In vitro wear performance of standard, crosslinked, and vitamin-E-blended UHMWPE.

Saverio Affatato1, Pierangiola Bracco, Luigi Costa, Tomaso Villa, Virginio Quaglini, Aldo Toni.   

Abstract

Crosslinked vitamin-E-stabilized polyethylene acetabular cups were compared with both commercially available conventional and custom-crosslinked polyethylene acetabular cups in terms of wear behavior, in a hip joint simulator for five millions cycles, using bovine calf serum as lubricant. We correlated the wear experiments results with the chemical characterization of the investigated materials: Fourier transformed infrared (FTIR) spectroscopic analyses, differential scanning calorimetry, and crosslink density measurements were used to assess the chemical characteristics of the pristine materials. In addition, further FTIR analyses and cyclohexane extraction were carried out after the simulator experiments. Lipids absorption was observed in all tested specimens and it has been shown to strongly affect the results of the wear test. Corrected gravimetric wear measurements showed that vitamin-E blended, crosslinked polyethylene wore more than the traditional crosslinked polyethylene but exhibited a much lower wear than conventional ultrahigh-molecular weight polyethylene. The chemical analyses showed that the addition of vitamin E reduced the crosslinking efficiency. Given the correlation between crosslink density and wear resistance, this gave an explanation for the observed wear performances.
Copyright © 2011 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2011        PMID: 22162269     DOI: 10.1002/jbm.a.33297

Source DB:  PubMed          Journal:  J Biomed Mater Res A        ISSN: 1549-3296            Impact factor:   4.396


  16 in total

1.  Retrieval analysis of contemporary antioxidant polyethylene: multiple material and design changes may decrease implant performance.

Authors:  Arianna Cerquiglini; Johann Henckel; Harry Hothi; Lukas B Moser; Antti Eskelinen; Michael T Hirschmann; Alister J Hart
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2019-02-10       Impact factor: 4.342

2.  Wear testing and particle characterisation of sequentially crosslinked polyethylene acetabular liners using different femoral head sizes.

Authors:  Carmen Zietz; Christian Fabry; Lars Middelborg; Gerhard Fulda; Wolfram Mittelmeier; Rainer Bader
Journal:  J Mater Sci Mater Med       Date:  2013-04-25       Impact factor: 3.896

3.  Reasons for Revision, Oxidation, and Damage Mechanisms of Retrieved Vitamin E-Stabilized Highly Crosslinked Polyethylene in Total Knee Arthroplasty.

Authors:  Hannah Spece; Jaclyn T Schachtner; Daniel W MacDonald; Gregg R Klein; Michael A Mont; Gwo-Chin Lee; Steven M Kurtz
Journal:  J Arthroplasty       Date:  2019-07-18       Impact factor: 4.757

4.  In vitro effects on mobile polyethylene insert under highly demanding daily activities: stair climbing.

Authors:  Sami Abdel Jaber; Paola Taddei; Silvia Tozzi; Alessandra Sudanese; Saverio Affatato
Journal:  Int Orthop       Date:  2014-12-14       Impact factor: 3.075

Review 5.  Highly cross-linked polyethylene may not have an advantage in total knee arthroplasty.

Authors:  Vasileios I Sakellariou; Peter Sculco; Lazaros Poultsides; Timothy Wright; Thomas P Sculco
Journal:  HSS J       Date:  2013-08-10

6.  Wear performance of inverted non-conforming bearings in anatomic total shoulder arthroplasty.

Authors:  Lorenzo Banci; Alessio Meoli; Martin Hintner; Hans Rudolf Bloch
Journal:  Shoulder Elbow       Date:  2019-02-07

Review 7.  Effectiveness of Vitamin-E-Doped Polyethylene in Joint Replacement: A Literature Review.

Authors:  Antonio Gigante; Carlo Bottegoni; Vincenza Ragone; Lorenzo Banci
Journal:  J Funct Biomater       Date:  2015-09-08

8.  Micro X-Ray Computed Tomography Mass Loss Assessment of Different UHMWPE: A Hip Joint Simulator Study on Standard vs. Cross-Linked Polyethylene.

Authors:  Saverio Affatato; Filippo Zanini; Simone Carmignato
Journal:  PLoS One       Date:  2017-01-20       Impact factor: 3.240

9.  Quantification of Wear and Deformation in Different Configurations of Polyethylene Acetabular Cups Using Micro X-ray Computed Tomography.

Authors:  Saverio Affatato; Filippo Zanini; Simone Carmignato
Journal:  Materials (Basel)       Date:  2017-03-03       Impact factor: 3.623

10.  A multicenter approach evaluating the impact of vitamin e-blended polyethylene in cementless total hip replacement.

Authors:  Marcus Jäger; Andrea van Wasen; Sebastian Warwas; Stefan Landgraeber; Marcel Haversath; Vitas Group
Journal:  Orthop Rev (Pavia)       Date:  2014-04-22
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.