Onur Dincer Kose1, Burcin Karataslı2, Sabit Demircan1, Taha Emre Kose3, Erhan Cene4, Serhan Aydın Aya5, Mehmet Ali Erdem6, Abdulkadir Burak Cankaya6.
Abstract
Preload is applied to screws manually or using a torque wrench in dental implant systems, and the preload applied must be appropriate for the purpose. The aim of this study was to assess screw loosening and bending/torsional moments applied by clinicians of various specialties following application of manual tightening torque to combinations of implants and abutments. Ten-millimeter implants of 3.7 and 4.1 mm diameters and standard or solid abutments were used. Each group contained five implant-abutment combinations. The control and experimental groups comprised 20 and 160 specimens, respectively. Implants in the experimental group were tightened by dentists of different specialties. Torsional and bending moments during tightening were measured using a strain gauge. Control group and implants with preload values close to the ideal preload were subjected to a dynamic loading test at 150 N, 15 Hz, and 85,000 cycles. The implants that deformed in this test were examined using an optical microscope to assess deformities. Manual tightening did not yield the manufacturer-recommended preload values. Dynamic loading testing suggested early screw loosening/fracture in samples with insufficient preload.Entities:
Mesh:
Substances:
Year: 2017 PMID: 28473988 PMCID: PMC5394346 DOI: 10.1155/2017/7376261
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1Experimental set-up.
Figure 2(A) Implant placement in the experimental set-up. (B) Handheld screwdriver placement in the experimental set-up.
Figure 3Measurement of loosening torque using a torque tester.
Mean bending and torsional moment values according to specialty.
| Moment type | Specialty | Mean (Ncm) |
|---|---|---|
| Bending moment (Ch1) | Oral and maxillofacial surgeon | 22.868 |
| Prosthodontist | 29.420 | |
| Periodontist | 23.075 | |
| Dentist | 17.546 | |
|
| ||
| Bending moment (Ch1) | Oral and maxillofacial surgeon | 22.917 |
| Prosthodontist | 20.736 | |
| Periodontist | 14.392 | |
| Dentist | 12.476 | |
|
| ||
| Torsional moment (Ch3) | Control | 25.988 |
| Oral and maxillofacial surgeon | 9.773 | |
| Prosthodontist | 10.989 | |
| Periodontist | 8.941 | |
| Dentist | 7.283 | |
ANOVA for bending and torsional moments according to specialty.
| Moment type |
| Sig. |
|---|---|---|
| Bending moment (Ch1) | 2.230 | 0.067 |
| Bending moment (Ch2) | 4.389 | 0.002 |
| Torsional moment (Ch3) | 134.490 | <0.001 |
∗ refers to <0.05.
Bonferroni test for bending and torsional moments according to specialty.
| Moment | ( | ( | Mean difference | Std. error | Sig. |
|---|---|---|---|---|---|
| Bending moment (Ch1) | Oral and maxillofacial surgeon | Prosthodontist | −6.552 | 4.245 | 0.748 |
| Periodontist | −0.207 | 4.245 | 1.000 | ||
| Dentist | 5.322 | 4.245 | 1.000 | ||
| Prosthodontist | Oral and maxillofacial surgeon | 6.552 | 4.245 | 0.748 | |
| Periodontist | 6.345 | 4.245 | 0.822 | ||
| Dentist | 11.874 | 4.245 | 0.035 | ||
| Periodontist | Oral and maxillofacial surgeon | 0.207 | 4.245 | 1.000 | |
| Prosthodontist | −6.345 | 4.245 | 0.822 | ||
| Dentist | 5.529 | 4.245 | 1.000 | ||
| Dentist | Oral and maxillofacial surgeon | −5.322 | 4.245 | 1.000 | |
| Prosthodontist | −11.874 | 4.245 | 0.035 | ||
| Periodontist | −5.529 | 4.245 | 1.000 | ||
|
| |||||
| Bending moment (Ch2) | Oral and maxillofacial surgeon | Prosthodontist | 2.181 | 3.585 | 1.000 |
| Periodontist | 8.525 | 3.585 | 0.112 | ||
| Dentist | 10.440 | 3.585 | 0.025 | ||
| Prosthodontist | Oral and maxillofacial surgeon | −2.181 | 3.585 | 1.000 | |
| Periodontist | 6.344 | 3.585 | 0.473 | ||
| Dentist | 8.259 | 3.585 | 0.135 | ||
| Periodontist | Oral and maxillofacial surgeon | −8.525 | 3.585 | 0.112 | |
| Prosthodontist | −6.344 | 3.585 | 0.473 | ||
| Dentist | 1.915 | 3.585 | 1.000 | ||
| Dentist | Oral and maxillofacial surgeon | −10.440 | 3.585 | 0.025 | |
| Prosthodontist | −8.259 | 3.585 | 0.135 | ||
| Periodontist | −1.915 | 3.585 | 1.000 | ||
|
| |||||
| Torsional moment (Ch3) | Control | Oral and maxillofacial surgeon | 16.215 | 0.928 | <0.001 |
| Prosthodontist | 14.999 | 0.928 | <0.001 | ||
| Periodontist | 17.047 | 0.928 | <0.001 | ||
| Dentist | 18.704 | 0.928 | <0.001 | ||
| Oral and maxillofacial surgeon | Control | −16.215 | 0.928 | <0.001 | |
| Prosthodontist | −1.216 | 0.928 | 1.000 | ||
| Periodontist | 0.832 | 0.928 | 1.000 | ||
| Dentist | 2.490 | 0.928 | 0.079 | ||
| Prosthodontist | Control | −14.999 | 0.928 | <0.001 | |
| Oral and maxillofacial surgeon | 1.216 | 0.928 | 1.000 | ||
| Periodontist | 2.048 | 0.928 | 0.285 | ||
| Dentist | 3.705 | 0.928 | 0.001 | ||
| Periodontist | Control | −17.047 | 0.928 | <0.001 | |
| Oral and maxillofacial surgeon | −0.832 | 0.928 | 1.000 | ||
| Prosthodontist | −2.048 | 0.928 | 0.285 | ||
| Dentist | 1.658 | 0.928 | 0.755 | ||
| Dentist | Control | −18.704 | 0.928 | <0.001 | |
| Oral and maxillofacial surgeon | −2.490 | 0.928 | 0.079 | ||
| Prosthodontist | −3.705 | 0.928 | <0.001 | ||
| Periodontist | −1.658 | 0.928 | 0.755 | ||
∗ refers to <0.05.
Mean bending and torsional moment values according to gender.
| Moment type | Gender | Mean (Ncm) |
|---|---|---|
| Bending moment (Ch1) | Female | 19.718 |
| Male | 26.737 | |
|
| ||
| Bending moment (Ch2) | Female | 15.693 |
| Male | 19.567 | |
|
| ||
| Torsional moment (Ch3) | Female | 8.199 |
| Male | 10.295 | |
| Control | 25.988 | |
t-test for equality of means according to gender and bending moment.
| Moment type |
| Two-tailed ( |
|---|---|---|
| Bending moment (Ch1) | −2.336 | 0.021 |
| Bending moment (Ch2) | −1.494 | 0.137 |
∗ refers to <0.05.
ANOVA for Ch3 torsional moment according to gender.
| Moment type |
| Sig. |
|---|---|---|
| Torsional moment (Ch3) | 259.502 | <0.001 |
∗ refers to <0.05.
Bonferroni test for Ch3 torsional moment according to gender.
| Moment type | ( | ( | Mean difference | Std. error | Sig. |
|---|---|---|---|---|---|
| Torsional moment (Ch3) | Female | Male | −2.096 | 0.664 | 0.006 |
| Control | −17.789 | 0.813 | <0.001 | ||
| Male | Female | 2.096 | 0.664 | 0.006 | |
| Control | −15.693 | 0.813 | <0.001 | ||
| Control | Female | 17.789 | 0.664 | <0.001 | |
| Male | 15.693 | 0.813 | <0.001 |
∗ refers to <0.05.
Evaluation of the correlation between initial tightening torque and gap/fraction on the loading side using Spearman's rho test.
| Correlation coefficient | Starting | Case |
|---|---|---|
| Spearman's Rho | ||
| Starting | ||
| Correlation coefficient | 1.000 | −0.001 |
| Two-tailed ( | — | 0.996 |
| | 20 | 20 |
| Case | ||
| Correlation coefficient | −0.001 | 1.000 |
| Two-tailed ( | 0.996 | — |
| | 20 | 20 |
Complications of implants undergoing dynamic loading testing.
| Implant | Experimental group | Control group | |||
|---|---|---|---|---|---|
| Fractures | Opening | Loosening | Fractures | Loosening | |
| 3.7 mm + standard | 3 | 2 | 0 | 1 | 4 |
| 3.7 mm + solid | 4 | 1 | 0 | 1 | 4 |
| 4.1 mm + standard | 2 | 1 | 2 | 1 | 4 |
| 4.1 mm + solid | 0 | 4 | 1 | 0 | 5 |
| Total | 9 | 8 | 3 | 3 | 17 |