| Literature DB >> 32651654 |
Stephan Haussig1, Axel Linke2, Norman Mangner2.
Abstract
PURPOSE OF REVIEW: Stroke remains a devastating complication of cardiovascular interventions. This review is going to discuss stroke rates and outcomes in different cardiovascular procedures with a highlight on the current evidence for the use of cerebral protection devices (CPD). RECENTEntities:
Keywords: Cardiovascular interventions; Cerebral protection; Stroke
Mesh:
Year: 2020 PMID: 32651654 PMCID: PMC7351861 DOI: 10.1007/s11886-020-01335-9
Source DB: PubMed Journal: Curr Cardiol Rep ISSN: 1523-3782 Impact factor: 2.931
Cerebral protection devices
| Sentinel® cerebral protection system (Boston Scientific) | TriGuard3™ Embolic Deflection Device (Venus Medtech, Inc.) | Point-guard™ dynamic cerebral embolic protection (Transverse Medical, Inc.) | ProtEmbo® cerebral protection system (Protembis, GmbH) | Emblok™ embolic protection system (ICS, LLC.) | Emboliner™ embolic protection catheter (Emboline, Inc.) | Captis™ Embolic Protection System (Filterlex Medical Ltd.) | |
|---|---|---|---|---|---|---|---|
| Mechanism | Capture | Deflection | Deflection | Deflection | Capture | Capture | Capture |
| Coverage | Partial | Full arch | Full arch | Full arch | Full arch | Full arch | Full arch |
| Access | 6F, right transradial | 8F, contralateral femoral | 10F, contralateral femoral | 6F, left transradial | 12F, contralateral femoral | 9F, contralateral femoral | Ipsilateral femoral |
| Mesh pore size | 140 μm | 115–145 μm | 105 μm | 60 μm | 125 μm | 150 μm | 115 μm (?) |
| Current clinical trials | PROTECTED TAVI trial (ongoing, NCT04149535) | REFLECT trial (completed, not published) | CENTER trial (on hold) | PROTEMBO SF trial (completed, not published) | European Study Evaluating the EMBLOK EPS During TAVR (recently published, Latib et al.55) | SafePass 2 trial (early clinical results presented at TCT 2019 | unknown |
| EU/USA status | CE marked/FDA approved | Investigational use | Investigational use | Investigational use | Investigational use | Investigational use | Investigational use |
F indicates French (1F~0.33 mm); partial = two vessel protection (brachiocephalic and left common carotid artery); full arch = three vessel protection (brachiocephalic, left common carotid and left subclavian artery)
Fig. 1Schematic draw of a filter-based capture (a), deflection (b), and complete protection device (c)