Zhale Rajavi1, Hiva Parsafar, Alireza Ramezani, Mehdi Yaseri.
Abstract
PURPOSE: To compare the accuracy of noncycloplegic photorefraction (NCP) with that of cycloplegic refraction (CR) for detecting refractive amblyopia risk factors (RARFs) and to determine cutoff points.Entities:
Keywords: Amblyopia; Cyclorefraction; Photorefraction; Screening
Year: 2012 PMID: 22737380 PMCID: PMC3381105
Source DB: PubMed Journal: J Ophthalmic Vis Res ISSN: 2008-322X
Comparison of refractive amblyopia risk factors between the two methods
| Cyclorefraction | n (%) | 22 (11.9) | 6 (3.2) | 28 (15.1) | 26 (14.1) |
| mean ± SD | +5.96 ± 2.13 | −5.63 ± 4.72 | −2.21 ± 0.80 | −1.76 ± 0.84 | |
| Noncycloplegic Photorefraction | n (%) | 15 (8.1) | 5 (2.7) | 33 (17.8) | 31 (16.8) |
| mean ± SD | +2.37 ± 1.36 | −4.38 ± 2.27 | −2.44 ± 0.97 | −1.56 ± 1.08 | |
| Difference | 95% CI | 2.68 (1.74 to 3.62) | −1.25 (−6.44 to −3.94) | 0.22 (−0.03 to −0.47) | −0.20 (−0.47 to −0.06) |
| P-value | < 0.001 | 0.563 | 0.077 | 0.127 | |
D, diopter; WTR, with the rule; ATR, against the rule; n, number; SD, standard deviation; CI, confidence interval
Figure 1.Pearson correlation between noncycloplegic photorefraction and cyclorefraction for spherical error, cylindrical error and spherical equivalent (r=0.76, r=0.86, r=0.76, respectively).
Agreement coefficient values for noncycloplegic photorefraction compared to standard cyclorefraction together with the calculated cutoff points based on ROC curve analysis
| Frequency | 22 | 6 | 43 | 57 |
| Sensitivity (%) | 45.4 | 83.3 | 85.7 | 79.0 |
| Specificity (%) | 96.9 | 100 | 93.0 | 94.5 |
| Positive predictive value (%) | 66.7 | 100 | 78.3 | 86.5 |
| Negative predictive value (%) | 92.9 | 99.4 | 95.7 | 91.0 |
| True positive | 10 | 5 | 36 | 45 |
| True negative | 158 | 179 | 132 | 121 |
| False positive | 5 | 0 | 10 | 7 |
| False negative | 12 | 1 | 7 | 12 |
| Overall agreement | 90.8 | 99.5 | 91.3 | 89.7 |
| Prescreening prevalence | 11.9 | 3.2 | 23.2 | 30.8 |
| Kappa index | 0.492 | 0.906 | 0.762 | 0.753 |
| Cutoff points (Diopters) | 1.87 | −3.0 | ±1.12 | - |
| Area under the curve | 89.6 | 99.1 | 94.2 | - |
| Sensitivity for cutoff points (%) | 81.8 | 83.3 | 95.3 | 91.2 |
| Specificity for cutoff points (%) | 84.7 | 100 | 87.4 | 82.0 |
| Positive predictive value (%) | 41.9 | 100 | 69.5 | 69.3 |
| Negative predictive value (%) | 97.2 | 99.4 | 99.2 | 95.4 |
| True positive | 18 | 5 | 41 | 52 |
| True negative | 138 | 179 | 125 | 105 |
| False positive | 25 | 0 | 18 | 23 |
| False negative | 4 | 1 | 1 | 5 |
Figure 2.ROC curves for noncycloplegic photorefraction in detecting hyperopia ≥ +3.5 D, myopia ≥ −3.0 D, and astigmatism ≥ 1.5 D. Areas under the curve were 89.6%, 99.1%, and 94.2%, respectively.