Low-dose computed tomography of the lumbar spine: a phantom study on imaging parameters and image quality.

Muhammed Alshamari1, Mats Geijer2, Eva Norrman3, Håkan Geijer4.   

Abstract

BACKGROUND: Lumbar spine radiography has limited diagnostic value but low radiation dose compared with computed tomography (CT). The average effective radiation dose from lumbar spine radiography is about 1.1 mSv. Low-dose lumbar spine CT may be an alternative to increase the diagnostic value at low radiation dose, around 1 mSv.
PURPOSE: To determine the optimal settings for low-dose lumbar spine CT simultaneously aiming for the highest diagnostic image quality possible.
MATERIAL AND METHODS: An ovine lower thoracic and lumbar spine phantom, with all soft tissues around the vertebrae preserved except the skin, was placed in a 20 L plastic container filled with water. The phantom was scanned repeatedly with various technical settings; different tube potential, reference mAs, and with different convolution filters. Five radiologists evaluated the image quality according to a modification of the European guidelines for multislice computed tomography (MSCT) quality criteria for lumbar spine CT 2004. In a visual comparison the different scans were also ranked subjectively according to perceived image quality. Image noise and contrast were measured.
RESULTS: A tube potential of 120 kV with reference mAs 30 and medium or medium smooth convolution filter gave the best image quality at a sub-millisievert dose level, i.e. with an effective dose comparable to that from lumbar spine radiography.
CONCLUSION: Low-dose lumbar spine CT thus opens a possibility to substitute lumbar spine radiography with CT without obvious increase in radiation dose. © The Foundation Acta Radiologica 2013 Reprints and permissions: sagepub.co.uk/journalsPermissions.nav.

Entities:  

Keywords:  CT; conventional radiography; spine; structures; techniques

Mesh:

Year:  2013        PMID: 24215904     DOI: 10.1177/0284185113509615

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  6 in total

1.  EOS microdose protocol for the radiological follow-up of adolescent idiopathic scoliosis.

Authors:  Brice Ilharreborde; Emmanuelle Ferrero; Marianne Alison; Keyvan Mazda
Journal:  Eur Spine J       Date:  2015-04-24       Impact factor: 3.134

2.  Diagnostic usefulness  of low-dose lumbar multi-detector CT with iterative reconstruction in trauma patients: acomparison with standard-dose CT.

Authors:  Sun Hwa Lee; Seong Jong Yun; Dong Hyeon Kim; Hyeon Hwan Jo; Jae Gwang Song; Yong Sung Park
Journal:  Br J Radiol       Date:  2017-07-14       Impact factor: 3.039

3.  Diagnostic accuracy of low-dose versus ultra-low-dose CT for lumbar disc disease and facet joint osteoarthritis in patients with low back pain with MRI correlation.

Authors:  Sun Hwa Lee; Seong Jong Yun; Hyeon Hwan Jo; Dong Hyeon Kim; Jae Gwang Song; Yong Sung Park
Journal:  Skeletal Radiol       Date:  2017-11-06       Impact factor: 2.199

4.  Vertebrae segmentation in reduced radiation CT imaging for augmented reality applications.

Authors:  Ethan Schonfeld; Madeleine de Lotbiniere-Bassett; Tatiana Jansen; Diana Anthony; Anand Veeravagu
Journal:  Int J Comput Assist Radiol Surg       Date:  2022-01-13       Impact factor: 2.924

5.  Highly reduced-dose CT of the lumbar spine in a human cadaver model.

Authors:  Malte Lennart Warncke; Nis Jesper Wiese; Enver Tahir; Susanne Sehner; Axel Heinemann; Marc Regier; Klaus Püschel; Gerhard Adam; Julius Matthias Weinrich; Azien Laqmani
Journal:  PLoS One       Date:  2020-10-08       Impact factor: 3.240

Review 6.  The development of musculoskeletal radiology for 100 years as presented in the pages of Acta Radiologica.

Authors:  Mats Geijer; Fatih Inci; Nektarios Solidakis; Pawel Szaro; Bariq Al-Amiry
Journal:  Acta Radiol       Date:  2021-10-19       Impact factor: 1.990