Literature DB >> 24738949

Comparing the SureSight autorefractor and the plusoptiX photoscreener for pediatric vision screening.

David Silbert1, Noelle Matta, Jing Tian, Eric Singman.   

Abstract

PURPOSE: Both the plusoptiX photoscreener photoscreener and the SureSight autorefractor can be used to objectively screen for amblyogenic risk factors in children. We compare the plusoptiX and the SureSight to a pediatric ophthalmology examination on a single cohort of children.
METHODS: We performed a retrospective medical records review of 90 children. Each patient was screened with both the PlusoptiX A09 and SureSight devices on the same day as part of a comprehensive pediatric ophthalmic examination. The ages of the patients ranged from <1 to 17 years. All patients had a cycloplegic refraction on the day of their exam, or within the previous 6 months.
RESULTS: Of the children reviewed, 71% had amblyopia risk factors. The plusoptiX demonstrated a sensitivity of 98% and a specificity of 88%. The SureSight demonstrated a sensitivity of 95% and a specificity of 65% utilizing the manufacturer's criteria.
CONCLUSION: The SureSight and plusoptiX were both found to be reliable objective vision screening devices, though the plusoptiX did have a higher sensitivity and specificity than the SureSight using the manufacturer's referral criteria. Using the Vision In Preschoolers Study (VIPS) referral criteria, the specificity of the SureSight improved but the sensitivity was reduced.

Entities:  

Keywords:  Amblyopia; SureSight; pediatric; photoscreening; plusoptiX; strabismus; vision screening; vision screening/diagnosis

Mesh:

Year:  2014        PMID: 24738949     DOI: 10.3109/09273972.2014.904896

Source DB:  PubMed          Journal:  Strabismus        ISSN: 0927-3972


  9 in total

1.  Performance of Plusoptix A09 Photo Screener in Refractive Error Screening in School Children Aged between 5 and 15 Years in the Southern Part of India.

Authors:  Avinash V Prabhu; Jyothi Thomas; Ramesh S Ve; Sayantan Biswas
Journal:  J Curr Ophthalmol       Date:  2020-07-04

2.  Accuracy of PlusOptix A09 distance refraction in pediatric myopia and hyperopia.

Authors:  Arnaud Payerols; Claudie Eliaou; Véronique Trezeguet; Max Villain; Vincent Daien
Journal:  BMC Ophthalmol       Date:  2016-06-01       Impact factor: 2.209

3.  Development of refractive error in children treated for retinopathy of prematurity with anti-vascular endothelial growth factor (anti-VEGF) agents: A meta-analysis and systematic review.

Authors:  Qing-Qing Tan; Stephen P Christiansen; Jingyun Wang
Journal:  PLoS One       Date:  2019-12-02       Impact factor: 3.240

4.  Does the Accuracy and Repeatability of Refractive Error Estimates Depend on the Measurement Principle of Autorefractors?

Authors:  Debananda Padhy; Shrikant R Bharadwaj; Suryasmita Nayak; Suryasnata Rath; Taraprasad Das
Journal:  Transl Vis Sci Technol       Date:  2021-01-05       Impact factor: 3.283

Review 5.  Scope and costs of autorefraction and photoscreening for childhood amblyopia-a systematic narrative review in relation to the EUSCREEN project data.

Authors:  Anna M Horwood; Helen J Griffiths; Jill Carlton; Paolo Mazzone; Arinder Channa; Mandy Nordmann; Huibert J Simonsz
Journal:  Eye (Lond)       Date:  2020-11-30       Impact factor: 3.775

6.  Comparative Validation of PlusoptiX and AI-Optic Photoscreeners in Children with High Amblyopia Risk Factor Prevalence.

Authors:  Robert W Arnold
Journal:  Clin Ophthalmol       Date:  2022-08-16

7.  Instrument referral criteria for PlusoptiX and SureSight based on 2021 AAPOS guidelines: A population-based study.

Authors:  Qi Yan; Rui Li; YingXiao Qian; Xiao Lin; Hui Zhu; Yue Wang; Xiaoyan Zhao; Xiaohan Zhang; Qigang Sun; Qingfeng Hao; Haohai Tong; Yue Zhu; Zhitong Li; Yan Zhu; Hu Liu; Dan Huang
Journal:  Front Public Health       Date:  2022-09-26

8.  Canon CP-TX1 camera - As a screening tool for amblyogenic risk factors.

Authors:  Syed Aisha Raza; Abadan Khan Amitava; Yogesh Gupta; Kamran Afzal; Farnaz Kauser; Juhi Saxena; Anam Masood; Aparna Bose; Ramprakash Singh
Journal:  Indian J Ophthalmol       Date:  2022-04       Impact factor: 2.969

9.  Pediatric vision screening using the plusoptiX A12C photoscreener in Chinese preschool children aged 3 to 4 years.

Authors:  Dan Huang; Xuejuan Chen; Xiaohan Zhang; Yue Wang; Hui Zhu; Hui Ding; Jing Bai; Ji Chen; Zhujun Fu; Zijin Wang; Hu Liu
Journal:  Sci Rep       Date:  2017-05-17       Impact factor: 4.379

  9 in total