Evaluating iterative reconstruction performance in computed tomography.

Baiyu Chen1, Juan Carlos Ramirez Giraldo2, Justin Solomon1, Ehsan Samei3.   

Abstract

PURPOSE: Iterative reconstruction (IR) offers notable advantages in computed tomography (CT). However, its performance characterization is complicated by its potentially nonlinear behavior, impacting performance in terms of specific tasks. This study aimed to evaluate the performance of IR with both task-specific and task-generic strategies.
METHODS: The performance of IR in CT was mathematically assessed with an observer model that predicted the detection accuracy in terms of the detectability index (d'). d' was calculated based on the properties of the image noise and resolution, the observer, and the detection task. The characterizations of image noise and resolution were extended to accommodate the nonlinearity of IR. A library of tasks was mathematically modeled at a range of sizes (radius 1-4 mm), contrast levels (10-100 HU), and edge profiles (sharp and soft). Unique d' values were calculated for each task with respect to five radiation exposure levels (volume CT dose index, CTDIvol: 3.4-64.8 mGy) and four reconstruction algorithms (filtered backprojection reconstruction, FBP; iterative reconstruction in imaging space, IRIS; and sinogram affirmed iterative reconstruction with strengths of 3 and 5, SAFIRE3 and SAFIRE5; all provided by Siemens Healthcare, Forchheim, Germany). The d' values were translated into the areas under the receiver operating characteristic curve (AUC) to represent human observer performance. For each task and reconstruction algorithm, a threshold dose was derived as the minimum dose required to achieve a threshold AUC of 0.9. A task-specific dose reduction potential of IR was calculated as the difference between the threshold doses for IR and FBP. A task-generic comparison was further made between IR and FBP in terms of the percent of all tasks yielding an AUC higher than the threshold.
RESULTS: IR required less dose than FBP to achieve the threshold AUC. In general, SAFIRE5 showed the most significant dose reduction potentials (11-54 mGy, 77%-84%), followed by SAFIRE3 (7-36 mGy, 50%-61%) and IRIS (6-26 mGy, 37%-50%). The dose reduction potentials highly depended on task size and task contrast, with tasks of lower contrasts and smaller sizes, i.e., more challenging tasks, indicating higher dose reductions. Softer edge profile showed higher dose reduction potentials with SAFIRE3 and SAFIRE5, but not with IRIS. The task-generic comparison between IR and FBP demonstrated the overall superiority of IR performance, as IR allowed a larger percent of tasks to exceed the threshold AUC: IRIS, 8%-12%; SAFIRE3, 10%-16%; and SAFIRE5, 20%-33%. The improvement with IR was generally more pronounced at lower dose levels.
CONCLUSIONS: Expanding beyond traditional contrast and noise based assessments of IR, we performed both task-specific and task-generic evaluations of IR performance. The task-specific evaluation demonstrated the dependency of IR's dose reduction potential on task attributes, which can be employed to optimize IR for clinical indications with specific range of size and contrast. The task-generic evaluation demonstrated IR's overall superiority over FBP in terms of the range of tasks exceeding a threshold performance level, which can be employed for general comparisons between algorithms.

Entities:  

Mesh:

Year:  2014        PMID: 25471973     DOI: 10.1118/1.4901670

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  18 in total

Review 1.  Imaging Techniques to Aid IR Treatment of Musculoskeletal Malignancy.

Authors:  A Kyle Jones; Steven Yevich
Journal:  Semin Intervent Radiol       Date:  2019-03-29       Impact factor: 1.513

2.  Correlation between human detection accuracy and observer model-based image quality metrics in computed tomography.

Authors:  Justin Solomon; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2016-09-22

3.  Dose reduction potential of iterative reconstruction algorithms in neck CTA-a simulation study.

Authors:  Stephan Ellmann; Ferdinand Kammerer; Thomas Allmendinger; Michael Brand; Rolf Janka; Matthias Hammon; Michael M Lell; Michael Uder; Manuel Kramer
Journal:  Dentomaxillofac Radiol       Date:  2016-08-19       Impact factor: 2.419

4.  Predicting detection performance with model observers: Fourier domain or spatial domain?

Authors:  Baiyu Chen; Lifeng Yu; Shuai Leng; James Kofler; Christopher Favazza; Thomas Vrieze; Cynthia McCollough
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-30

5.  Use of a channelized Hotelling observer to assess CT image quality and optimize dose reduction for iteratively reconstructed images.

Authors:  Christopher P Favazza; Andrea Ferrero; Lifeng Yu; Shuai Leng; Kyle L McMillan; Cynthia H McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2017-10-03

6.  Pediatric head computed tomography with advanced modeled iterative reconstruction: focus on image quality and reduction of radiation dose.

Authors:  Hyun-Hae Cho; So Mi Lee; Sun Kyoung You
Journal:  Pediatr Radiol       Date:  2019-10-19

7.  Effect of Radiation Dose Reduction and Reconstruction Algorithm on Image Noise, Contrast, Resolution, and Detectability of Subtle Hypoattenuating Liver Lesions at Multidetector CT: Filtered Back Projection versus a Commercial Model-based Iterative Reconstruction Algorithm.

Authors:  Justin Solomon; Daniele Marin; Kingshuk Roy Choudhury; Bhavik Patel; Ehsan Samei
Journal:  Radiology       Date:  2017-02-07       Impact factor: 11.105

8.  Detection of Colorectal Hepatic Metastases Is Superior at Standard Radiation Dose CT versus Reduced Dose CT.

Authors:  Corey T Jensen; Nicolaus A Wagner-Bartak; Lan N Vu; Xinming Liu; Bharat Raval; David Martinez; Wei Wei; Yuan Cheng; Ehsan Samei; Shiva Gupta
Journal:  Radiology       Date:  2018-11-27       Impact factor: 11.105

9.  Statistical model based iterative reconstruction in clinical CT systems. Part III. Task-based kV/mAs optimization for radiation dose reduction.

Authors:  Ke Li; Daniel Gomez-Cardona; Jiang Hsieh; Meghan G Lubner; Perry J Pickhardt; Guang-Hong Chen
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

10.  Construction of realistic phantoms from patient images and a commercial three-dimensional printer.

Authors:  Shuai Leng; Baiyu Chen; Thomas Vrieze; Joel Kuhlmann; Lifeng Yu; Amy Alexander; Jane Matsumoto; Jonathan Morris; Cynthia H McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2016-07-07