| Literature DB >> 33458303 |
Mary Adjeiwaah1, Anders Garpebring1, Tufve Nyholm1.
Abstract
BACKGROUND ANDEntities:
Keywords: ACR; Image quality; Large geometric accuracy phantom; MRI in RT; Quality assurance; Sensitivity analysis
Year: 2020 PMID: 33458303 PMCID: PMC7807625 DOI: 10.1016/j.phro.2020.03.001
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Sequence protocol for the phantom image acquisitions.
| Parameter | ACR Phantom | Geometric Accuracy phantom | ||
|---|---|---|---|---|
| Sequence | 2D-SE (T1) | 2D-SE (T2) | 3D-FSE | 3D-GRE |
| Coil | Built-in | Built-in | ||
| Scan plane | Axial | Axial | Axial | Axial |
| Repetition time (ms) | ||||
| Echo time | ||||
| Slice thickness (mm) | ||||
| Slice gap (mm) | ||||
| Flip angle ( | ||||
| Bandwidth ( | ||||
| 3D correction | ||||
| Field of view (mm2) | ||||
UAA*: Upper anterior array.
The most frequent cause of image quality degradation based on the survey. These were the input errors introduced to intentionally degrade MR images of the large geometric accuracy and ACR phantoms.
| Simulated provocations | Expected image quality parameter to be affected | Mode of detection |
|---|---|---|
| Slice position and Geometric accuracy | Distortion measurements and ACR Slice position test | |
| 40 mm metallic paper clip in magnet’s bore | ||
| image uniformity and SNR | ACR PIU and SNR measurements | |
| No image intensity correction | ||
| Coil failure | ||
| Geometric accuracy | Distortion measurements | |
| No 2D/3D gradient | ||
| nonlinearity correction | ||
| SNR image uniformity | SNR Spike artifacts and | |
| Static electric discharges/sparks |
PIU – Percent image uniformity; SNR – signal-to-noise ratio.
Both phantoms.
ACR phantom only.
Large FoV geometric accuracy phantom only.
Fig. 2Sample phantom images with and without provocations. Phantom images with flexible coils corrected (a–c) and uncorrected (d–f) for coil intensity variations. Geometric distorted images due to a piece metal in the bore of scanner (g–i) with the measured phantom lengths. This is in comparison to the original phantom length of 190 mm. Images with spike artifacts from electric interference are in (j–l).
Results of the repeated measurements under RT settings using the ACR MRI QA phantom. Shown are the range and machine-specific action limits for a 3 T PET/MR scanner when using surface coils. The limits were set at ±3 standard deviations (σ) from the mean. The corresponding acceptance values prescribed according to the ACR MRI QA protocol are also shown.
| Image quality Index | Range | Machine-specific Limits ( | ACR Acceptance Values |
|---|---|---|---|
| Geometric Accuracy (mm) | [189–191] | 190 ± 2 | 190 ± 2 |
| Slice Thickness Accuracy (mm) | [5 | 5 | 5 |
| Slice Position Accuracy (mm) | [2 | 3 | ±5 |
| Image Intensity Uniformity (%) | [95 | ≥94 | ≥82 |
| Percent Signal Ghosting (%) | [0 | <2 | |
| Signal-to-noise ratio | |||
| T1 | [698–649 | – | |
| T2 | [693–627 | – |
–; no defined ACR acceptance value.
Fig. 1Effect of the provocations on geometric distortion measurements using the commercial geometric accuracy phantom: Scatter plot of the individual marker positions with (a) and without gradient nonlinearity correction (b). A piece of metal in the bore of the scanner produced the effects in (c) after correcting for gradient nonlinearity effects.
The sensitivity of the ACR MRI QA protocol to objectively detect individual provocations.
| Geometric | Slice | Slice | Ghosting (%) | Uniformity (%) | SNR (%) | |
|---|---|---|---|---|---|---|
| PROVOCATIONS | Accuracy (%) | Position (%) | Thickness (%) | |||
| Metal in scanner | 100 | 83.3 | 0 | 0 | 0 | 0 |
| Uncorrected coil intensity variation | 100 | 33.3 | 66.7 | 100 | 100 | 100 |
| Coil failure (after 4-faulty elements) | 0 | 0 | 0 | 0 | 66.7 | 66.7 |
| Electric sparks | 16.7 | 33.3 | 66.7 | 0 | 83.3 | 83.3 |