Correlation between model observer and human observer performance in CT imaging when lesion location is uncertain.

Shuai Leng1, Lifeng Yu, Yi Zhang, Rickey Carter, Alicia Y Toledano, Cynthia H McCollough.   

Abstract

PURPOSE: The purpose of this study was to investigate the correlation between model observer and human observer performance in CT imaging for the task of lesion detection and localization when the lesion location is uncertain.
METHODS: Two cylindrical rods (3-mm and 5-mm diameters) were placed in a 35×26 cm torso-shaped water phantom to simulate lesions with -15 HU contrast at 120 kV. The phantom was scanned 100 times on a 128-slice CT scanner at each of four dose levels (CTDIvol=5.7, 11.4, 17.1, and 22.8 mGy). Regions of interest (ROIs) around each lesion were extracted to generate images with signal-present, with each ROI containing 128×128 pixels. Corresponding ROIs of signal-absent images were generated from images without lesion mimicking rods. The location of the lesion (rod) in each ROI was randomly distributed by moving the ROIs around each lesion. Human observer studies were performed by having three trained observers identify the presence or absence of lesions, indicating the lesion location in each image and scoring confidence for the detection task on a 6-point scale. The same image data were analyzed using a channelized Hotelling model observer (CHO) with Gabor channels. Internal noise was added to the decision variables for the model observer study. Area under the curve (AUC) of ROC and localization ROC (LROC) curves were calculated using a nonparametric approach. The Spearman's rank order correlation between the average performance of the human observers and the model observer performance was calculated for the AUC of both ROC and LROC curves for both the 3- and 5-mm diameter lesions.
RESULTS: In both ROC and LROC analyses, AUC values for the model observer agreed well with the average values across the three human observers. The Spearman's rank order correlation values for both ROC and LROC analyses for both the 3- and 5-mm diameter lesions were all 1.0, indicating perfect rank ordering agreement of the figures of merit (AUC) between the average performance of the human observers and the model observer performance.
CONCLUSIONS: In CT imaging of different sizes of low-contrast lesions (-15 HU), the performance of CHO with Gabor channels was highly correlated with human observer performance for the detection and localization tasks with uncertain lesion location in CT imaging at four clinically relevant dose levels. This suggests the ability of Gabor CHO model observers to meaningfully assess CT image quality for the purpose of optimizing scan protocols and radiation dose levels in detection and localization tasks for low-contrast lesions.

Entities:  

Mesh:

Year:  2013        PMID: 23927322      PMCID: PMC3724792          DOI: 10.1118/1.4812430

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  37 in total

1.  Using human and model performance to compare MRI reconstructions.

Authors:  M Dylan Tisdall; M Stella Atkins
Journal:  IEEE Trans Med Imaging       Date:  2006-11       Impact factor: 10.048

2.  Small nodule detectability evaluation using a generalized scan-statistic model.

Authors:  Lucreţiu M Popescu; Robert M Lewitt
Journal:  Phys Med Biol       Date:  2006-11-15       Impact factor: 3.609

Review 3.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

4.  Effect of dose reduction on the detection of mammographic lesions: a mathematical observer model analysis.

Authors:  Amarpreet S Chawla; Ehsan Samei; Robert Saunders; Craig Abbey; David Delong
Journal:  Med Phys       Date:  2007-08       Impact factor: 4.071

5.  Evaluation of internal noise methods for Hotelling observer models.

Authors:  Yani Zhang; Binh T Pham; Miguel P Eckstein
Journal:  Med Phys       Date:  2007-08       Impact factor: 4.071

6.  A channelized Hotelling observer study of lesion detection in SPECT MAP reconstruction using anatomical priors.

Authors:  S Kulkarni; P Khurd; I Hsiao; L Zhou; G Gindi
Journal:  Phys Med Biol       Date:  2007-05-23       Impact factor: 3.609

7.  Automated computer evaluation and optimization of image compression of x-ray coronary angiograms for signal known exactly detection tasks.

Authors:  Miguel Eckstein; Jay Bartroff; Craig Abbey; James Whiting; Francois Bochud
Journal:  Opt Express       Date:  2003-03-10       Impact factor: 3.894

8.  Visual signal detection. IV. Observer inconsistency.

Authors:  A E Burgess; B Colborne
Journal:  J Opt Soc Am A       Date:  1988-04       Impact factor: 2.129

9.  Adaptive detection mechanisms in globally statistically nonstationary-oriented noise.

Authors:  Yani Zhang; Craig K Abbey; Miguel P Eckstein
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2006-07       Impact factor: 2.129

10.  Human observer detection experiments with mammograms and power-law noise.

Authors:  A E Burgess; F L Jacobson; P F Judy
Journal:  Med Phys       Date:  2001-04       Impact factor: 4.071

View more
  41 in total

1.  Achieving routine submillisievert CT scanning: report from the summit on management of radiation dose in CT.

Authors:  Cynthia H McCollough; Guang Hong Chen; Willi Kalender; Shuai Leng; Ehsan Samei; Katsuyuki Taguchi; Ge Wang; Lifeng Yu; Roderic I Pettigrew
Journal:  Radiology       Date:  2012-06-12       Impact factor: 11.105

2.  Computational and human observer image quality evaluation of low dose, knowledge-based CT iterative reconstruction.

Authors:  Brendan L Eck; Rachid Fahmi; Kevin M Brown; Stanislav Zabic; Nilgoun Raihani; Jun Miao; David L Wilson
Journal:  Med Phys       Date:  2015-10       Impact factor: 4.071

3.  Anthropomorphic model observer performance in three-dimensional detection task for low-contrast computed tomography.

Authors:  Alexandre Ba; Miguel P Eckstein; Damien Racine; Julien G Ott; Francis Verdun; Sabine Kobbe-Schmidt; François O Bochud
Journal:  J Med Imaging (Bellingham)       Date:  2015-12-29

4.  Channelized Hotelling observer correlation with human observers for low-contrast detection in liver CT images.

Authors:  Alexandre Ba; Craig K Abbey; Damien Racine; Anaïs Viry; Francis R Verdun; Sabine Schmidt; François O Bochud
Journal:  J Med Imaging (Bellingham)       Date:  2019-05-20

5.  A Web-Based Software Platform for Efficient and Quantitative CT Image Quality Assessment and Protocol Optimization.

Authors:  Mingdong Fan; Theodore Thayib; Liqiang Ren; Scott Hsieh; Cynthia McCollough; David Holmes; Lifeng Yu
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2021-02-15

6.  Optimization of digital breast tomosynthesis (DBT) acquisition parameters for human observers: effect of reconstruction algorithms.

Authors:  Rongping Zeng; Aldo Badano; Kyle J Myers
Journal:  Phys Med Biol       Date:  2017-02-02       Impact factor: 3.609

7.  Technical Note: Measuring contrast- and noise-dependent spatial resolution of an iterative reconstruction method in CT using ensemble averaging.

Authors:  Lifeng Yu; Thomas J Vrieze; Shuai Leng; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

8.  Impact of Number of Repeated Scans on Model Observer Performance for a Low-contrast Detection Task in CT.

Authors:  Chi Ma; Lifeng Yu; Baiyu Chen; Thomas Vrieze; Shuai Leng; Cynthia McCollough
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015-02-21

9.  Correlation between human detection accuracy and observer model-based image quality metrics in computed tomography.

Authors:  Justin Solomon; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2016-09-22

10.  Determination of Optimal Image Type and Lowest Detectable Concentration for Iodine Detection on a Photon Counting Detector-Based Multi-Energy CT System.

Authors:  Wei Zhou; Rachel Schornak; Gregory Michalak; Jayse Weaver; Dilbar Abdurakhimova; Andrea Ferrero; Kenneth A Fetterly; Cynthia H McCollough; Shuai Leng
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2018-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.