Literature DB >> 21708083

Accuracy of Plusoptix S04 in children and teens.

Yilmaz Kiyak Ayse1, Uretmen Onder, Kose Suheyla.   

Abstract

OBJECTIVE: To compare refraction measurements of young patients obtained using Plusoptix S04 with those obtained using cycloplegic retinoscopy.
DESIGN: Cohort study. PARTICIPANTS: Data were collected from 64 patients (128 eyes), aged 2-19 years.
METHODS: All eyes underwent photorefraction using Plusoptix S04 with or without cycloplegia and cycloplegic retinoscopy. Eyes with a maximum spherical and cylinder range of 7.00 to +5.00 were excluded from the study.
RESULTS: The difference in spherical equivalents between Plusoptix S04 without cycloplegia and cycloplegic retinoscopy was more than ±0.50 D in 70.2% of the eyes. In 100 eyes (78.1%), photorefraction underestimated the hyperopic refractive error. The cylinder powers determined by Plusoptix S04 and cycloplegic retinoscopy were similar. The difference in the spherical equivalents between Plusoptix S04 with cycloplegia and cycloplegic retinoscopy was ±0.50 D in 89.8% of the eyes. In 96 eyes (75%), photorefraction with cycloplegia overestimated the hyperopic refractive error. The difference in cylindrical power determined using Plusoptix S04 with cycloplegia and cycloplegic retinoscopy was statistically significant (paired t test, p <0.009).
CONCLUSIONS: The statistically significant difference in spherical equivalent values obtained using Plusoptix S04 with or without cycloplegia and cycloplegic retinoscopy suggests that Plusoptix S04 is more in evaluating the cylinder power and axis in young patients. Our results also suggest that Plusoptix S04 is not an accurate tool to estimate the refraction in children.

Entities:  

Mesh:

Year:  2011        PMID: 21708083     DOI: 10.3129/i10-110

Source DB:  PubMed          Journal:  Can J Ophthalmol        ISSN: 0008-4182            Impact factor:   1.882


  10 in total

1.  Comparison of non-cycloplegic photorefraction, cycloplegic photorefraction and cycloplegic retinoscopy in children.

Authors:  Ozdemir Ozdemir; Zuhal Özen Tunay; Ikbal Seza Petriçli; Damla Ergintürk Acar; Muhammet Kazım Erol
Journal:  Int J Ophthalmol       Date:  2015-02-18       Impact factor: 1.779

2.  Performance of Photoscreener in Detection of Refractive Error in All Age Groups and Amblyopia Risk Factors in Children in a Tribal District of Odisha: The Tribal Odisha Eye Disease Study (TOES) # 3.

Authors:  Lapam Panda; Umasankar Barik; Suryasmita Nayak; Biswajit Barik; Gyanaranjan Behera; Ramesh Kekunnaya; Taraprasad Das
Journal:  Transl Vis Sci Technol       Date:  2018-06-04       Impact factor: 3.283

3.  Performance of the Plusoptix A09 photoscreener in detecting amblyopia risk factors in Chinese children attending an eye clinic.

Authors:  Xiao-Ran Yan; Wan-Zhen Jiao; Zhi-Wei Li; Wen-Wen Xu; Feng-Jiao Li; Li-Hua Wang
Journal:  PLoS One       Date:  2015-06-01       Impact factor: 3.240

4.  Vision screening for children 36 to <72 months: recommended practices.

Authors:  Susan A Cotter; Lynn A Cyert; Joseph M Miller; Graham E Quinn
Journal:  Optom Vis Sci       Date:  2015-01       Impact factor: 1.973

5.  Accuracy of PlusOptix A09 distance refraction in pediatric myopia and hyperopia.

Authors:  Arnaud Payerols; Claudie Eliaou; Véronique Trezeguet; Max Villain; Vincent Daien
Journal:  BMC Ophthalmol       Date:  2016-06-01       Impact factor: 2.209

6.  Comparison of Autorefraction and Photorefraction with and without Cycloplegia Using 1% Tropicamide in Preschool Children.

Authors:  Ertuğrul Tan Yassa; Cihan Ünlü
Journal:  J Ophthalmol       Date:  2019-05-02       Impact factor: 1.909

7.  Comparison between Plusoptix A09 and gold standard cycloplegic refraction in preschool children and agreement to detect refractive amblyogenic risk factors.

Authors:  Jyothi Thomas; B Rajashekar; Asha Kamath; Parikshit Gogate
Journal:  Oman J Ophthalmol       Date:  2021-02-27

8.  Accuracy of Photorefraction and Wavefront-Based Autorefraction in Children Under 3 Years of Age.

Authors:  Caner Kara; Ikbal Seza Petricli
Journal:  Beyoglu Eye J       Date:  2021-02-13

9.  Accuracy and Repeatability of Refractive Error Measurements by Photorefractometry.

Authors:  Zhale Rajavi; Hamideh Sabbaghi; Ahmad Shojaei Baghini; Mehdi Yaseri; Koroush Sheibani; Ghazal Norouzi
Journal:  J Ophthalmic Vis Res       Date:  2015 Jul-Sep

10.  Pediatric vision screening using the plusoptiX A12C photoscreener in Chinese preschool children aged 3 to 4 years.

Authors:  Dan Huang; Xuejuan Chen; Xiaohan Zhang; Yue Wang; Hui Zhu; Hui Ding; Jing Bai; Ji Chen; Zhujun Fu; Zijin Wang; Hu Liu
Journal:  Sci Rep       Date:  2017-05-17       Impact factor: 4.379

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.