Literature DB >> 29576224

MDCT in suspected lumbar spine fracture: comparison of standard and reduced dose settings using iterative reconstruction.

J M Weinrich1, L Well2, M Regier2, C Behzadi2, S Sehner3, G Adam2, A Laqmani2.   

Abstract

AIM: To compare standard (SD-) and reduced-dose computed tomography (RD-CT) in combination with iterative reconstruction (IR) in emergency patients with suspected lumbar spine fracture.
MATERIAL AND METHODS: Forty emergency patients with suspected lumbar spinal disorder who underwent RD-CT and 40 body mass index-matched patients undergoing SD-CT were enrolled in this retrospective study. Raw data for RD-CT were reconstructed using two increasing IR levels (IRL) 4 and 6, while SD-CT was reconstructed with IRL3. Two radiologists assessed image quality, image noise, and reader confidence in interpreting findings of spinal fractures in a blinded manner.
RESULTS: Effective radiation dose was reduced by 50% using RD-CT. Overall subjective image quality (SIQ) was high for both protocols and slightly superior in the RD-CT protocol for both IRL compared to SD-CT. The detection rate of spinal disorders was high for both protocols with a high interobserver agreement.
CONCLUSION: RD-CT with higher levels of IR results in substantial dose reduction of 50% in lumbar spine CT while maintaining an excellent subjective image quality resulting in a high diagnostic confidence.
Copyright © 2018. Published by Elsevier Ltd.

Entities:  

Mesh:

Year:  2018        PMID: 29576224     DOI: 10.1016/j.crad.2018.02.015

Source DB:  PubMed          Journal:  Clin Radiol        ISSN: 0009-9260            Impact factor:   2.350


  3 in total

Review 1.  Update on Imaging-Based Measurement of Bone Mineral Density and Quality.

Authors:  Thomas M Link; Galateia Kazakia
Journal:  Curr Rheumatol Rep       Date:  2020-04-09       Impact factor: 4.592

2.  Multi-detector CT imaging: impact of virtual tube current reduction and sparse sampling on detection of vertebral fractures.

Authors:  Nico Sollmann; Kai Mei; Dennis M Hedderich; Christian Maegerlein; Felix K Kopp; Maximilian T Löffler; Claus Zimmer; Ernst J Rummeny; Jan S Kirschke; Thomas Baum; Peter B Noël
Journal:  Eur Radiol       Date:  2019-03-22       Impact factor: 5.315

3.  Highly reduced-dose CT of the lumbar spine in a human cadaver model.

Authors:  Malte Lennart Warncke; Nis Jesper Wiese; Enver Tahir; Susanne Sehner; Axel Heinemann; Marc Regier; Klaus Püschel; Gerhard Adam; Julius Matthias Weinrich; Azien Laqmani
Journal:  PLoS One       Date:  2020-10-08       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.