| Literature DB >> 35493852 |
Jonathan J Wyatt1,2, Hazel M McCallum1,2, Ross J Maxwell1.
Abstract
Background and purpose Simultaneous Positron Emission Tomography - Magnetic Resonance (PET-MR) imaging can potentially improve radiotherapy by enabling more accurate tumour delineation and dose painting. The use of PET-MR imaging for radiotherapy planning requires a comprehensive Quality Assurance (QA) programme to be developed. This study aimed to develop the QA tests required and assess their repeatability and stability. Materials and methods QA tests were developed for: MR image quality, MR geometric accuracy, electromechanical accuracy, PET-MR alignment accuracy, Diffusion Weighted (DW)-MR Apparent Diffusion Coefficient (ADC) accuracy and PET Standard Uptake Value (SUV) accuracy. Each test used a dedicated phantom and was analysed automatically or semi-automatically, with in-house software. Repeatability was evaluated by three same-day measurements with independent phantom positions. Stability was assessed through 12 monthly measurements. Results The repeatability Standard Deviations (SDs) of distortion for the MR geometric accuracy test were ⩽ 0.7 mm . The repeatability SDs in ADC difference from reference were ⩽ 3 % for the DW-MR accuracy test. The PET SUV difference from reference repeatability SD was 0.3 % . The stability SDs agreed within 0.6 mm , 1 percentage point and 1.4 percentage points of the repeatability SDs for the geometric, ADC and SUV accuracy tests respectively. There were no monthly trends apparent. These results were representative of the other tests. Conclusions QA Tests for radiotherapy planning PET-MR have been developed. The tests appeared repeatable and stable over a 12-month period. The developed QA tests could form the basis of a QA programme that enables high-quality, robust PET-MR imaging for radiotherapy planning.Entities:
Keywords: Magnetic resonance imaging; PET-MR; Positron Emission Tomography; QA; Quality Assurance; Radiotherapy
Year: 2022 PMID: 35493852 PMCID: PMC9048159 DOI: 10.1016/j.phro.2022.03.003
Source DB: PubMed Journal: Phys Imaging Radiat Oncol ISSN: 2405-6316
Fig. 1Photographs of the phantoms and setup used for each of the six QA tests evaluated. Each test was carried out three times on the same day using independent phantom setups (repeatability measurements) and once a month for 12 months (stability measurements). Images were acquired using the spine coil and anterior array coil for MR image quality and DW-MR ADC accuracy tests, and the in-built body coil/PET detector for the other tests.
The repeatability and stability results for all tests. T1/T2 refers to the ACR T1/T2 image series respectively. 2D/3D refers to the 2D and 3D geometric accuracy sequences respectively. indicates distance from the scanner isocentre. EL refers to the external lasers mounted on the laser bridge and IL to the scanner internal lasers.The reference column indicates the tolerance values from the ACR manual for the MR image quality test, the recommended tolerance for geometric distortion in MR for radiotherapy and the reference values for the other tests.
| 1) MR Image Quality | Spatial Resolution (T1) [ | |||||
| Spatial Resolution (T2) [ | ||||||
| Slice Thickness (T1) [ | ||||||
| Slice Thickness (T2) [ | ||||||
| Slice Position (T1) [ | ||||||
| Slice Position (T2) [ | ||||||
| Image Uniformity (T1) [ | ||||||
| Image Uniformity (T2) [ | ||||||
| Ghosting (T1) [ | ||||||
| Ghosting (T2) [ | ||||||
| Low-Contrast Detection (T1) | 33 | 1 | ||||
| Low-Contrast Detection (T2) | 24 | 2 | ||||
| 2) MR Geometric Accuracy | SD of Distortion (2D) [ | – | – | – | ||
| SD of Distortion (3D)[ | – | – | – | |||
| Range of Distortion (2D) [ | – | |||||
| Range of Distortion (3D) [ | – | |||||
| Distortion | ||||||
| Distortion | ||||||
| Distortion | ||||||
| Distortion | ||||||
| Distortion | – | |||||
| Distortion | ||||||
| Distortion | ||||||
| Distortion | ||||||
| Distortion | ||||||
| Distortion | – | |||||
| 3) Mechanical Accuracy | EL Right-Left Offset [ | |||||
| EL Ant-Post Offset [ | ||||||
| EL Pitch Angle [o] | ||||||
| EL Roll Angle [o] | ||||||
| EL Yaw Angle [o] | ||||||
| EL Lateral Coincidence [ | ||||||
| EL Movements [ | ||||||
| EL-IL Right-Left Difference [ | ||||||
| IL Sup-Inf Offset [ | ||||||
| Couch Movements [ | ||||||
| 4) PET-MR Alignment | Right-Left Difference [ | |||||
| Ant-Post Difference [ | ||||||
| Sup-Inf Difference [ | ||||||
| Pitch Angle [o] | ||||||
| Roll Angle [o] | ||||||
| Yaw Angle [o] | ||||||
| 5) DW-MR ADC Accuracy | Nonane Difference [ | |||||
| Undecane Difference [ | ||||||
| Tridecane Difference [ | ||||||
| 6) PET SUV Accuracy | SUV Difference [ | |||||
Fig. 2Monthly stability plots of the MR image quality test. The plots show slice thickness and position (a), low contrast detection (b), image uniformity (c) and image ghosting (d) measurements.
Fig. 3Boxplots of the distributions of distortions at different distances from the isocentre for the 3D sequence for the repeatability (a) and monthly stability (b) measurements. The repeat and monthly measurements are displayed in the order acquired and shown as different colours. For the monthly measurements each colour represents two months, six months apart. The 2D sequence results showed very similar pattern.
Fig. 4Monthly stability plot of mechanical accuracy (a) and PET-MR alignment (b) measurements. Both plots show translational (left axis, solid lines) and rotational (right axis, dashed lines) differences.
Fig. 5Monthly stability plot of percentage difference in mean ADC value to temperature-corrected literature reference value for each vial (a) and plot of percentage difference in mean SUV within phantom to reference activity (b). Plot (a) also shows measured temperature of phantom (blue dashed line).