| Literature DB >> 27246869 |
Arnaud Payerols1, Claudie Eliaou1, Véronique Trezeguet1, Max Villain1,2, Vincent Daien3,4,5.
Abstract
BACKGROUND: The PlusOptix photoscreeners (PlusOptix GmbH, Nuremberg, Germany) is used in many vision screening programs. The purpose of the present study was to further explore the accuracy of the PlusOptix A09 photoscreener in children with ametropia (myopia or hyperopia).Entities:
Keywords: Amblyopia; Cycloplegia; Refraction; Screening
Mesh:
Year: 2016 PMID: 27246869 PMCID: PMC4888737 DOI: 10.1186/s12886-016-0247-8
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Spherical equivalents, sphere, cylinder and axis for each method for the right eye
| Spherical equivalent | Sphere | Cylinder | Axis | |
|---|---|---|---|---|
| Non-cycloplegic autorefraction ( | −0.70 (3.14)a | −0.02 (3.33)d | −1.16 (1.35)g | 72.75 (66.00)j |
| Nidek ( | 0.04 (2.57) | 0.76 (2.72) | −1.13 (1.51) | 82.19 (72.05) |
| Retinomax ( | −2.10 (3.74) | −1.50 (3.98) | −1.21 (1.05) | 54.73 (50.79) |
| PlusOptix A09 | 0.54 (1.83)b | 1.27 (2.07)e | −1.46 (1.22)h | 84.41 (68.56)k |
| Cycloplegic autorefraction ( | 1.06 (2.04)c | 1.77 (2.20)f | −1.41 (0.98)i | 73.50 (67.29)l |
| Nidek ( | 1.06 (2.29) | 1.75 (2.46) | −1.38 (1.09) | 75.14 (71.47) |
| Retinomax ( | 1.06 (1.57) | 1.80 (1.71) | −1.48 (0.79) | 70.36 (61.69) |
Data are mean (SD)
a P = 0.004 between autorefractor (Nidek/Retinomax) and cycloplegic autorefraction
b P = 0.02 between autorefractor (Nidek/Retinomax) and PlusOptix A09
c P = 0.04 between PlusOptix A09 and cycloplegic autorefraction
d P < 0.01 between autorefractor (Nidek/Retinomax) and cycloplegic autorefraction
e P = 0.007 between autorefractor (Nidek/Retinomax) and PlusOptix A09
f P = 0.044 between PlusOptix A09 and cycloplegic autorefraction
g P = 0.09 between autorefractor (Nidek/Retinomax) and cycloplegic autorefraction
h P = 0.19 between autorefractor (Nidek/Retinomax) and PlusOptix A09
i P = 0.69 between PlusOptix A09 and cycloplegic autorefraction
j P < 0.01 between autorefractor (Nidek/Retinomax) and cycloplegic autorefraction
k P = 0.16 between autorefractor (Nidek/Retinomax) and PlusOptix A09
l P = 0.28 between PlusOptix A09 and cycloplegic autorefraction
Correlation of refraction value between
-PlusOptix A09 and cycloplegic autorefraction: r = 0.81, p < 0.001
-PlusOptix A09 and non-cycloplegic autorefraction: r = 0.70, p < 0.001
-Non-cycloplegic and cycloplegic autorefraction: r = 0.77, p < 0.001
Mean difference in spherical equivalents between PlusOptix A09 and cycloplegic autorefraction among children with hyperopia and myopia measured by cycloplegia
| Total | Hyperopia ( | Myopia ( |
| |
|---|---|---|---|---|
| Difference between PlusOptix A09 and cycloplegia autorefraction mean (SD) | 0.52 (1.2) | 0.73 (1.34) | 0.05 (0.66) | 0.01 |
| Children with < 1-D difference between PlusOptix A09 and cycloplegia autorefraction (%) | 68.75 % | 54.5 % | 90.0 % | 0.01 |
Correlation between cycloplegic autorefraction with PlusOptix A09 and cycloplegic autorefraction
-in myopic children (r = 0.91, p = 0.0002)
-in hyperopic children (r = 0.52, p = 0.01)
Fig. 1Correlation between autorefractor and PlusOptix A09. Top left graph: Correlation between non-cycloplegic autorefraction with PlusOptix A09 and cycloplegic autorefraction with a autorefractor (r = 0.81, p < 0.001); top right graph: non-cycloplegic autorefraction with PlusOptix A09 and with autorefractor (r = 0.70, p < 0.001), down left graph non-cycloplegic autorefraction and cycloplegic autorefraction (r = 0.77, p < 0.001)
Fig. 2Bland and Altman analysist. Top left graph: Correlation between the mean difference and difference in means between non-cycloplegic autorefraction with PlusOptix A09 and cycloplegic autorefraction (r = 0.19; p = 0.31), red curve: bias (mean difference between the measures = 0.52, blue curves: limit of agreements [−1.55 to +3.15 D]; top right graph: Correlation between the mean difference and difference in means between PlusOptix A09 refraction and non-cycloplegic autorefraction (r = −0.58; p = 0.0005), red curve: biais (mean difference between the measures = 1.24, blue curves: limit of agreements [−4.94 to 4.79 D]; down left graph: Correlation between the mean difference and difference in means between non-cycloplegic autorefraction and cycloplegic autorefraction (r = −0.51; p = 0.0029), red curve: bias (mean difference between the measures = 1.75, blue curves: limit of agreements [−4.42 to 4.79 D]
Fig. 3Correlation between non-cycloplegic autorefraction with PlusOptix A09 and cycloplegic autorefraction among myopic and hyperopic children: left graph: in myopic children (r = 0.91, p = 0.0002), right graph: in hyperopic children (r = 0.52, p = 0.01)