Avinash V Prabhu1, Jyothi Thomas1, Ramesh S Ve1, Sayantan Biswas1.
Abstract
PURPOSE: To evaluate the performance of Plusoptix A09 in detecting ametropia, warranted against frequently-used technique of retinoscopy in children attending school (5-15 years) and its probability as a screening tool.Entities:
Keywords: Plusoptix; Refractive errors; Retinoscopy; School children; Sensitivity
Year: 2020 PMID: 32775802 PMCID: PMC7382526 DOI: 10.4103/JOCO.JOCO_76_20
Source DB: PubMed Journal: J Curr Ophthalmol ISSN: 2452-2325
Comparison between the Plusoptix A09, objective retinoscopy, and cycloplegic retinoscopy for spherical, cylindrical, and spherical equivalent values
| Total ( | Sphere (D) | Cylinder (D) | Spherical equivalent (D) | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Mean±SD | 95% CI | Mean±SD | 95% CI | Mean±SD | 95% CI | ||||
| Cycloplegic retinoscopy | 0.8±0.82 | 0.67-0.94 | NA | 0.34±0.93 | −0.55-0.25 | NA | 0.61±0.81 | 0.48-0.74 | NA |
| Objective retinoscopy | 0.65±0.69 | 0.54-0.76 | NA | 0.32±0.86 | −0.46-0.18 | NA | 0.45±0.7 | 0.38-0.60 | NA |
| Plusoptix | 1.12±1.16 | 0.94-1.31 | 0.0001 | 0.83±1.27 | −1.04-0.63 | 0.0001 | 0.71±1.06 | 0.53-0.88 | 0.097 |
D: Diopter, SD: Standard deviation, CI: Confidence interval, NA: Not available
Values on interobserver repeatability coefficient, coefficient of variance, within subject standard of deviation, and intraclass correlation of Plusoptix A09 spherical power, cylindrical power and axis
| 95% CI | ||||
|---|---|---|---|---|
| RC | CV | Sw | ICC | |
| Spherical power | 1.26 (1.12-1.40) | 40.54 (35.96-45.11) | 0.45 (0.40-0.51) | 0.92 (0.89-0.94) |
| Cylindrical power | 0.5 (0.45-0.56) | -21.76 (−19.31-24.21) | 0.18 (0.16-0.20) | 0.99 (0.98-0.99) |
| Cylindrical axis | 96.44 (85.52-107.35) | 45.57 (40.42-50.73) | 34.79 (30.85-38.73) | 0.70 (0.59-0.78) |
RC: Repeatability coefficient, CI: Confidence interval, CV: Coefficient of variance, Sw: Within subject standard deviation, ICC: Intraclass correlation coefficient
Figure 1Agreement between Plusoptix sphere and objective retinoscopy sphere using Bland Altman plot
Figure 2Agreement between Plusoptix sphere and cycloretinoscopy sphere using Bland Altman plot
Figure 3Agreement between Plusoptix spherical equivalent and objective retinoscopy spherical equivalent using Bland Altman plot
Figure 4Agreement between Plusoptix spherical equivalent and cycloretinoscopy spherical equivalent using Bland Altman plot
Figure 5Receiver operating characteristics curve to plot the sensitivity and specificity of Plusoptix. Area under the curve: 0.614, 95% confidence interval: 0.444–0.783, Standard error: 0.086