Xin Liu1, Xiaohui Song1, Wei Wang1, Yanan Zhu1, Danni Lyu1, Xingchao Shentu1, Peiqing Chen1, Yibo Yu1, Ke Yao1.
Abstract
PURPOSE: To compare the clinical outcomes of echelette extended range of vision (ERV) and diffractive bifocal intraocular lenses (IOLs).Entities:
Year: 2019 PMID: 31662895 PMCID: PMC6778892 DOI: 10.1155/2019/5815040
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Participant characteristics.
| Parameter | ZXR00 | ZMB00 |
|
|---|---|---|---|
| Patients/eyes ( | 38/56 | 35/53 | |
| Implantation type | 0.840 | ||
| Monocular ( | 20 (35.7%) | 17 (32.1%) | |
| Binocular ( | 36 (64.3%) | 36 (67.9%) | |
| Age (years), mean ± SD | 68.77 ± 8.22 | 66.87 ± 6.53 | 0.186 |
| Gender | 0.825 | ||
| Male ( | 13 (23.2%) | 14 (26.4%) | |
| Female ( | 43 (76.8%) | 39 (73.6%) | |
| UCDVA (logMAR), mean ± SD | 0.58 ± 0.38 | 0.70 ± 0.45 | 0.145 |
| Keratometry (D), mean ± SD | 43.78 ± 1.51 | 43.90 ± 1.43 | 0.677 |
| Axial length (mm), mean ± SD | 23.65 ± 0.70 | 23.73 ± 0.94 | 0.627 |
| Anterior chamber depth (mm), mean ± SD | 2.81 ± 0.46 | 2.76 ± 0.45 | 0.604 |
| IOL power (D), mean ± SD | 21.12 ± 1.49 | 20.83 ± 2.13 | 0.417 |
| Angle kappa (mm), mean ± SD | 0.20 ± 0.11 | 0.24 ± 0.14 | 0.176 |
| Corneal astigmatism (D), mean ± SD | 0.58 ± 0.22 | 0.53 ± 0.27 | 0.314 |
| Corneal endothelial cell count (mm2), mean ± SD | 2532.0 ± 260.4 | 2596.21 ± 234.9 | 0.180 |
| Nuclear hardness | 0.337 | ||
| Emery Grade I ( | 20 (35.7%) | 21 (39.6%) | |
| Emery Grade II ( | 30 (53.6%) | 22 (41.5%) | |
| Emery Grade III ( | 6 (10.7%) | 10 (18.9%) | |
| Surgical technique | 0.099 | ||
| Standard ( | 22 (39.3%) | 13 (24.5%) | |
| Femtosecond-assisted ( | 34 (60.7%) | 40 (75.5%) |
n number of eyes; UCDVA = uncorrected distance visual acuity; logMAR = logarithm of the minimum angle of resolution; mm = millimeter; D = diopter; IOL = intraocular lens; SD = standard deviation.
Repeated measurements of visual acuities after IOL implantation.
| Parameter | Postoperative visit |
| ||
|---|---|---|---|---|
| 1 week | 1 month | 3 months | ||
| UCDVA (logMAR), mean ± SD | ||||
| ZXR00 (eyes = 39) | 0.10 ± 0.14 | 0.11 ± 0.15 | 0.09 ± 0.13 | 0.606 |
| ZMB00 (eyes = 28) | 0.22 ± 0.21 | 0.19 ± 0.16 | 0.19 ± 0.17 | 0.415 |
| UCIVA (logMAR), mean ± SD | ||||
| ZXR00 (eyes = 39) | 0.19 ± 0.16 | 0.18 ± 0.17 | 0.14 ± 0.12 | 0.161 |
| ZMB00 (eyes = 28) | 0.38 ± 0.24 | 0.35 ± 0.15 | 0.30 ± 0.15 | 0.112 |
| UCNVA (logMAR), mean ± SD | ||||
| ZXR00 (eyes = 39) | 0.44 ± 0.19 | 0.35 ± 0.20† | 0.34 ± 0.20† | 0.008 |
| ZMB00 (eyes = 28) | 0.32 ± 0.18 | 0.28 ± 0.19 | 0.25 ± 0.18 | 0.126 |
| SE (D), mean ± SD | ||||
| ZXR00 (eyes = 39) | −0.19 ± 0.49 | −0.21 ± 0.60 | −0.19 ± 0.64 | 0.964 |
| ZMB00 (eyes = 28) | 0.23 ± 0.69 | 0.22 ± 0.71 | 0.12 ± 0.79 | 0.088 |
† P < 0.05 compared to the visual acuity of 1 week after operation. UCDVA = uncorrected distance visual acuity; UCIVA = uncorrected intermediate visual acuity; UCNVA = uncorrected near visual acuity; SE = spherical equivalent; D = diopter; SD = standard deviation.
Visual acuities and refractive outcomes 3 months after IOL implantation.
| ZXR00 (eyes = 45) | ZMB00 (eyes = 34) |
| |
|---|---|---|---|
| UCDVA (logMAR), mean ± SD | 0.10 ± 0.13 | 0.19 ± 0.19 | 0.012 |
| UCIVA (logMAR), mean ± SD | 0.15 ± 0.13 | 0.29 ± 0.17 | <0.001 |
| UCNVA (logMAR), mean ± SD | 0.35 ± 0.19 | 0.26 ± 0.21 | 0.057 |
| CDVA (logMAR), mean ± SD | −0.01 ± 0.07 | 0.03 ± 0.08 | 0.008 |
| CIVA (logMAR), mean ± SD | 0.03 ± 0.11 | 0.08 ± 0.15 | 0.134 |
| CNVA (logMAR), mean ± SD | 0.10 ± 0.20 | 0.10 ± 0.18 | 0.719 |
| DCIVA (logMAR), mean ± SD | 0.12 ± 0.13 | 0.32 ± 0.19 | <0.001 |
| DCNVA (logMAR), mean ± SD | 0.38 ± 0.17 | 0.22 ± 0.24 | 0.001 |
| Int SE add (D), mean ± SD | 0.72 ± 0.52 | 0.94 ± 0.38 | 0.036 |
| Near SE add (D), mean ± SD | 1.84 ± 0.62 | 0.45 ± 1.14 | <0.001 |
| SE (D), mean ± SD | −0.22 ± 0.61 | 0.12 ± 0.73 | 0.025 |
UCDVA = uncorrected distance visual acuity; UCIVA = uncorrected intermediate visual acuity; UCNVA = uncorrected near visual acuity; CDVA = corrected distance visual acuity; CIVA = corrected intermediate visual acuity; CNVA = corrected near visual acuity; DCIVA = distance-corrected intermediate visual acuity; DCNVA = distance-corrected near visual acuity; SE = spherical equivalent; int SE add = addition of diopters to spectacle from the best corrected distance vision to achieve the best corrected intermediate vision; near SE add = addition of diopters to spectacle from the best corrected distance vision to achieve the best corrected near vision; D = diopter; SD = standard deviation.
Multivariate analysis on the DCIVA (logMAR) 3 months after IOL implantation.
| IOL | Correlation indicators | Variables | ||||
|---|---|---|---|---|---|---|
| Age (year) | Gender† | Keratometry (D) | Axial length (mm) | Anterior chamber depth (mm) | ||
| ZXR00 (eyes = 45) | Coefficient | −0.001 | 0.024 | 0.018 | 0.039 | −0.012 |
|
| 0.810 | 0.730 | 0.326 | 0.334 | 0.811 | |
| LCI | −0.007 | −0.115 | −0.019 | −0.042 | −0.116 | |
| UCI | 0.005 | 0.163 | 0.055 | 0.120 | 0.092 | |
| ZMB00 (eyes = 34) | Coefficient | 0.007 | 0.147 | 0.024 | 0.138 | −0.002 |
|
| 0.154 | 0.053 | 0.401 | <0.001 | 0.978 | |
| LCI | −0.003 | −0.002 | −0.034 | 0.067 | −0.153 | |
| UCI | 0.016 | 0.296 | 0.082 | 0.209 | 0.149 | |
†For gender, 0 indicates male while 1 indicates female. DCIVA = distance-corrected intermediate visual acuity; IOL = intraocular lens; D = diopter; LCI = lower bound of 95% confidence interval; UCI = upper bound of 95% confidence interval.
Multivariate analysis on the DCNVA (logMAR) 3 months after IOL implantation.
| IOL | Correlation indicators | Variables | ||||
|---|---|---|---|---|---|---|
| Age (year) | Gender† | Keratometry (D) | Axial length (mm) | Anterior chamber depth (mm) | ||
| ZXR00 (eyes = 45) | Coefficient | −0.008 | 0.015 | 0.008 | 0.003 | −0.043 |
|
| 0.018 | 0.852 | 0.705 | 0.956 | 0.485 | |
| LCI | −0.015 | −0.148 | −0.035 | −0.092 | −0.165 | |
| UCI | −0.002 | 0.178 | 0.052 | 0.098 | 0.080 | |
| ZMB00 (eyes = 34) | Coefficient | 0.009 | 0.116 | 0.009 | 0.134 | −0.144 |
|
| 0.187 | 0.286 | 0.822 | 0.013 | 0.192 | |
| LCI | −0.004 | −0.102 | −0.075 | 0.030 | −0.364 | |
| UCI | 0.022 | 0.333 | 0.094 | 0.238 | 0.077 | |
†For gender, 0 indicates male while 1 indicates female. DCNVA = distance-corrected near visual acuity; D = diopter; LCI = lower bound of 95% confidence interval; UCI = upper bound of 95% confidence interval.
Figure 1Defocus curves 3 months after IOL implantation (D = diopter; = P < 0.05; = P < 0.01; = P < 0.001).
Figure 2Contrast sensitivities with (a) and without (b) glare under mesopic condition 3 months after IOL implantation (CS = contrast sensitivity; ° = degree of angle; = P < 0.05; = P < 0.01; = P < 0.001).
Figure 3Modulation transfer function for 3.0 mm (a) and 5.0 mm (b) pupil 3 months after IOL implantation (cpd = cycle per degree; = P < 0.05; = P < 0.01; = P < 0.001).
Subjective evaluation by questionnaire 3 months after IOL implantation.
| Questionnaire | ZXR00 (eyes = 54) | ZMB00 (eyes = 44) |
|
|---|---|---|---|
| VF-14 score, mean ± SD | 90.54 ± 12.63 | 85.54 ± 13.34 | 0.021 |
| QoV score, mean ± SD | 5.06 ± 6.15 | 8.54 ± 8.35 | 0.022 |
| Frequency score | 2.20 ± 2.43 | 3.39 ± 2.99 | 0.042 |
| Severity score | 1.65 ± 2.13 | 2.64 ± 2.69 | 0.023 |
| Bothersome score | 1.20 ± 1.74 | 2.52 ± 2.82 | 0.006 |
| Self-reported vision | |||
| Day score, mean ± SD | 9.30 ± 1.24 | 8.39 ± 1.54 | 0.001 |
| Night score, mean ± SD | 8.74 ± 1.46 | 7.73 ± 1.88 | 0.004 |
| Spectacle dependence | 0.426 | ||
| Independent ( | 32 (59.3%) | 21 (47.7%) | |
| Occasionally ( | 18 (33.3%) | 17 (38.7%) | |
| Often ( | 4 (7.4%) | 6 (13.6%) | |
| Most of time ( | 0 | 0 | |
| Always ( | 0 | 0 | |
| Satisfaction grade | 0.045 | ||
| Very satisfied ( | 26 (48.1%) | 11 (25.0%) | |
| Good ( | 19 (35.2%) | 17 (38.6%) | |
| Partial improvement ( | 9 (16.7%) | 12 (27.3%) | |
| Little improvement ( | 0 | 2 (4.5%) | |
| No improvement ( | 0 | 1 (2.3%) | |
| Worse ( | 0 | 1 (2.3%) | |
| Recommendation grade | 0.045 | ||
| Strong ( | 24 (44.4%) | 11 (25.0%) | |
| Possible ( | 17 (31.5%) | 13 (29.6%) | |
| Probable ( | 13 (24.1%) | 18 (40.9%) | |
| Against ( | 0 | 2 (4.5%) | |
| No opinion ( | 0 | 0 |
n number of eyes; VF-14 = Visual Function-14; QoV = Quality of Vision; SD = standard deviation.