Predicting detection performance with model observers: Fourier domain or spatial domain?

Baiyu Chen1, Lifeng Yu1, Shuai Leng1, James Kofler1, Christopher Favazza1, Thomas Vrieze1, Cynthia McCollough1.   

Abstract

The use of Fourier domain model observer is challenged by iterative reconstruction (IR), because IR algorithms are nonlinear and IR images have noise texture different from that of FBP. A modified Fourier domain model observer, which incorporates nonlinear noise and resolution properties, has been proposed for IR and needs to be validated with human detection performance. On the other hand, the spatial domain model observer is theoretically applicable to IR, but more computationally intensive than the Fourier domain method. The purpose of this study is to compare the modified Fourier domain model observer to the spatial domain model observer with both FBP and IR images, using human detection performance as the gold standard. A phantom with inserts of various low contrast levels and sizes was repeatedly scanned 100 times on a third-generation, dual-source CT scanner at 5 dose levels and reconstructed using FBP and IR algorithms. The human detection performance of the inserts was measured via a 2-alternative-forced-choice (2AFC) test. In addition, two model observer performances were calculated, including a Fourier domain non-prewhitening model observer and a spatial domain channelized Hotelling observer. The performance of these two mode observers was compared in terms of how well they correlated with human observer performance. Our results demonstrated that the spatial domain model observer correlated well with human observers across various dose levels, object contrast levels, and object sizes. The Fourier domain observer correlated well with human observers using FBP images, but overestimated the detection performance using IR images.

Entities:  

Keywords:  Computed tomography (CT); channelized Hotelling observer (CHO); detectability index (d’); iterative reconstruction (IR); model observer; non-prewhitening

Year:  2016        PMID: 27239086      PMCID: PMC4879813          DOI: 10.1117/12.2216962

Source DB:  PubMed          Journal:  Proc SPIE Int Soc Opt Eng        ISSN: 0277-786X


  12 in total

1.  Characteristic image quality of a third generation dual-source MDCT scanner: Noise, resolution, and detectability.

Authors:  Justin Solomon; Joshua Wilson; Ehsan Samei
Journal:  Med Phys       Date:  2015-08       Impact factor: 4.071

2.  Comparison of model and human observer performance for detection and discrimination tasks using dual-energy x-ray images.

Authors:  Samuel Richard; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

3.  Technical Note: Measuring contrast- and noise-dependent spatial resolution of an iterative reconstruction method in CT using ensemble averaging.

Authors:  Lifeng Yu; Thomas J Vrieze; Shuai Leng; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

4.  Evaluating iterative reconstruction performance in computed tomography.

Authors:  Baiyu Chen; Juan Carlos Ramirez Giraldo; Justin Solomon; Ehsan Samei
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

5.  Impact of Number of Repeated Scans on Model Observer Performance for a Low-contrast Detection Task in CT.

Authors:  Chi Ma; Lifeng Yu; Baiyu Chen; Thomas Vrieze; Shuai Leng; Cynthia McCollough
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-02-21

6.  Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms.

Authors:  Samuel Richard; Daniela B Husarik; Girijesh Yadava; Simon N Murphy; Ehsan Samei
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

7.  Assessment of volumetric noise and resolution performance for linear and nonlinear CT reconstruction methods.

Authors:  Baiyu Chen; Olav Christianson; Joshua M Wilson; Ehsan Samei
Journal:  Med Phys       Date:  2014-07       Impact factor: 4.071

8.  Analysis of Fourier-domain task-based detectability index in tomosynthesis and cone-beam CT in relation to human observer performance.

Authors:  Grace J Gang; Junghoon Lee; J Webster Stayman; Daniel J Tward; W Zbijewski; Jerry L Prince; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

Review 9.  Visual perception studies and observer models in medical imaging.

Authors:  Arthur E Burgess
Journal:  Semin Nucl Med       Date:  2011-11       Impact factor: 4.446

10.  Correlation between model observer and human observer performance in CT imaging when lesion location is uncertain.

Authors:  Shuai Leng; Lifeng Yu; Yi Zhang; Rickey Carter; Alicia Y Toledano; Cynthia H McCollough
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

View more
  2 in total

1.  Tilted-wire method for measuring resolution properties of CT images under extremely low-contrast and high-noise conditions.

Authors:  Chiaki Tominaga; Hiroki Azumi; Mitsunori Goto; Masaaki Taura; Noriyasu Homma; Issei Mori
Journal:  Radiol Phys Technol       Date:  2018-02-23

2.  Determination of iodine detectability in different types of multiple-energy images for a photon-counting detector computed tomography system.

Authors:  Wei Zhou; Gregory Michalak; Jayse Weaver; Andrea Ferrero; Hao Gong; Kenneth A Fetterly; Cynthia H McCollough; Shuai Leng
Journal:  J Med Imaging (Bellingham)       Date:  2019-10-15
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.