Literature DB >> 22830744

Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms.

Samuel Richard1, Daniela B Husarik, Girijesh Yadava, Simon N Murphy, Ehsan Samei.   

Abstract

PURPOSE: To investigate a measurement method for evaluating the resolution properties of CT imaging systems across reconstruction algorithms, dose, and contrast.
METHODS: An algorithm was developed to extract the task-based modulation transfer function (MTF) from disk images generated from the rod inserts in the ACR phantom (model 464 Gammex, WI). These inserts are conventionally employed for HU accuracy assessment. The edge of the disk objects was analyzed to determine the edge-spread function, which was differentiated to yield the line-spread function and Fourier-transformed to generate the object-specific MTF for task-based assessment, denoted MTF(Task). The proposed MTF measurement method was validated against the conventional wire technique and further applied to measure the MTF of CT images reconstructed with an adaptive statistical iterative algorithm (ASIR) and a model-based iterative (MBIR) algorithm. Results were further compared to the standard filtered back projection (FBP) algorithm. Measurements were performed and compared across different doses and contrast levels to ascertain the MTF(Task) dependencies on those factors.
RESULTS: For the FBP reconstructed images, the MTF(Task) measured with the inserts were the same as the MTF measured from the wire-based method. For the ASIR and MBIR data, the MTF(Task) using the high contrast insert was similar to the wire-based MTF and equal or superior to that of FBP. However, results for the MTF(Task) measured using the low-contrast inserts, the MTF(Task) for ASIR and MBIR data was lower than for the FBP, which was constant throughout all measurements. Similarly, as a function of mA, the MTF(Task) for ASIR and MBIR varied as a function of noise--with MTF(Task) being proportional to mA. Overall greater variability of MTF(Task) across dose and contrast was observed for MBIR than for ASIR.
CONCLUSIONS: This approach provides a method for assessing the task-based MTF of a CT system using conventional and iterative reconstructions. Results demonstrated that the object-specific MTF can vary as a function of dose and contrast. The analysis highlighted the paradigm shift for iterative reconstructions when compared to FBP, where iterative reconstructions generally offer superior noise performance but with varying resolution as a function of dose and contrast. The MTF(Task) generated by this method is expected to provide a more comprehensive assessment of image resolution across different reconstruction algorithms and imaging tasks.

Mesh:

Year:  2012        PMID: 22830744     DOI: 10.1118/1.4725171

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  75 in total

1.  Low-dose CT screening using hybrid iterative reconstruction: confidence ratings of diagnoses of simulated lesions other than lung cancer.

Authors:  N Sakai; H Yabuuchi; M Kondo; Y Matsuo; T Kamitani; M Nagao; M Jinnouchi; M Yonezawa; T Kojima; Y Yano; H Honda
Journal:  Br J Radiol       Date:  2015-07-08       Impact factor: 3.039

2.  Full Dose-Reduction Potential of Statistical Iterative Reconstruction for Head CT Protocols in a Predominantly Pediatric Population.

Authors:  A E Mirro; S L Brady; R A Kaufman
Journal:  AJNR Am J Neuroradiol       Date:  2016-04-07       Impact factor: 3.825

3.  Effects of Image Quality on the Fundamental Limits of Image Registration Accuracy.

Authors:  Michael D Ketcha; Tharindu De Silva; Runze Han; Ali Uneri; Joseph Goerres; Matthew W Jacobson; Sebastian Vogt; Gerhard Kleinszig; Jeffrey H Siewerdsen
Journal:  IEEE Trans Med Imaging       Date:  2017-07-11       Impact factor: 10.048

4.  Technical Note: Measuring contrast- and noise-dependent spatial resolution of an iterative reconstruction method in CT using ensemble averaging.

Authors:  Lifeng Yu; Thomas J Vrieze; Shuai Leng; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

5.  Optimizing radiation dose by using advanced modelled iterative reconstruction in high-pitch coronary CT angiography.

Authors:  Sonja Gordic; Lotus Desbiolles; Martin Sedlmair; Robert Manka; André Plass; Bernhard Schmidt; Daniela B Husarik; Francesco Maisano; Simon Wildermuth; Hatem Alkadhi; Sebastian Leschka
Journal:  Eur Radiol       Date:  2015-06-03       Impact factor: 5.315

6.  Aliased noise in X-ray CT images and band-limiting processing as a preventive measure.

Authors:  Kazuhiro Sato; Miho Shidahara; Mitsunori Goto; Isao Yanagawa; Noriyasu Homma; Issei Mori
Journal:  Radiol Phys Technol       Date:  2015-01-11

7.  Radiation dose optimization for the bolus tracking technique in abdominal computed tomography: usefulness of real-time iterative reconstruction for monitoring scan.

Authors:  Yuya Ishikawa; Atsushi Urikura; Tsukasa Yoshida; Keisuke Takiguchi; Yoshihiro Nakaya
Journal:  Radiol Phys Technol       Date:  2016-09-30

8.  Correlation between human detection accuracy and observer model-based image quality metrics in computed tomography.

Authors:  Justin Solomon; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2016-09-22

9.  Quality evaluation of image-based iterative reconstruction for CT: Comparison with hybrid iterative reconstruction.

Authors:  Hiroki Kawashima; Katsuhiro Ichikawa; Kosuke Matsubara; Hiroji Nagata; Tadanori Takata; Satoshi Kobayashi
Journal:  J Appl Clin Med Phys       Date:  2019-05-02       Impact factor: 2.102

10.  Model-based iterative reconstruction: effect on patient radiation dose and image quality in pediatric body CT.

Authors:  Ethan A Smith; Jonathan R Dillman; Mitchell M Goodsitt; Emmanuel G Christodoulou; Nahid Keshavarzi; Peter J Strouse
Journal:  Radiology       Date:  2013-10-29       Impact factor: 11.105

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.