstantinos Georgiou; Pantelis Karaiskos; Evangelos Georgiou
Journal:  Technol Cancer Res Treat       Date:  2017-10-11
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.