| Literature DB >> 33031418 |
Malte Lennart Warncke1, Nis Jesper Wiese1, Enver Tahir1, Susanne Sehner2, Axel Heinemann3, Marc Regier1, Klaus Püschel3, Gerhard Adam1, Julius Matthias Weinrich1, Azien Laqmani1.
Abstract
PURPOSE: Feasibility of a highly reduced-dose lumbar spine CT protocol using iterative reconstruction (IR) in a human cadaver model.Entities:
Mesh:
Year: 2020 PMID: 33031418 PMCID: PMC7544118 DOI: 10.1371/journal.pone.0240199
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
CT protocol parameters.
| SDCT | RDCT-1 | RDCT-2 | RDCT-3 | |
|---|---|---|---|---|
| Reference tube current-exposure time product (mAs) | 70 | 50 | 30 | 10 |
| Tube voltage (kV) | 140 | |||
| Collimation (mm) | 64 x 0.625 | |||
| Rotation time (s) | 0.5 | |||
| Pitch | 0.579 | |||
| Automatic tube current modulation | Automatic current selection with z-axis dose modulation (Z-DOM, Philips Healthcare) | |||
| Level of iteration | FBP and IR levels 4 and 6 | |||
| Kernel | Bone (D) | |||
| Section thickness (mm) | 3 | |||
| Reformations | Axial, coronal, sagittal | |||
SDCT = standard-dose CT; RDCT = reduced-dose CT; FBP = filtered back projection; IR = iterative reconstruction.
Fig 1a. All CT protocols (SDCT and RDCT-1, 2 and 3) reconstructed with FBP and IRL4 and IRL6 in sagittal reformations. b. As for Fig 1a, but images were obtained from a male cadaver (age at death: 53 years; BMI: 20.8 kg/m2).
Fig 2Objective image noise.
A Measurement of CT number and objective image noise (OIN). Axial CT image shows placement of ROI (circle) in the center of the body of the third lumbar vertebra. Measurements were performed in three adjacent slices and the mean values were calculated. B Quantitative analyses of OIN (defined as SD of CT number) for SDCT and three different RDCT protocols, as measured in the center of the body of the third lumbar vertebra. Images were reconstructed with FBP and IRL4 and IRL6. Horizontal midpoints of diamonds represent mean values, and error bars denote 95% CIs. Compared to SDCT with FBP, even the RDCT protocol with lowest dose (RDCT-3) reconstructed with IR resulted in lower OIN.
Radiation dose.
| Measurement | SDCT | RDCT-1 | RDCT-2 | RDCT-3 |
|---|---|---|---|---|
| Tube current-exposure time product (mAs) | 65.6 | 46.5 | 27.0 | 12.5 |
| CTDIvol (mGy) | 6.7 | 4.7 | 2.8 | 1.1 |
| DLP (mGy*cm) | 148.3 | 111.5 | 64.9 | 26.1 |
| Effective dose (mSv) | 2.3 | 1.7 | 1.2 | 0.4 |
| SSDE (mGy) | 7.9 | 5.5 | 3.5 | 1.4 |
Each table cell data is the mean (± SD). CTDIvol = volume CT dose index; DLP = dose-length product; SSDE = size-specific dose estimate.
a Values are stated as modulated tube current values.
Fig 3Mean image quality ratings of all analyzed anatomical structures for SDCT and RDCT protocols.
Number of diagnostically acceptable studies of the lumbar spine.
| SDCT | RDCT-1 | RDCT-2 | RDCT-3 | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| FBP | IRL4 | IRL6 | FBP | IRL4 | IRL6 | FBP | IRL4 | IRL6 | FBP | IRL4 | IRL6 | ||
| 20 | 20 | 20 | 19 | 20 | 20 | 13 | 19 | 20 | 8 | 13 | 18 | ||
| 0 | 0 | 0 | 1 | 0 | 0 | 7 | 1 | 0 | 12 | 7 | 2 | ||
Table cell data is number of individual studies.
FBP = filtered back projection; IR = iterative reconstruction.
a RDCT-2 with IRL6 was the lowest reduced-dose CT protocol which produced diagnostically acceptable image quality for all anatomical structures in all studies.
b Protocols and reconstructions whose image quality was not diagnostically acceptable for at least one anatomical structure (cortical and trabecular bone, intervertebral foramina, pedicles and intervertebral joints, spinous and transverse processes).
Fig 4Direct comparison in quality of axial, coronal and sagittal reformatted images produced by the SDCT reconstructed with FBP, and RDCT-2 reconstructed with IRL6.