| Literature DB >> 30596748 |
Laura Guidetti1, Marco Meucci2, Francesco Bolletta1, Gian Pietro Emerenziani3, Maria Chiara Gallotta1, Carlo Baldari1,4.
Abstract
PURPOSE: This study aimed to examine the validity, reliability and minimum detectable change (MDC) of the Cosmed K5 in breath by breath (BxB) mode, against VacuMed metabolic simulator. Intra and inter-units reliability was also assessed.Entities:
Mesh:
Substances:
Year: 2018 PMID: 30596748 PMCID: PMC6312326 DOI: 10.1371/journal.pone.0209925
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Schematic of the COSMED K5 and VacuMed systems.
Protocol used to simulate 14 different metabolic rates (step) using VacuMed simulator.
| Simulator settings | Simulated values | |||||
|---|---|---|---|---|---|---|
| Step | Pump Stroke Volume (L) | Pump Stroke Rate (rev.min-1) | Mass Flow of Cal Gas (L.min-1) | VE (L.min-1) | VO2 (mL.min-1) | VCO2 (mL.min-1) |
| 1 | 1.5 | 15 | 4.5 | 25 | 941 | 952 |
| 2 | 1.5 | 25 | 6.0 | 42 | 1252 | 1268 |
| 3 | 1.5 | 30 | 8.0 | 49 | 1670 | 1692 |
| 4 | 1.5 | 40 | 9.0 | 65 | 1878 | 1904 |
| 5 | 2 | 20 | 7.0 | 44 | 1459 | 1478 |
| 6 | 2 | 35 | 9.5 | 77 | 1980 | 2009 |
| 7 | 2 | 40 | 11.0 | 88 | 2293 | 2325 |
| 8 | 2 | 55 | 14.0 | 120 | 2922 | 2960 |
| 9 | 2.5 | 35 | 12.0 | 97 | 2504 | 2536 |
| 10 | 2.5 | 40 | 13.0 | 111 | 2716 | 2729 |
| 11 | 2.5 | 55 | 18.0 | 152 | 3760 | 3778 |
| 12 | 3 | 40 | 15.5 | 136 | 3238 | 3253 |
| 13 | 3 | 45 | 17.5 | 149 | 3658 | 3673 |
| 14 | 3 | 50 | 19.0 | 166 | 3969 | 3988 |
“Cal Gas”; calibration gas mixture (21% CO2 and 79% N2, as detailed in VacuMed manual)
Agreement between values generated by the VacuMed simulator and values measured by the K5 as assessed by OLP regression analysis.
| R2 | Slope (95% CI) | Intercept (95% CI) | mean % diff (min to max) | ICC (95% CI) | ||||
|---|---|---|---|---|---|---|---|---|
| VE (L.min-1) | 0.9995 | 0.994 (0.980 to 1.008) | 0.063 (-1.283 to 1.391) | -0.50 (-2.79 to 0.66) | 0.11 | 1.000 (0.999 to 1.000) | 0.73 | 2.01 |
| VO2 (mL.min-1) | 0.9969 | 0.984 (0.950 to 1.018) | 36.5 (-48.6 to 118.7) | -0.04 (-2.55 to 3.52) | 0.80 | 0.999 (0.998 to 1.000) | 1.37 | 3.79 |
| VCO2 (mL.min-1) | 0.9973 | 0.983 (0.952 to 1.016) | 15.9 (-64.3 to 93.4) | -1.03 (-3.82 to 2.13) | 0.09 | 0.999 (0.997 to 1.000) | 1.34 | 3.71 |
Determinant coefficient (R2), slope and intercept of the regression equations, as well as mean percentage difference (mean % diff), p values, intra-class correlation coefficient (ICC), typical percentage error (TE), and minimum detectable change (MDC) are reported.
Fig 2Regression and difference plots of pulmonary ventilation (VE), oxygen uptake (VO2) and carbon dioxide production (VCO2) measured by K5 and generated by the VacuMed simulator.
Intra- and inter-system K5 reliability.
| Intra-K5 system reliability | Inter-K5 systems reliability | |||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| MAPE (95% CI) | ICC (95% CI) | TE | MDC | MAPE (95% CI) | ICC (95% CI) | |||||
| VE (L.min-1) | 0.66 (0.08 to 1.23) | 0.90 | 1000 (1.000 to 1.000) | 0.71 | 1.96 | 0.99 (0.41 to 1.58) | 0.73 | 1000 (0.999 to 1.000) | 0.72 | 1.99 |
| VO2 (mL.min-1) | 1.11 (0.44 to 1.78) | 0.41 | 0.999(0.997to 1.000) | 0.82 | 2.28 | 1.36 (0.91 to 1.81) | 0.65 | 0.999 (0.997 to 1.000) | 1.15 | 3.19 |
| VCO2 (mL.min-1) | 1.16 (0.86 to 2.37) | 0.98 | 0.999 (0.997to 1.000) | 0.93 | 2.57 | 1.85 (1.35 to 2.35) | 0.68 | 0.998 (0.995 to 0.999) | 1.18 | 3.28 |
Mean absolute percentage difference (MAPE), intra-class correlation coefficient (ICC), p values, typical percentage error (TE), and minimum detectable change (MDC) are reported.