Literature DB >> 28516046

Image quality assessment of a 1.5T dedicated magnetic resonance-simulator for radiotherapy with a flexible radio frequency coil setting using the standard American College of Radiology magnetic resonance imaging phantom test.

Oi Lei Wong1, Jing Yuan1, Siu Ki Yu1, Kin Yin Cheung1.   

Abstract

BACKGROUND: A flexible RF coil setting has to be used on an MR-simulator (MR-sim) in the head and neck simulation scan for radiotherapy (RT) purpose, while the image quality might be compromised due to the sub-optimized flexible coil compared to the normal diagnostic radiological (DR) head coil. In this study, we assessed the image quality of an MR-sim by conducting the standard American College of Radiology (ACR) MRI phantom test on a 1.5T MR-sim under RT-setting and comparing it to DR-setting.
METHODS: A large ACR MRI phantom was carefully positioned, aligned and scanned 9 times for each under RT- and DR-setting on a 1.5T MR-sim, following the ACR scanning instruction. Images were analyzed following the ACR guidance. Measurement results under two coil settings were quantitatively compared. Inter-observer disagreements under RT-setting between two physicists were compared using Bland-Altman (BA) analysis and intra-class correlation coefficient (ICC).
RESULTS: The MR-sim with RT-setting obtained sufficiently good image quality to pass all ACR recommended criteria. No significant difference was found in phantom length accuracy, high-contrast spatial resolution, slice thickness accuracy, slice position accuracy, and percent-signal ghosting. RT-setting significantly under-performed in low-contrast object detectability, while better performed in image intensity uniformity. BA analysis showed that 95% limit of agreement and biases of phantom test measurement under RT-setting between two observers were very small. Excellent inter-observer agreement (ICC >0.75) was achieved in all measurements except for slice thickness accuracy (ICC =0.42, moderate agreement) under RT-setting.
CONCLUSIONS: Very good and highly reproducible image quality could be achieved on a 1.5T MR-sim with a flexible coil setting as revealed by the standard ACR MRI phantom test. The flexible RT-setting compromised in image signal-to-noise ratio (SNR) compared to the normal DR-setting, and resulted in reduced low-contrast object detectability.

Keywords:  ACR MRI phantom; MR-simulator (MR-sim); inter-observer agreement; radiotherapy (RT)

Year:  2017        PMID: 28516046      PMCID: PMC5418154          DOI: 10.21037/qims.2017.02.08

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  9 in total

1.  On replacing the manual measurement of ACR phantom images performed by MRI technologists with an automated measurement approach.

Authors:  Lawrence P Panych; Jr-Yuan George Chiou; Lei Qin; Vera L Kimbrell; Lisa Bussolari; Robert V Mulkern
Journal:  J Magn Reson Imaging       Date:  2015-09-23       Impact factor: 4.813

2.  Quality assurance of clinical MRI scanners using ACR MRI phantom: preliminary results.

Authors:  Chien-Chuan Chen; Yung-Liang Wan; Yau-Yau Wai; Ho-Ling Liu
Journal:  J Digit Imaging       Date:  2004-12       Impact factor: 4.056

Review 3.  Intraclass correlations: uses in assessing rater reliability.

Authors:  P E Shrout; J L Fleiss
Journal:  Psychol Bull       Date:  1979-03       Impact factor: 17.737

4.  MRI quality assurance using the ACR phantom in a multi-unit imaging center.

Authors:  Toni M Ihalainen; Nadja T Lönnroth; Juha I Peltonen; Jouni K Uusi-Simola; Marjut H Timonen; Linda J Kuusela; Sauli E Savolainen; Outi E Sipilä
Journal:  Acta Oncol       Date:  2011-08       Impact factor: 4.089

Review 5.  Magnetic resonance imaging (MRI): considerations and applications in radiotherapy treatment planning.

Authors:  V S Khoo; D P Dearnaley; D J Finnigan; A Padhani; S F Tanner; M O Leach
Journal:  Radiother Oncol       Date:  1997-01       Impact factor: 6.280

6.  Fully-automated quality assurance in multi-center studies using MRI phantom measurements.

Authors:  Mathias Davids; Frank G Zöllner; Michaela Ruttorf; Frauke Nees; Herta Flor; Gunter Schumann; Lothar R Schad
Journal:  Magn Reson Imaging       Date:  2014-02-03       Impact factor: 2.546

Review 7.  Magnetic resonance imaging acquisition techniques for radiotherapy planning.

Authors:  Gary P Liney; Marinus A Moerland
Journal:  Semin Radiat Oncol       Date:  2014-07       Impact factor: 5.934

8.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

9.  Radiotherapy planning using MRI.

Authors:  Maria A Schmidt; Geoffrey S Payne
Journal:  Phys Med Biol       Date:  2015-10-28       Impact factor: 3.609

  9 in total
  7 in total

1.  Comparison of treatment position with mask immobilization and standard diagnostic setup in intracranial MRI radiotherapy simulation.

Authors:  Vesna Mekiš; Valerija Žager Marciuš; Dominika Rogina; Laura Dolenc; Nejc Mekiš
Journal:  Strahlenther Onkol       Date:  2021-04-21       Impact factor: 3.621

2.  3D T1-weighted turbo spin echo contrast-enhanced MRI at 1.5 T for frameless brain metastases radiotherapy.

Authors:  Jing Yuan; Stephen C K Law; Ka Kin Wong; Gladys G Lo; Michael K M Kam; Wing Hong Kwan; Cindy Xue; Oi Lei Wong; Siu Ki Yu; Kin Yin Cheung
Journal:  J Cancer Res Clin Oncol       Date:  2021-08-06       Impact factor: 4.553

Review 3.  A narrative review of MRI acquisition for MR-guided-radiotherapy in prostate cancer.

Authors:  Jing Yuan; Darren M C Poon; Gladys Lo; Oi Lei Wong; Kin Yin Cheung; Siu Ki Yu
Journal:  Quant Imaging Med Surg       Date:  2022-02

4.  Selection and application of coils in temporomandibular joint MRI.

Authors:  Qi Sun; Min-Jun Dong; Xiao-Feng Tao; Meng-da Jiang; Chi Yang
Journal:  Dentomaxillofac Radiol       Date:  2019-09-27       Impact factor: 2.419

5.  Sensitivity analysis of different quality assurance methods for magnetic resonance imaging in radiotherapy.

Authors:  Mary Adjeiwaah; Anders Garpebring; Tufve Nyholm
Journal:  Phys Imaging Radiat Oncol       Date:  2020-03-13

6.  Assessment of positional reproducibility in the head and neck on a 1.5-T MR simulator for an offline MR-guided radiotherapy solution.

Authors:  Yihang Zhou; Jing Yuan; Oi Lei Wong; Winky Wing Ki Fung; Ka Fai Cheng; Kin Yin Cheung; Siu Ki Yu
Journal:  Quant Imaging Med Surg       Date:  2018-10

7.  Implementation of a dedicated 1.5 T MR scanner for radiotherapy treatment planning featuring a novel high-channel coil setup for brain imaging in treatment position.

Authors:  Veit Mengling; Christoph Bert; Rosalind Perrin; Siti Masitho; Thomas Weissmann; Sina Mansoorian; Hadi Siavooshhaghighi; Rolf Janka; Sylvain Doussin; Melanie Habatsch; Rainer Fietkau; Florian Putz
Journal:  Strahlenther Onkol       Date:  2020-10-25       Impact factor: 3.621

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.