| Literature DB >> 22945269 |
Abstract
Knowing the inter-unit variability, especially the technological error, is important when using many physiological measurement systems, yet no such inter-unit analysis has been undertaken on duplicate automated gas analysis systems. This study investigated the inter-unit performance of two identical ParvoMedics TrueOne 2400 automated gas analysis systems during a range of submaximal steady-state exercises performed on an electromagnetic cycle ergometer. Fifteen adult males were tested on two separate days a rest, 30, 60, 90, and 120 Watts with the duplicate gas analysis units arranged (1) collaterally (2 min of steady-state expired gas was alternately passed through each system), and (2) simultaneously (identical steady-state expired gas was passed simultaneously through both systems). Total within-subject variation (biological + technological) was determined from the collateral tests, but the unique inter-unit variability (technological error between identical systems) was shown by the simultaneous tests. Absolute percentage errors (APE), coefficient of variations (CV), effect sizes and Bland-Altman analyses were undertaken on the metabolic data, including expired ventilation (V (E)), oxygen consumption (VO(2)) and carbon dioxide production (VCO(2)). The few statistically significant differences detected between the two duplicate systems were determined to have small or trivial effect sizes, and their magnitudes to be of little physiological importance. The total within-subject variations for VO(2), VCO(2) and V (E) each equated to a mean CV and mean APE value of ~4 and ~6 %, whilst the respective inter-unit technological errors equated to ~1.5 and ~2.1 %. The two ParvoMedics TrueOne 2400 systems demonstrated excellent inter-unit agreement.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22945269 PMCID: PMC3569593 DOI: 10.1007/s00421-012-2483-9
Source DB: PubMed Journal: Eur J Appl Physiol ISSN: 1439-6319 Impact factor: 3.078
Fig. 1Illustration of the two experimental set-ups. a Collateral configuration with alternating measurements of the same steady-state activity made by the two identical gas analysis units. b Simultaneous configuration with simultaneous measurements of the same steady-state activity made by both identical gas analysis units, but sharing a common mixing chamber
Physiological responses from the graded exercise test using two “collateral” systems (1 and 2: mean ± SD), and p value from paired t tests, absolute percentage error (APE), coefficient of variation (CV), and effect size
| Variable | System 1 | System 2 | Paired | APE mean (range) | CV mean (range) | Effect size |
|---|---|---|---|---|---|---|
|
| ||||||
| Rest | 11.91 ± 6.29 | 11.84 ± 6.14 | 0.850 | 12.4 (2.0–39.7) | 8.4 (1.4–23.4) | 0.01 |
| 30 W | 19.57 ± 4.29 | 20.01 ± 4.47 | 0.110 | 4.3 (0.5–12.9) | 2.9 (0.4–8.6) | 0.11 |
| 60 W | 26.83 ± 3.40 | 27.41 ± 3.50 | 0.280 | 6.7 (0.0–18.1) | 4.6 (0.0–11.7) | 0.17 |
| 90 W | 37.37 ± 4.49 | 37.94 ± 3.43 | 0.160 | 3.1 (0.1–11.7) | 2.2 (0.1–7.8) | 0.15 |
| 120 W | 49.33 ± 7.49 | 50.71 ± 5.64 | 0.170 | 6.2 (0.4–17.3) | 4.2 (0.3–11.2) | 0.21 |
|
| ||||||
| Rest | 0.30 ± 0.06 | 0.31 ± 0.06 | 0.780 | 11.6 (1.1–26.2) | 8.1 (0.8–20.2) | 0.05 |
| 30 W | 0.73 ± 0.08 | 0.74 ± 0.08 | 0.200 | 4.7 (0.0–16.5) | 3.2 (0.0–10.8) | 0.18 |
| 60 W | 1.06 ± 0.08 | 1.07 ± 0.08 | 0.380 | 4.2 (0.7–9.4) | 2.9 (0.5–7.0) | 0.17 |
| 90 W | 1.44 ± 0.09 | 1.46 ± 0.07 | 0.340 | 3.1 (0.1–10.5) | 2.2 (0.1–7.1) | 0.20 |
| 120 W | 1.83 ± 0.14 | 1.85 ± 0.10 | 0.320 | 3.7 (0.2–8.5) | 2.6 (0.1–5.7) | 0.18 |
|
| ||||||
| Rest | 0.28 ± 0.09 | 0.28 ± 0.10 | 0.720 | 12.6 (1.0–37.4) | 8.6 (0.7–22.3) | 0.04 |
| 30 W | 0.63 ± 0.08 | 0.64 ± 0.09 | 0.160 | 4.5 (0.5–14.9) | 3.1 (0.4–9.8) | 0.17 |
| 60 W | 0.95 ± 0.08 | 0.97 ± 0.08 | 0.120 | 5.2 (0.1–13.5) | 3.5 (0.1–9.0) | 0.32 |
| 90 W | 1.36 ± 0.11 | 1.39 ± 0.09 | 0.090 | 3.3 (0.2–14.0) | 2.2 (0.2–9.2) | 0.27 |
| 120 W | 1.77 ± 0.19 | 1.82 ± 0.17 | 0.070 | 3.9 (0.1–13.4) | 2.7 (0.1–8.9) | 0.26 |
|
| ||||||
| Rest | 17.58 ± 0.60 | 17.57 ± 0.55 | 0.890 | 1.1 (0.2–2.9) | 0.8 (0.1–2.0) | 0.02 |
| 30 W | 16.42 ± 0.54 | 16.43 ± 0.58 | 0.810 | 0.6 (0.1–1.9) | 0.4 (0.1–1.3) | 0.02 |
| 60 W | 16.14 ± 0.45 | 16.18 ± 0.43 | 0.500 | 0.9 (0.0–2.8) | 0.6 (0.0–2.0) | 0.08 |
| 90 W | 16.21 ± 0.37 | 16.24 ± 0.31 | 0.600 | 0.8 (0.0–2.5) | 0.6 (0.0–1.7) | 0.08 |
| 120 W | 16.37 ± 0.37 | 16.44 ± 0.36 | 0.170 | 1.1 (0.2–2.1) | 0.8 (0.1–1.5) | 0.20 |
|
| ||||||
| Rest | 3.06 ± 0.40 | 3.06 ± 0.37 | 0.840 | 3.7 (0.1–12.6) | 2.6 (0.0–8.4) | 0.02 |
| 30 W | 4.03 ± 0.46 | 4.02 ± 0.47 | 0.860 | 2.2 (0.4–7.6) | 1.6 (0.3–5.6) | 0.01 |
| 60 W | 4.41 ± 0.42 | 4.42 ± 0.43 | 0.560 | 2.1 (0.1–5.4) | 1.5 (0.1–3.7) | 0.05 |
| 90 W | 4.55 ± 0.38 | 4.55 ± 0.33 | 0.890 | 2.6 (0.6–7.9) | 1.8 (0.4–5.4) | 0.01 |
| 120 W | 4.48 ± 0.32 | 4.46 ± 0.34 | 0.610 | 3.1 (0.4–6.6) | 2.2 (0.3–4.5) | 0.07 |
| RER | ||||||
| Rest | 0.90 ± 0.13 | 0.90 ± 0.15 | 0.930 | 4.1 (0.0–14.3) | 2.9 (0.0–10.9) | 0.01 |
| 30 W | 0.86 ± 0.04 | 0.86 ± 0.06 | 0.770 | 2.6 (0.1–7.5) | 1.8 (0.0–5.1) | 0.05 |
| 60 W | 0.89 ± 0.04 | 0.90 ± 0.04 | 0.090 | 2.5 (0.2–7.6) | 1.7 (0.1–5.2) | 0.29 |
| 90 W | 0.95 ± 0.03 | 0.95 ± 0.05 | 0.210 | 2.0 (0.0–7.3) | 1.4 (0.0–5.0) | 0.22 |
| 120 W | 0.97 ± 0.04 | 0.98 ± 0.05 | 0.070 | 2.6 (0.3–4.9) | 1.8 (0.2–3.4) | 0.30 |
* Significantly different means by paired t test with Holm’s sequential Bonferroni adjustment
Fig. 2Bland–Altman plots from the collateral (a, c, e) and simultaneous (b, d, f) tests, showing the error scores for the two gas analysis systems (Test unit 1 − Test unit 2); data shown for V E (a, b), VO2 (c, d), and VCO2 (e, f). Solid horizontal line indicates the mean error, whilst the dashed horizontal lines indicate the 95 % LOA
Physiological responses from the graded exercise test using two “simultaneous” systems (1 and 2: mean ± SD), and p value from paired t tests, absolute percentage error (APE), coefficient of variation (CV), and effect size
| Variable | System 1 | System 2 | Paired | APE mean (range) | CV mean (range) | Effect size |
|---|---|---|---|---|---|---|
|
| ||||||
| Rest | 12.26 ± 3.26 | 12.21 ± 3.30 | 0.555 | 2.6 (0.1–9.2) | 1.8 (0.0–6.2) | 0.02 |
| 30 W | 19.70 ± 2.49 | 19.63 ± 1.94 | 0.752 | 2.6 (0.1–10.8) | 1.9 (0.1–7.3) | 0.03 |
| 60 W | 27.16 ± 2.75 | 27.15 ± 2.38 | 0.991 | 2.1 (0.2–6.4) | 1.5 (0.2–4.4) | 0.00 |
| 90 W | 37.13 ± 5.08 | 37.19 ± 4.93 | 0.750 | 1.7 (0.1–4.1) | 1.2 (0.1–2.8) | 0.01 |
| 120 W | 47.86 ± 6.09 | 48.27 ± 5.73 | 0.173 | 1.8 (0.1–6.1) | 1.2 (0.1–4.2) | 0.07 |
|
| ||||||
| Rest | 0.32 ± 0.05 | 0.31 ± 0.04 | 0.102 | 2.9 (0.9–7.0) | 2.0 (0.6–4.8) | 0.10 |
| 30 W | 0.73 ± 0.08 | 0.73 ± 0.07 | 0.420 | 2.6 (0.1–9.2) | 1.8 (0.1–6.8) | 0.09 |
| 60 W | 1.05 ± 0.09 | 1.04 ± 0.08 | 0.423 | 1.9 (0.1–7.2) | 1.4 (0.1–5.3) | 0.08 |
| 90 W | 1.42 ± 0.08 | 1.41 ± 0.07 | 0.327 | 1.6 (0.1–5.2) | 1.1 (0.1–3.8) | 0.10 |
| 120 W | 1.78 ± 0.13 | 1.78 ± 0.12 | 0.823 | 1.6 (0.1–4.5) | 1.1 (0.0–3.3) | 0.02 |
|
| ||||||
| Rest | 0.29 ± 0.06 | 0.29 ± 0.06 | 0.702 | 2.4 (0.3–9.2) | 1.7 (0.2–6.2) | 0.02 |
| 30 W | 0.64 ± 0.07 | 0.64 ± 0.06 | 0.885 | 2.6 (0.3–10.7) | 1.8 (0.2–7.2) | 0.01 |
| 60 W | 0.95 ± 0.07 | 0.95 ± 0.07 | 0.856 | 2.1 (0.2–6.1) | 1.5 (0.1–4.5) | 0.02 |
| 90 W | 1.37 ± 0.11 | 1.37 ± 0.12 | 0.698 | 1.7 (0.0–3.9) | 1.2 (0.0–2.7) | 0.03 |
| 120 W | 1.75 ± 0.16 | 1.77 ± 0.15 | 0.163 | 1.7 (0.0–5.9) | 1.2 (0.0–4.0) | 0.10 |
|
| ||||||
| Rest | 17.73 ± 0.57 | 17.75 ± 0.57 | 0.008 | 0.1 (0.0–0.4) | 0.1 (0.0–0.3) | 0.03 |
| 30 W | 16.46 ± 0.41 | 16.48 ± 0.41 | 0.007 | 0.2 (0.0–0.6) | 0.1 (0.0–0.4) | 0.06 |
| 60 W | 16.27 ± 0.35 | 16.29 ± 0.34 | 0.012 | 0.2 (0.0–0.6) | 0.1 (0.0–0.4) | 0.07 |
| 90 W | 16.22 ± 0.45 | 16.25 ± 0.44 | 0.006 | 0.2 (0.0–0.6) | 0.1 (0.0–0.4) | 0.06 |
| 120 W | 16.34 ± 0.42 | 16.37 ± 0.41 | 0.003 | 0.2 (0.0–0.6) | 0.1 (0.0–0.4) | 0.07 |
|
| ||||||
| Rest | 2.99 ± 0.39 | 3.00 ± 0.39 | 0.081 | 0.3 (0.0–1) | 0.2 (0.0–0.7) | 0.01 |
| 30 W | 4.05 ± 0.37 | 4.05 ± 0.37 | 0.497 | 0.2 (0.0–0.8) | 0.2 (0.0–0.6) | 0.01 |
| 60 W | 4.37 ± 0.37 | 4.37 ± 0.37 | 0.491 | 0.3 (0.0–1.0) | 0.2 (0.0–0.7) | 0.01 |
| 90 W | 4.60 ± 0.37 | 4.60 ± 0.37 | 0.863 | 0.3 (0.0–0.7) | 0.2 (0.0–0.5) | <0.01 |
| 120 W | 4.56 ± 0.36 | 4.56 ± 0.36 | 0.792 | 0.3 (0.0–0.8) | 0.2 (0.0–0.5) | <0.01 |
| RER | ||||||
| Rest | 0.92 ± 0.12 | 0.93 ± 0.13 | 0.010 | 1.1 (0.1–4.1) | 0.8 (0.1–2.8) | 0.09 |
| 30 W | 0.87 ± 0.05 | 0.88 ± 0.05 | 0.002 | 0.8 (0.2–2.1) | 0.6 (0.1–1.5) | 0.14 |
| 60 W | 0.91 ± 0.04 | 0.92 ± 0.04 | 0.002 | 0.8 (0.1–2.0) | 0.6 (0.1–1.4) | 0.15 |
| 90 W | 0.96 ± 0.06 | 0.97 ± 0.06 | 0.003 | 0.9 (0.2–1.9) | 0.6 (0.1–1.3) | 0.12 |
| 120 W | 0.98 ± 0.04 | 0.99 ± 0.05 | 0.002* | 0.8 (0.0–1.8) | 0.6 (0.0–1.3) | 0.16 |
* Significantly different means by paired t test with Holm’s sequential Bonferroni adjustment