Abstract
BACKGROUND: To compare the intraocular scattering before and after implantation of implantable collamer lens (ICL) V4c for correction of high myopia in a short term.Entities:
Keywords: High myopia; Implantable collamer lens; Intraocular light scattering; Phakic intraocular lens
Mesh:
Year: 2020 PMID: 32552888 PMCID: PMC7301533 DOI: 10.1186/s12886-020-01482-1
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Patient demographic and characteristics
| Age | 26.11 ± 6.38 |
|---|---|
| Spherical refraction (D) | −10.62 ± 3.18 |
| Cylinder (D) | 2.17 ± 28.87 |
| AXL (mm) | 27.50 ± 1.53 |
AXL axial length.
Demographic data of patients before and after surgery
| Age (years) | 26.11 ± 6.38 (19 ~ 39) | |||
|---|---|---|---|---|
| Gender (%female) | 84 | |||
| Preoperative | Postoperative | |||
| SE (D) | −10.77 ± 3.00 | −0.86 ± 0.60 | −23.35 | < 0.05* |
| UCVA (LogMAR) | 1.62 ± 0.24 | −0.04 ± 0.08 | 40.5 | < 0.05* |
| BCVA (LogMAR) | 0.02 ± 0.05 | −0.2 ± 0.07 | 15.64 | < 0.05* |
| IOP (mmHg) | 14.53 ± 3.64 | 15.79 ± 3.86 | −1.52 | > 0.05 |
| Pupil size (mm) | 5.28 ± 1.23 | 5.96 ± 0.92 | −3.16 | < 0.05* |
SE Spherical Equivalent, D diopter, UCVA uncorrected visual acuity, BCVA best corrected visual acuity, IOP intraocular pressure P* means significant difference.
Fig. 1Postoperative predictability of ICL implantation at 1w, 1 m and 3 m postoperatively
Fig. 2Stability of ICL implantation at pre operatively, 1w, 1 m and 3 m postoperatively
The parameters of OQAS test before and after ICL implantation
| Preoperation | 1 week | 1 month | 3 months | |||
|---|---|---|---|---|---|---|
| OSI | 2.37 ± 1.6 | 1.63 ± 0.94 | 1.5 ± 0.86 | 1.43 ± 1.05 | 12.92 | |
| S/R | 0.16 ± 0.07 | 0.16 ± 0.04 | 0.16 ± 0.04 | 0.17 ± 0.05 | 0.582 | |
| MTF | 28.74 ± 12.77 | 29.33 ± 8.7 | 29.72 ± 8.49 | 31.19 ± 9.52 | 0.609 | |
| OV100% | 0.07 ± 0.22 | 0.06 ± 0.28 | 0.08 ± 0.24 | 0.13 ± 0.34 | 0.966 | |
| OV20% | 0.23 ± 0.23 | 0.21 ± 0.18 | 0.22 ± 0.16 | 0.22 ± 0.21 | 0.121 | |
| OV9% | 0.45 ± 0.25 | 0.42 ± 0.12 | 0.42 ± 0.15 | 0.39 ± 0.16 | 1.013 |
OSI objective scatter index, S/R strehl ratio, MTF Modulation Transfer Function cutoff frequency, OV OQAS value, P* means significant difference.
The rate of PDQ questionnaire
| PDQ question | 1 week | 1 month | 3 months | ||
|---|---|---|---|---|---|
| 1 bright light in general | 6.11 ± 1.94 | 3.63 ± 2.18 | 1.63 ± 1.38 | 25.941 | |
| 2 headlights at night | 5.53 ± 1.57 | 4.21 ± 1.94 | 2.11 ± 2.2 | 14.571 | |
| 3 halos around light | 5.53 ± 2.39 | 4.05 ± 2.39 | 2.21 ± 2.48 | 8.457 | |
| 4 flashes avoid light | 4.26 ± 2.55 | 2.47 ± 1.79 | 0.95 ± 1.19 | 13.370 | |
| 5 dark/grey shadow | 2.84 ± 3.22 | 1.53 ± 2.21 | 0.79 ± 1.54 | 3.317 | |
| 6 glares blind me | 2.58 ± 2.56 | 1.21 ± 1.67 | 0.58 ± 0.88 | 5.573 | |
| 7 side/above light | 3.16 ± 2.21 | 1.11 ± 1.29 | 0.89 ± 1.55 | 9.429 | |
| 8 flickering shadow | 2.84 ± 3.01 | 1.37 ± 2.23 | 0.68 ± 1.42 | 4.090 | |
| 9 semi-circular shadow | 1.79 ± 2.28 | 0.68 ± 1.03 | 0.21 ± 0.52 | 5.414 | |
The minimum score for each question was 0 (no symptoms) and the maximum score was 10 (feel the worst with symptoms) P = difference among 1w, 1 m and 3 ml; P = difference between 1w and 1 m; P = difference between 1 m and 3 m, P = difference between 1w and 3 m; P* means significantly difference.
The relationship between the PDQ scale and 4 mm OSI tested by OQAS postoperatively
| 1 week | 1 month | 3 months | |
|---|---|---|---|
| Score | 3.84 ± 2.86 | 2.25 ± 2.3 | 1.12 ± 1.7 |
| OSI | 1.63 ± 0.94 | 1.5 ± 0.85 | 1.43 ± 1.05 |
| 0.3 | 0.2 | 0.13 | |
| 0.06 | 0.19 | 0.43 |
Score = PDQ mean score.