Mehmet Ali Erdem1, Burcin Karatasli2, Onur Dinçer Kose3, Taha Emre Kose4, Erhan Çene5, Serhan Aydın Aya6, Abdulkadir Burak Cankaya1.
Abstract
PURPOSE: Friction-style and spring-style torque wrenches are used to tighten implant abutments and prosthetic screws. The mechanical stability of these torque wrenches is crucial for the implant-abutment connection. The purposes of this study were to assess the performance of five brands (Straumann, Zimmer, Implant KA, Bredent, and Biohorizons) of wrench and to evaluate possible changes in applied torque values of aged wrenches.Entities:
Mesh:
Substances:
Year: 2017 PMID: 29238725 PMCID: PMC5697386 DOI: 10.1155/2017/8652720
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Figure 1The P3 Strain Indicator and Recorder (a), and positioning of the implant in the device (b).
Minimum, maximum, and mean values for all new and aged wrenches.
| Implant | Minimum | Maximum | Mean | Std. deviation |
|
|---|---|---|---|---|---|
| Biohorizons (aged) | 17.570 | 32.024 | 23.703 | 2.770 |
|
| Biohorizons (new) | 19.554 | 27.489 | 23.544 | 1.888 | |
|
| |||||
| Bredent (aged) | 17.287 | 45.060 | 30.39 | 1.5069 |
|
| Bredent (new) | 21.538 | 38.825 | 29.643 | 0.9666 | |
|
| |||||
| Implant KA (aged) | 18.9878 | 27.2064 | 23.1594 | 2.3690 |
|
| Implant KA (new) | 18.9878 | 25.2226 | 22.3886 | 1.5173 | |
|
| |||||
| Straumann (aged) | 19.27 | 33.16 | 29.8250 | 2.9557 |
|
| Straumann (new) | 18.70 | 37.98 | 30.1311 | 3.4132 | |
|
| |||||
| Zimmer (aged) | 17.5708 | 31.1740 | 26.2201 | 3.2001 |
|
| Zimmer (new) | 15.0202 | 37.4088 | 27.8412 | 3.8774 | |
Accuracies of new and aged wrenches.
| Implant | Target |
| 95% confidence interval of the difference | Mean |
| |
|---|---|---|---|---|---|---|
| Lower | Upper | |||||
| Biohorizons (used) | 30 | −11.363 | −7.4400 | −5.1527 | −6.29642 |
|
| Biohorizons (new) | 30 | −17.090 | −7.2347 | −5.6755 | −6.4551 |
|
|
| ||||||
| Bredent (used) | 30 | 0.358 | −2.5709 | 3.6493 | 0.5391 |
|
| Bredent (new) | 30 | −0.369 | −2.3513 | 1.6386 | −0.3563 |
|
|
| ||||||
| Implant KA (used) | 25 | −3.885 | −2.8184 | −0.8626 | −1.8405 |
|
| Implant KA (new) | 25 | −8.605 | −3.2377 | −1.985 | −2.6114 |
|
|
| ||||||
| Straumann (used) | 35 | −8.754 | −6.3950 | −3.9549 | −5.1749 |
|
| Straumann (new) | 35 | −7.132 | −6.2778 | −3.4599 | −4.8689 |
|
|
| ||||||
| Zimmer (used) | 30 | −5.906 | −5.1007 | −2.4588 | −3.7798 |
|
| Zimmer (new) | 30 | −2.784 | −3.7592 | −0.5582 | −2.1587 |
|
Differences between mean values and target values in N·cm and as percentages.
| Biohorizons | Bredent | Implant KA | Straumann | Zimmer | |
|---|---|---|---|---|---|
| Target value (N·cm) | 30 | 30 | 25 | 35 | 30 |
|
| |||||
| All wrenches | |||||
| Mean value | 23.624 | 30.091 | 22.774 | 29.978 | 27.031 |
| Difference | −6.376 | 0.091 | −2.226 | −5.022 | −2.969 |
| Difference% | −21.253 | 0.305 | −8.904 | −14.348 | −9.898 |
|
| |||||
| Aged wrenches | |||||
| Mean value | 23.704 | 30.539 | 23.159 | 29.825 | 26.220 |
| Difference | −6.296 | 0.539 | −1.841 | −5.175 | −3.780 |
| Difference% | −20.988 | 1.797 | −7.362 | −14.786 | −12.599 |
|
| |||||
| New wrenches | |||||
| Mean value | 23.545 | 29.644 | 22.389 | 30.131 | 27.841 |
| Difference | −6.455 | −0.356 | −2.611 | −4.869 | −2.159 |
| Difference% | −21.517 | −1.188 | −10.446 | −13.911 | −7.196 |
Figure 2Torques delivered by aged and new Bredent ratchet wrenches.
Figure 3Torques delivered by aged and new Biohorizons ratchet wrenches.
Figure 4Torques delivered by aged and new Straumann ratchet wrenches.
Figure 5Torques delivered by aged and new Implant KA ratchet wrenches.
Figure 6Torques delivered by aged and new Zimmer ratchet wrenches.