In vitro quantification of wear in tibial inserts using microcomputed tomography.

Matthew G Teeter1, Douglas D R Naudie, David D McErlain, Jan-M Brandt, Xunhua Yuan, Steven J Macdonald, David W Holdsworth.   

Abstract

BACKGROUND: Wear of polyethylene tibial inserts can decrease the longevity of total knee arthroplasty. Wear is currently assessed using laboratory methods that may not permit backside wear measurements or do not quantify surface deviation. QUESTIONS/PURPOSES: We developed and validated a technique to quantify polyethylene wear in tibial inserts using microcomputed tomography (micro-CT), a nondestructive high-resolution imaging technique that provides detailed images of surface geometry in addition to volumetric measurements.
METHODS: Six unworn and six wear-simulated polyethylene tibial inserts were evaluated. Each insert was scanned three times using micro-CT at a resolution of 50 μm. The insert surface was reconstructed for each scan and the insert volume was calculated. Gravimetric analysis was performed for all inserts, and the micro-CT and gravimetric volumes were compared to determine accuracy. We created three-dimensional surface deviation maps.
RESULTS: Micro-CT generated high-quality three-dimensional renderings of the insert surface geometry. Between-scan precision was 0.07%; we observed no difference between micro-CT and gravimetric volume measurements.
CONCLUSIONS: Micro-CT can provide precise and accurate volumetric measurements in addition to quantifiable three-dimensional surface deviation maps for the entire insert surface. The technique has the potential to evaluate wear in wear simulator trials and retrieval studies. CLINICAL RELEVANCE: This micro-CT technique combines the benefits of volumetric and surface scanning methods to quantify wear across all surfaces of polyethylene components with a single tool. When applied in wear simulator and retrieval studies, these measurements can be used to evaluate and predict the wear properties of the components.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 20676810      PMCID: PMC3008888          DOI: 10.1007/s11999-010-1490-6

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  16 in total

1.  Micro-wear patterns on UHMWPE tibial inserts in total knee joint simulation.

Authors:  Jiro Tamura; Ian C Clarke; Keiichi Kawanabe; Masao Akagi; Victoria D Good; Paul A Williams; Toshinori Masaoka; Dave Schroeder; Hironobu Oonishi
Journal:  J Biomed Mater Res       Date:  2002-08

2.  Assessment of backside wear from the analysis of 55 retrieved tibial inserts.

Authors:  S Li; G Scuderi; B D Furman; S Bhattacharyya; J J Schmieg; J N Insall
Journal:  Clin Orthop Relat Res       Date:  2002-11       Impact factor: 4.176

3.  Metrology to quantify wear and creep of polyethylene tibial knee inserts.

Authors:  Orhun K Muratoglu; Rebecca S Perinchief; Charles R Bragdon; Daniel O O'Connor; Reto Konrad; William H Harris
Journal:  Clin Orthop Relat Res       Date:  2003-05       Impact factor: 4.176

Review 4.  Micro-computed tomography-current status and developments.

Authors:  Erik L Ritman
Journal:  Annu Rev Biomed Eng       Date:  2004       Impact factor: 9.590

5.  Factors affecting the severity of backside wear of modular tibial inserts.

Authors:  Michael A Conditt; Jason A Stein; Philip C Noble
Journal:  J Bone Joint Surg Am       Date:  2004-02       Impact factor: 5.284

6.  Computational wear prediction of a total knee replacement from in vivo kinematics.

Authors:  Benjamin J Fregly; W Gregory Sawyer; Melinda K Harman; Scott A Banks
Journal:  J Biomech       Date:  2005-02       Impact factor: 2.712

7.  Validation of a micro-CT technique for measuring volumetric wear in retrieved acetabular liners.

Authors:  A E Bowden; S M Kurtz; A A Edidin
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2005-10       Impact factor: 3.368

8.  Quantification of polyethylene degradation in mobile bearing knees: a retrieval analysis of the Anterior-Posterior-Glide (APG) and Rotating Platform (RP) Low Contact Stress (LCS) knee.

Authors:  Alan M Kop; Eric Swarts
Journal:  Acta Orthop       Date:  2007-06       Impact factor: 3.717

9.  Wear characteristics of UHMW polyethylene: a method for accurately measuring extremely low wear rates.

Authors:  H McKellop; I C Clarke; K L Markolf; H C Amstutz
Journal:  J Biomed Mater Res       Date:  1978-11

10.  Simulated normal gait wear testing of a highly cross-linked polyethylene tibial insert.

Authors:  Orhun K Muratoglu; Harry E Rubash; Charles R Bragdon; Brian R Burroughs; Anna Huang; William H Harris
Journal:  J Arthroplasty       Date:  2007-04       Impact factor: 4.757

View more
  12 in total

1.  The 2012 Mark Coventry award: a retrieval analysis of high flexion versus posterior-stabilized tibial inserts.

Authors:  Nicholas R Paterson; Matthew G Teeter; Steven J MacDonald; Richard W McCalden; Douglas D R Naudie
Journal:  Clin Orthop Relat Res       Date:  2013-01       Impact factor: 4.176

2.  Rotating-platform has no surface damage advantage over fixed-bearing TKA.

Authors:  Kirsten Stoner; Seth A Jerabek; Stephanie Tow; Timothy M Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2013-01       Impact factor: 4.176

3.  Can microcomputed tomography measure retrieved polyethylene wear? Comparing fixed-bearing and rotating-platform knees.

Authors:  Charles A Engh; Rebecca L Zimmerman; Robert H Hopper; Gerard A Engh
Journal:  Clin Orthop Relat Res       Date:  2013-01       Impact factor: 4.176

4.  Relationship of surface damage appearance and volumetric wear in retrieved TKR polyethylene liners.

Authors:  Christopher B Knowlton; Priyanka Bhutani; Markus A Wimmer
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2016-07-12       Impact factor: 3.368

5.  Radiostereometric Analysis Permits In Vivo Measurement of Very Small Levels of Wear in TKA.

Authors:  Matthew G Teeter; Jacob Wihlidal; Richard W McCalden; Xunhua Yuan; Steven J MacDonald; Brent A Lanting; Douglas D Naudie
Journal:  Clin Orthop Relat Res       Date:  2019-01       Impact factor: 4.176

6.  Current Total Knee Designs: Does Baseplate Roughness or Locking Mechanism Design Affect Polyethylene Backside Wear?

Authors:  Zachary W Sisko; Matthew G Teeter; Brent A Lanting; James L Howard; Richard W McCalden; Douglas D Naudie; Steven J MacDonald; Edward M Vasarhelyi
Journal:  Clin Orthop Relat Res       Date:  2017-09-13       Impact factor: 4.176

7.  An autonomous mathematical reconstruction to effectively measure volume loss on retrieved polyethylene tibial inserts.

Authors:  Christopher B Knowlton; Markus A Wimmer
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2012-08-22       Impact factor: 3.368

8.  Inter and intra-system size variability of reverse shoulder arthroplasty polyethylene inserts.

Authors:  Matthew G Teeter; Matthew T Dawson; George S Athwal
Journal:  Int J Shoulder Surg       Date:  2016 Jan-Mar

9.  Micro X-Ray Computed Tomography Mass Loss Assessment of Different UHMWPE: A Hip Joint Simulator Study on Standard vs. Cross-Linked Polyethylene.

Authors:  Saverio Affatato; Filippo Zanini; Simone Carmignato
Journal:  PLoS One       Date:  2017-01-20       Impact factor: 3.240

10.  Peripheral snap-fit locking mechanisms and smooth surface finish of tibial trays reduce backside wear in fixed-bearing total knee arthroplasty.

Authors:  Łukasz Łapaj; Adrian Mróz; Paweł Kokoszka; Jacek Markuszewski; Justyna Wendland; Celina Helak-Łapaj; Jacek Kruczyński
Journal:  Acta Orthop       Date:  2016-10-26       Impact factor: 3.717

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.