Literature DB >> 26469074

Effect of Large Apertures on the Optical Quality of Three Multifocal Lenses.

Jose Juan Esteve-Taboada, Alberto Domínguez-Vicent, Antonio J Del Águila-Carrasco, Teresa Ferrer-Blasco, Robert Montés-Micó.   

Abstract

PURPOSE: To compare the optical quality under large apertures among three multifocal intraocular lenses (IOLs): the TECNIS Symfony ZXR00 (Abbott Laboratories, Abbott Park, IL), the AT LISA tri 839MP (Carl Zeiss Meditec, Jena, Germany), and the Finevision (PhysIOL, Liège, Belgium).
METHODS: The in vitro optical quality of each lens was assessed with an instrument that measured the modulation transfer function (MTF). The optical quality of each lens was described in terms of MTF and through focus average MTF. The Strehl ratio, cut-off frequency, area of visibility, and percentage energy were calculated to objectively describe the optical quality of each lens. These metrics were assessed for the best lens distance focus and at four vergences (from -1.50 to -3.00 D in 0.50-D steps) at a 4.5-mm aperture.
RESULTS: The through focus average MTF of the AT LISA and Finevision IOLs showed three mean areas corresponding to distance, intermediate, and near vision. The TECNIS Symfony IOL showed two main areas corresponding to distance and intermediate vision. All metrics revealed that the Finevision IOL showed the best optical quality at distance vision, the TECNIS Symfony IOL at intermediate vision, and the AT LISA IOL at near vision. The TECNIS Symfony IOL showed the most homogeneous light distribution between its best vision foci.
CONCLUSIONS: The TECNIS Symfony IOL is less vergence dependent than the AT LISA and Finevision IOLs under dim conditions. These results may help clinicians to choose the proper lens depending on the patient's visual requirements. Copyright 2015, SLACK Incorporated.

Entities:  

Mesh:

Year:  2015        PMID: 26469074     DOI: 10.3928/1081597X-20150928-01

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  13 in total

1.  Bias of near-infrared light in evaluation of patients implanted with multifocal intraocular lenses.

Authors:  Francisco de Asís Bartol-Puyal; Galadriel Giménez; Silvia Méndez-Martínez; Irene Altemir; José Manuel Larrosa; Vicente Polo; Luis Pablo
Journal:  Int Ophthalmol       Date:  2021-05-25       Impact factor: 2.031

2.  Repeatability of in-vitro optical quality measurements of intraocular lenses with a deflectometry technique effect of the toricity.

Authors:  Teresa Ferrer-Blasco; Alberto Domínguez-Vicent; Santiago García-Lázaro; María Amparo Díez-Ajenjo; José F Alfonso; José J Esteve-Taboada
Journal:  Int J Ophthalmol       Date:  2018-07-18       Impact factor: 1.779

3.  Functional assessment of a new extended depth-of-focus intraocular lens.

Authors:  Giacomo Savini; Nicole Balducci; Claudio Carbonara; Scipione Rossi; Manuel Altieri; Nicola Frugis; Emilia Zappulla; Roberto Bellucci; Giovanni Alessio
Journal:  Eye (Lond)       Date:  2018-09-28       Impact factor: 3.775

4.  Comparative analysis of visual outcomes, reading skills, contrast sensitivity, and patient satisfaction with two models of trifocal diffractive intraocular lenses and an extended range of vision intraocular lens.

Authors:  Rita Mencucci; Eleonora Favuzza; Orsola Caporossi; Alfonso Savastano; Stanislao Rizzo
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2018-07-06       Impact factor: 3.117

5.  Difference in Quality of Vision Outcome among Extended Depth of Focus, Bifocal, and Monofocal Intraocular Lens Implantation.

Authors:  Chen-Cheng Chao; Hung-Yuan Lin; Chia-Yi Lee; Elsa Lin-Chin Mai; Ie-Bin Lian; Chao-Kai Chang
Journal:  Healthcare (Basel)       Date:  2022-05-28

6.  Extended depth of focus intraocular lens: Chromatic performance.

Authors:  Maria S Millán; Fidel Vega
Journal:  Biomed Opt Express       Date:  2017-08-31       Impact factor: 3.732

7.  Preclinical metrics to predict through-focus visual acuity for pseudophakic patients.

Authors:  Aixa Alarcon; Carmen Canovas; Robert Rosen; Henk Weeber; Linda Tsai; Kendra Hileman; Patricia Piers
Journal:  Biomed Opt Express       Date:  2016-04-15       Impact factor: 3.732

8.  Influence of the level of monovision on visual outcome with an extended range of vision intraocular lens.

Authors:  Béatrice Cochener
Journal:  Clin Ophthalmol       Date:  2018-11-12

9.  A clinical assessment of visual performance of combining the TECNIS® Symfony Extended Range of Vision IOL (ZXR00) with the +3.25 D TECNIS Multifocal 1-piece IOL (ZLB00) in subjects undergoing bilateral cataract extraction.

Authors:  Sondra Black
Journal:  Clin Ophthalmol       Date:  2018-10-23

10.  Comparison of the Clinical Outcomes between Echelette Extended Range of Vision and Diffractive Bifocal Intraocular Lenses.

Authors:  Xin Liu; Xiaohui Song; Wei Wang; Yanan Zhu; Danni Lyu; Xingchao Shentu; Peiqing Chen; Yibo Yu; Ke Yao
Journal:  J Ophthalmol       Date:  2019-09-23       Impact factor: 1.909