Abstract
AIM: To compare the clinical performance of the Spot Vision Screener used to detect amblyopia risk factors (ARFs) in children before and after induction of cycloplegia; the children were referred because they met the screening criteria of the American Association for Pediatric Ophthalmology and Strabismus (AAPOS).Entities:
Year: 2019 PMID: 31511789 PMCID: PMC6714320 DOI: 10.1155/2019/5329121
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Ambliyopia risk factors targeted by automated vision screening (2013 AAOPS guideline).
| Age (months) | Refractive risk factors targets (cycloplegic refraction) | |||
|---|---|---|---|---|
| Astigmatism | Hyperopia | Anisometropia | Myopia | |
| 12–30 | >2.0 D | >4.5 D | >2.5 D | >−3.5 D |
| 31–48 | >2.0 D | >4.0 D | >2.0 D | >−3.0 D |
| >48 | >1.5 D | >3.5 D | >1.5 D | >1.5 D |
| All ages | Nonrefractive amblyopia risk factor targets: manifest strabismus > 8 PD in primary position media opacity > 1 mm | |||
D: diopters; PD: prism diopters.
Median values of refractive parameters and power vectors using cycloplegic autorefraction and Spot Vision Screener with or without cycloplegia.
| Cycloplegic refraction | Spot screener | Cycloplegic spot screener | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Min | Max | Median | Min | Max | Median | Min | Max | Median | |
| S | −3.25 | 7.5 | 1.25 | −3.0 | 6.50 | 0.50 | −3.0 | 7.50 | 1.75 |
| C | −3.50 | 0.0 | −0.5 | −3.00 | 0.0 | −0.50 | −3.00 | 0.0 | −0.75 |
| SE | −3.50 | 7.38 | 1.0 | −3.25 | 6.25 | 0.25 | −3.25 | 7.50 | 1.25 |
|
| −1.29 | 1.64 | 0.21 | −0.56 | 2.12 | 0.24 | −1.49 | 1.96 | 0.32 |
|
| −0.56 | 0.74 | 0.0 | −0.55 | 0.97 | 0.0 | −0.38 | 0.61 | 0.0 |
S: spherical; C: cylindrical; SE: spherical equivalent.
Figure 1Correlations between spherical and spherical equivalent values of cycloplegic autorefractometer and Spot with or without cycloplegia.
Figure 2Correlations between J0 and J45 values of cycloplegic autorefractometer and Spot with or without cycloplegia.
Comparison of cycloplegic autorefraction with Spot Screener.
| Spherical | Cylindrical | Spherical equivalent | |
|---|---|---|---|
| CA vs Ss |
|
|
|
| CA vs cycloplegic Ss |
|
|
|
| Cycloplegic SV vs noncycloplegic Ss |
|
|
|
CA: cycloplegic autorefraction; Ss: spot screener. Wilcoxon p < 0.001.
Comparison of previous studies to current study of Spot Screener in detecting ARFs accordingly to 2013 AAOPS criteria.
| Study | Device | Age | Sensitivity | Specificity | PPV | NPV | Compared with |
|---|---|---|---|---|---|---|---|
|
| Spot (v2.0.16) | 11–221 | 84.8 | 70.9 | 78.1 | 79.2 | CR |
|
| Spot (v2.0.16) | 20–119 | 92.6 | 90.6 | 58.1 | 98.9 | CR |
|
| Spot (v2.0.16) | 6–35 | 89.8 | 70.4 | 58.9 | 93.6 | CR |
|
| Spot (2.1.4) | 26.4–110.4 | 84 | 62 | 58 | 86 | CR |
|
| Spot (version missing) | 48–84 | 94.79 | 85 | — | — | CR |
|
| Spot (v 2.1.4) | 48–72 | 94 | 80 | — | — | CR |
|
| Spot (v3.0.04.06) | 36–120 | 60.9 | 94.9 | 75.7 | 90.4 | CA |
| Spot cycloplegic | 36–120 | 85.3 | 87.4 | 63.6 | 95.8 | CA |
Age: months; PPV: positive predictive value; NPV: negative predictive value; CR: cycloplegic retinoscopy; CA: cycloplegic autorefractometer.