Manual closing torque in five implant abutment systems: an in vitro comparative study.

M Gross1, D Kozak, B Z Laufer, E I Weiss.   

Abstract

STATEMENT OF
PURPOSE: A significant percentage of abutment/implant (A/I) assemblies tend to unscrew on functional loading.
PURPOSE: This in vitro study evaluated the maximum closing torque generated manually for comparison with recommended closing torque values in 5 implant systems.
METHODS: Closing torque generated by 9 operators with 5 manual torque drivers was measured and compared by using a mannequin and torque gauge assembly.
RESULTS: Closing torque values were found to be significantly different between operators and between implant systems (P <.0001). Mean closing torque values of 9 operators performing 5 closures for 5 systems ranged from 7 to 14.6 N/cm for habitual closures and from 9.4 to 19.9 N/cm for maximum closures.
CONCLUSIONS: The results demonstrated that maximum manual closure did not approach recommended closing torque in any of the system measured. Significant interoperator and intraoperator variability was found in the closing torque with manual drivers, and the driver diameter and grip were probably an important feature in the generation of high closing torque.

Mesh:

Substances:

Year:  1999        PMID: 10220662     DOI: 10.1016/s0022-3913(99)70212-x

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  7 in total

1.  Effect of early palliative care on quality of life in patients with non-small-cell lung cancer.

Authors:  H Zhuang; Y Ma; L Wang; H Zhang
Journal:  Curr Oncol       Date:  2018-02-28       Impact factor: 3.677

2.  Effect of cyclic load on vertical misfit of prefabricated and cast implant single abutment.

Authors:  Rudys Rodolfo de Jesus Tavarez; Wellington Cardoso Bonachela; Anuar Antônio Xible
Journal:  J Appl Oral Sci       Date:  2011 Jan-Feb       Impact factor: 2.698

3.  In Vitro Evaluation of Manual Torque Values Applied to Implant-Abutment Complex by Different Clinicians and Abutment Screw Loosening.

Authors:  Onur Dincer Kose; Burcin Karataslı; Sabit Demircan; Taha Emre Kose; Erhan Cene; Serhan Aydın Aya; Mehmet Ali Erdem; Abdulkadir Burak Cankaya
Journal:  Biomed Res Int       Date:  2017-04-03       Impact factor: 3.411

4.  Accuracy of different electronic torque drivers: A comparative evaluation.

Authors:  Byeong-Dae Ko; KeunBaDa Son; Seok Hyon Kang; Kyu-Bok Lee
Journal:  J Adv Prosthodont       Date:  2019-12-18       Impact factor: 1.904

5.  The effect of aging on the accuracy of new friction-style mechanical torque limiting devices for dental implants.

Authors:  Aboulfazl Saboury; Seyed Jalil Sadr; Ali Fayaz; Minoo Mahshid
Journal:  J Dent (Tehran)       Date:  2013-01-31

6.  Accuracy combining different brands of implants and abutments.

Authors:  María-Fernanda Solá-Ruíz; Eduardo Selva-Otaolaurruchi; Gisela Senent-Vicente; Inés González-de-Cossio; Vicente Amigó-Borrás
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2013-03-01

7.  Tightening torque of implant abutment using hand drivers against torque wrench and its effect on the internal surface of implant.

Authors:  Y Sameera; Rathika Rai
Journal:  J Indian Prosthodont Soc       Date:  2020-04-07
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.