Po-Jen Shih1, Chun-Ju Huang2, Tzu-Han Huang2, Hung-Chou Lin3, Jia-Yush Yen2, I-Jong Wang4, Hui-Jyun Cao2, Wen-Pin Shih2, Chi-An Dai5.
Abstract
Current intraocular pressure (IOP) measurement using air puff could be erroneous without applying proper corrections. Although noncontact tonometry is not considered to be accurate, it is still popularly used by eye clinics. It is thus necessary to extract the correct information from their results. This study proposes a practical approach to correctly measure IOP in vivo. By embedding a new model-based correction to the Corvis® ST, we can extract the corneal Young's modulus from the patient data. This Young's modulus can be used to correct the IOP readings. The tests were applied to 536 right eyes of 536 healthy subjects (228 male and 308 female) between March of 2012 and April of 2016. The tests were applied to patients at the Department of Ophthalmology, National Taiwan University Hospital and the Hung-Chuo Eye Clinics. The statistical analysis showed that the value for the Young's modulus was independent of all the other parameters collected from the Corvis ST, including the corneal thickness and the intraocular pressure. Therefore, it is important to independently measure the Young's modulus instead of depending on the correlation with the other parameters. This study adds the methodology of measuring corneal stiffness in vivo for ophthalmologists' reference in diagnosis.Entities:
Year: 2017 PMID: 29250440 PMCID: PMC5698833 DOI: 10.1155/2017/5410143
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Figure 1Deflection components for fluid-filled cornea: (a) dimple deflection, (b) stretching, and (c) bending.
Figure 2Images collected from the Corvis ST and the geometric relationship marked (a) before cornea deflection and (b) at the moment of maximum deflection.
Figure 3Flowchart for corneal Young's modulus estimation.
Corvis' parameters and the proposed Young's modulus from right eyes of 536 subjects in the three different age groups.
| Age (years) | 0–14 | 15–64 | > 64 |
|---|---|---|---|
| Mean (95% CI) | Mean (95% CI) | Mean (95% CI) | |
| Number | 13 | 498 | 25 |
| IOP (mmHg) | 15.19 (8.72, 21.67) | 14.73 (7.73, 21.72) | 17.66 (9.16, 26.16) |
| A1L (mm) | 1.78 (1.63, 1.93) | 1.77 (1.54, 2.00) | 1.81 (1.70, 1.93) |
| A1V (m/s) | 0.14 (0.09, 0.18) | 0.15 (0.10, 0.19) | 0.12 (0.07, 0.17) |
| A2L (mm) | 1.73 (1.10, 2.37) | 1.72 (1.07, 2.37) | 1.81 (1.39, 2.22) |
| A2V (m/s) | −0.35 (−0.51, −0.18) | −0.37 (−0.55, −0.20) | −0.32 (−0.45, −0.19) |
| PD (mm) | 3.91 (1.23, 6.5) | 3.79 (1.22, 6.36) | 3.56 (1.02, 6.09) |
| Radius (mm) | 7.11 (5.01, 9.20) | 7.17 (4.72, 9.63) | 7.63 (5.05, 10.21) |
| DA (mm) | 1.02 (0.75, 1.29) | 1.07 (0.83, 1.31) | 1.03 (0.81, 1.25) |
| CCT ( | 537.54 (482.52, 592.56) | 543.10 (469.31, 616.90) | 561.68 (484.96, 638.40) |
| A1T (msec) | 7.45 (6.63, 8.27) | 7.38 (6.53, 8.23) | 7.76 (6.73, 8.79) |
| A2T (msec) | 21.76 (20.60, 22.93) | 21.77 (20.77, 22.77) | 21.21 (20.26, 22.16) |
|
| 0.243 (0.089, 0.397) | 0.208 (0.053, 0.363) | 0.187 (0.104, 0.269) |
A1T and A2T: applanation times; A1L and A2L: applanation diameters; A1V and A2V: applanation velocities; PD: peak distance; DA: maximum deformation amplitude; CCT: central corneal thickness; E: Young's modulus.
Pearson correlation coefficients between different parameters of the Corvis ST measurement and the proposed Young's modulus.
| IOP | A1L | A1V | A2L | A2V | Radius | DA | CCT | A1T | A2T | PD |
| |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| IOP | 1 | |||||||||||
| A1L | .219∗∗ | 1 | ||||||||||
| A1V |
| −.072 | 1 | |||||||||
| A2L | .190∗∗ | .229∗∗ | −.169∗∗ | 1 | ||||||||
| A2V |
| .153∗∗ | −.493∗∗ | .269∗∗ | 1 | |||||||
| Radius | .297∗∗ | .198∗∗ | −.239∗∗ | .194∗∗ | .366∗∗ | 1 | ||||||
| DA |
| −.162∗∗ |
| −.200∗∗ |
| −.402∗∗ | 1 | |||||
| CCT | .366∗∗ | .315∗∗ | −.262∗∗ | .280∗∗ | .425∗∗ | .353∗∗ | −.318∗∗ | 1 | ||||
| A1T |
| .216∗∗ |
| .197∗∗ |
| .299∗∗ |
| .371∗∗ | 1 | |||
| A2T |
| −.190∗∗ |
| −.097∗ |
| −.295∗∗ |
| −.307∗∗ |
| 1 | ||
| PD |
| −.098∗ | .487∗∗ | −.164∗∗ |
| −.212∗∗ |
| −.299∗∗ |
|
| 1 | |
|
| .102∗ | −.042 | −.181∗∗ | −.070 | .054 | −.250∗∗ | −.097∗ | −.165∗∗ | .108∗ | −.087∗ | −.201∗∗ | 1 |
∗∗Correlation is significant at the 0.01 level (2-tailed); ∗Correlation is significant at the 0.05 level (2-tailed). A1T and A2T: applanation times; A1L and A2L: applanation diameters; A1V and A2V: applanation velocities; PD peak distance; DA: maximum deformation amplitude; CCT: central corneal thickness; E: Young's modulus.
Association of parameters with measured IOP using a multiple regression model.
| Multiple linear regression | |||
|---|---|---|---|
| Variable | Estimate |
|
|
| Intercept | −8.106 | 2.208 | 0.000 |
| Age | .034 | .010 | 0.001 |
| CCT | .036 | .004 | 0.000 |
| Young's modulus | 8.922 | 1.894 | 0.000 |