Torque generated by handheld screwdrivers and mechanical torquing devices for osseointegrated implants.

K L Goheen1, S G Vermilyea, J Vossoughi, J R Agar.   

Abstract

This study evaluated the ability of practitioners experienced in the use of Brãnemark implant components to impart a desired torque using handheld screwdrivers. The torque output and variability of mechanical torque devices was also examined. For clinicians using handheld screwdrivers, values ranged from 0.7 to 18.1 Ncm, 1.4 to 33.7 Ncm, and 8.2 to 36.2 Ncm for the 10-Ncm, 20-Ncm, and 32-Ncm torques respectively. The manually operated mechanical torquing devices produced torque values within the tolerances specified by their respective manufacturers at all levels. Significant variation from the manufacturer's set values were noted as a function of tightening speed with the electronic torquing device. The data indicate that there is wide variation in the ability of clinicians to perceive adequate torquing forces applied to implant components. Calibrated torquing devices are mandatory if proper torquing procedures are to be accomplished.

Entities:  

Mesh:

Substances:

Year:  1994        PMID: 8206549

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  16 in total

1.  Experimental study on the factors affecting torque of beam-type implant torque wrenches.

Authors:  Hiroki Shiba; Yuji Sato; Junichi Furuya; Tokiko Osawa; Akio Isobe; Myu Hayashi; Noboru Kitagawa
Journal:  BMC Oral Health       Date:  2021-07-15       Impact factor: 2.757

2.  Analytical and experimental position stability of the abutment in different dental implant systems with a conical implant-abutment connection.

Authors:  Wiebke Semper-Hogg; Silvan Kraft; Sebastian Stiller; Juergen Mehrhof; Katja Nelson
Journal:  Clin Oral Investig       Date:  2012-07-22       Impact factor: 3.573

3.  Evaluation of stability of interface between CCM (Co-Cr-Mo) UCLA abutment and external hex implant.

Authors:  Ki-Joon Yoon; Young-Bum Park; Hyunmin Choi; Youngsung Cho; Jae-Hoon Lee; Keun-Woo Lee
Journal:  J Adv Prosthodont       Date:  2016-12-15       Impact factor: 1.904

4.  In Vitro Evaluation of Manual Torque Values Applied to Implant-Abutment Complex by Different Clinicians and Abutment Screw Loosening.

Authors:  Onur Dincer Kose; Burcin Karataslı; Sabit Demircan; Taha Emre Kose; Erhan Cene; Serhan Aydın Aya; Mehmet Ali Erdem; Abdulkadir Burak Cankaya
Journal:  Biomed Res Int       Date:  2017-04-03       Impact factor: 3.411

5.  The Accuracy of New and Aged Mechanical Torque Devices Employed in Five Dental Implant Systems.

Authors:  Mehmet Ali Erdem; Burcin Karatasli; Onur Dinçer Kose; Taha Emre Kose; Erhan Çene; Serhan Aydın Aya; Abdulkadir Burak Cankaya
Journal:  Biomed Res Int       Date:  2017-11-07       Impact factor: 3.411

Review 6.  Systematic analysis of factors that cause loss of preload in dental implants.

Authors:  S Nithyapriya; A S Ramesh; A Kirubakaran; Jeevitha Mani; J Raghunathan
Journal:  J Indian Prosthodont Soc       Date:  2018 Jul-Sep

7.  Accuracy of different electronic torque drivers: A comparative evaluation.

Authors:  Byeong-Dae Ko; KeunBaDa Son; Seok Hyon Kang; Kyu-Bok Lee
Journal:  J Adv Prosthodont       Date:  2019-12-18       Impact factor: 1.904

8.  The influence of screw type, alloy and cylinder position on the marginal fit of implant frameworks before and after laser welding.

Authors:  Daniela Castilio; Ana Paula Ribeiro do Vale Pedreira; Paulo Henrique Orlato Rossetti; Leylha Maria Nunes Rossetti; Wellington Cardoso Bonachela
Journal:  J Appl Oral Sci       Date:  2006-04       Impact factor: 2.698

Review 9.  Implant fractures: Rare but not exceptional.

Authors:  Sahitya Sanivarapu; Srinivas Moogla; Rupa Sruthi Kuntcham; Lakshmi Kanth Kolaparthy
Journal:  J Indian Soc Periodontol       Date:  2016 Jan-Feb

10.  Tightening torque of implant abutment using hand drivers against torque wrench and its effect on the internal surface of implant.

Authors:  Y Sameera; Rathika Rai
Journal:  J Indian Prosthodont Soc       Date:  2020-04-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.