Optical quality comparison of conventional and hole-visian implantable collamer lens at different degrees of decentering.

Cari Pérez-Vives1, Teresa Ferrer-Blasco, David Madrid-Costa, Santiago García-Lázaro, Robert Montés-Micó.   

Abstract

PURPOSE: To compare the optical quality of implantable Collamer lens (ICL) with and without central hole (Hole ICL and conventional ICL) at different degrees of decentering.
DESIGN: Experimental laboratory investigation.
METHODS: Wavefront aberrations of the -3, -6, and -12 diopter (D) V4b and -3, -6, and -12 D V4c ICLs were measured in 3 conditions-centered and decentered 0.3 and 0.6 mm-at 3-mm and 4.5-mm pupils. The root mean square of total higher order aberrations, trefoil, coma, tetrafoil, secondary astigmatism, and spherical aberration were evaluated. In addition, point spread function and simulated retinal images of ICLs were calculated from the wavefront aberrations for each ICL and all conditions of decentering at 4.5-mm pupil.
RESULTS: No statistically significant differences in any Zernike coefficient terms evaluated were found between conventional and Hole ICLs for any ICL powers and pupils evaluated (P > .05). We could not appreciate differences in the point spread function images and in simulated retinal images. Regarding the effect of the ICL decentration, coma aberration increased significantly with ICL decentration (P < .05), although these differences were not visible in the point spread function images and simulated retinal images. The ICL decentration was affected in the same manner on the conventional and Hole ICLs.
CONCLUSIONS: The outcomes showed good and comparable optical quality of the conventional and Hole ICLs for all ICL powers evaluated. Despite that coma aberration increased with ICL decentering, these values were clinically negligible and did not have a significant effect on the simulated visual performance.
Copyright © 2013 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2013        PMID: 23540712     DOI: 10.1016/j.ajo.2013.01.030

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  14 in total

Review 1.  [Scheimpflug photography for the examination of phakic intraocular lenses].

Authors:  M Baumeister
Journal:  Ophthalmologe       Date:  2014-10       Impact factor: 1.059

2.  Theoretical study on the need for laser iridotomy in an implantable collamer lens with a hole using computational fluid dynamics.

Authors:  T Kawamorita; K Shimizu; N Shoji
Journal:  Eye (Lond)       Date:  2017-01-20       Impact factor: 3.775

3.  Comparison of postoperative optical quality according to the degree of decentering of V4c implantable collamer lens.

Authors:  Min Ji Park; Hye Min Jeon; Kyoung Heon Lee; Sang Youp Han
Journal:  Int J Ophthalmol       Date:  2017-04-18       Impact factor: 1.779

4.  In vitro optical performance of a new aberration-free intraocular lens.

Authors:  D Madrid-Costa; J Ruiz-Alcocer; T Ferrer-Blasco; S García-Lázaro; R Montés-Micó
Journal:  Eye (Lond)       Date:  2014-02-21       Impact factor: 3.775

5.  Effects of tilt and decentration of Visian Implantable Collamer Lens (ICL V4c) on visual quality: an observational study.

Authors:  Lingling Niu; Zhe Zhang; Huamao Miao; Jing Zhao; Meiyan Li; Ji C He; Peijun Yao; Xingtao Zhou
Journal:  BMC Ophthalmol       Date:  2022-07-05       Impact factor: 2.086

6.  Clinical Prediction of Inadequate Vault in Eyes With Thick Lens After Implantable Collamer Lens Implantation Using Iris Morphology.

Authors:  Zhikun Yang; Lihui Meng; Xinyu Zhao; Youxin Chen; Yan Luo
Journal:  Front Med (Lausanne)       Date:  2022-06-09

7.  Implantable collamer lens V4b and V4c for correction of high myopia.

Authors:  Vipul Bhandari; Smitha Karandikar; Jagdeesh K Reddy; Kirti Relekar
Journal:  J Curr Ophthalmol       Date:  2016-02-09

8.  Posterior chamber collagen copolymer phakic intraocular lens with a central hole for moderate-to-high myopia: First experience in China.

Authors:  Xinfang Cao; Weiliang Wu; Yang Wang; Chen Xie; Jianping Tong; Ye Shen
Journal:  Medicine (Baltimore)       Date:  2016-09       Impact factor: 1.889

9.  Comparison of intraocular pressure andvault after implantation of implantable collamer lens with and without a central hole.

Authors:  Haiting Chen; Guangzeng Niu; Yuxiang Fan; Jingxue Ma
Journal:  BMC Ophthalmol       Date:  2016-11-16       Impact factor: 2.209

10.  Short-time evaluation on intraocular scattering after implantable collamer lens implantation for correcting high myopia.

Authors:  Zhe Yu; Jun Li; Hui Song
Journal:  BMC Ophthalmol       Date:  2020-06-17       Impact factor: 2.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.