| Literature DB >> 28772616 |
Saverio Affatato1, Filippo Zanini2, Simone Carmignato3.
Abstract
Wear is currently quantified as mass loss of the bearing materials measured using gravimetric methods. However, this method does not provide other information, such as volumetric loss or surface deviation. In this work, we validated a technique to quantify polyethylene wear in three different batches of ultrahigh-molecular-polyethylene acetabular cups used for hip implants using nondestructive microcomputed tomography. Three different configurations of polyethylene acetabular cups, previously tested under the ISO 14242 parameters, were tested on a hip simulator for an additional 2 million cycles using a modified ISO 14242 load waveform. In this context, a new approach was proposed in order to simulate, on a hip joint simulator, high-demand activities. In addition, the effects of these activities were analyzed in terms of wear and deformations of those polyethylenes by means of gravimetric method and micro X-ray computed tomography. In particular, while the gravimetric method was used for weight loss assessment, microcomputed tomography allowed for acquisition of additional quantitative information about the evolution of local wear and deformation through three-dimensional surface deviation maps for the entire cups' surface. Experimental results showed that the wear and deformation behavior of these materials change according to different mechanical simulations.Entities:
Keywords: cross-linked PE; daily activity; hip simulator; standard PE; vitamin-E doped PE; μCT
Year: 2017 PMID: 28772616 PMCID: PMC5503355 DOI: 10.3390/ma10030259
Source DB: PubMed Journal: Materials (Basel) ISSN: 1996-1944 Impact factor: 3.623
Figure 1Distribution of high-demand daily activities during wear test.
Axial load ranges for each motor task and the rotation frequency imposed during simulation.
| Activity | Min Load (N) | Max Load (N) | Frequency (Hz) |
|---|---|---|---|
| Level walking | ISO 14242-3 | ISO 14242-3 | 1.1 |
| Chair sitting and rising | 351.20 | 2776.4 | 0.7 |
| Stair climbing | 546 | 3951.3 | 1.0 |
| Stair descending | 611.7 | 4082.2 | 1.0 |
| Deep squatting | 1339.1 | 2505.8 | 0.7 |
Figure 2Load waveforms and rotation frequency for each considered motor task and machine response.
Computed tomography (CT)-scanning parameters optimized for the standard polyethylene (STD_PE) acetabular cups.
| Parameter | Value |
|---|---|
| Voltage | 194 kV |
| Current | 46 μA |
| Exposure time | 1415 ms |
| Projections | 1500 |
| Scanning time | 35 min |
| Voxel size | 31 μm |
| Physical filtering | No |
Figure 3(a) Acetabular cup as it was positioned during the CT acquisition; (b) example of 2D projection image acquired by CT (RGB coded), showing that the polystyrene stand is not visible and does not disturb the measurement.
Figure 4Unworn and nondeformed regions selected to create the geometrical elements used for the alignment (left) and 2D representation of the resulting fitting sphere, fitting plane, and intersection circle (right).
Figure 5Wear trend of all different configurations of polyethylene (PE). XLPE: cross-linked PE; XLPE_VE: cross-linked PE plus vitamin E; UHMWPE: ultrahigh molecular weight PE.
Cumulative mass loss (±standard deviation) for the three sets of PE acetabular cups tested. The p-values were obtained using a Kruskal–Wallis nonparametric test.
| Cycles (Mc) | Mean (mg) ± Standard Deviation | K–W Test ( | LSD Post Hoc Test ( | ||||
|---|---|---|---|---|---|---|---|
| XLPE | XLPE_VE | STD_PE | XLPE vs. XLPE_VE | XLPE vs. STD_PE | XLPE_VE vs. STD_PE | ||
| 2 | 6.5 ± 4.0 | 16.1 ± 8.2 | 35.4 ± 0.6 | 0.039 | NS | 0.034 | NS |
| 2.4 | 6.9 ± 5.3 | 19.3 ± 8.7 | 39.6 ± 0.3 | 0.031 | NS | 0.026 | NS |
| 2.8 | 8.6 ± 6.0 | 23.0 ± 7.9 | 43.3 ± 1.5 | 0.027 | NS | 0.022 | NS |
| 3.2 | 10.8 ± 6.0 | 29.3 ± 10.4 | 48.4 ± 1.9 | 0.027 | NS | 0.022 | NS |
| 3.6 | 12.9 ± 6.5 | 36.5 ± 12.0 | 56.3 ± 2.9 | 0.027 | NS | 0.022 | NS |
| 4 | 15.0 ± 7.1 | 43.0 ± 13.3 | 63.4 ± 3.6 | 0.027 | NS | 0.022 | NS |
Not significant (NS) = p value 0.05; K–W = Kruskal–Wallis; Mc = million cycles.
Figure 6Comparison of deviation maps resulting from wear test under ISO 14242 and modified wear test under daily activities (one acetabular cup for each material). Histograms of the calculated deviations versus the entire analyzed surface of the selected object are shown.