Characteristic image quality of a third generation dual-source MDCT scanner: Noise, resolution, and detectability.

Justin Solomon1, Joshua Wilson2, Ehsan Samei3.   

Abstract

PURPOSE: The purpose of this work was to assess the inherent image quality characteristics of a new multidetector computed tomography system in terms of noise, resolution, and detectability index as a function of image acquisition and reconstruction for a range of clinically relevant settings.
METHODS: A multisized image quality phantom (37, 30, 23, 18.5, and 12 cm physical diameter) was imaged on a SOMATOM Force scanner (Siemens Medical Solutions) under variable dose, kVp, and tube current modulation settings. Images were reconstructed with filtered back projection (FBP) and with advanced modeled iterative reconstruction (ADMIRE) with iterative strengths of 3, 4, and 5. Image quality was assessed in terms of the noise power spectrum (NPS), task transfer function (TTF), and detectability index for a range of detection tasks (contrasts of approximately 45, 90, 300, -900, and 1000 HU, and 2-20 mm diameter) based on a non-prewhitening matched filter model observer with eye filter.
RESULTS: Image noise magnitude decreased with decreasing phantom size, increasing dose, and increasing ADMIRE strength, offering up to 64% noise reduction relative to FBP. Noise texture in terms of the NPS was similar between FBP and ADMIRE (<5% shift in peak frequency). The resolution, based on the TTF, improved with increased ADMIRE strength by an average of 15% in the TTF 50% frequency for ADMIRE-5. The detectability index increased with increasing dose and ADMIRE strength by an average of 55%, 90%, and 163% for ADMIRE 3, 4, and 5, respectively. Assessing the impact of mA modulation for a fixed average dose over the length of the phantom, detectability was up to 49% lower in smaller phantom sections and up to 26% higher in larger phantom sections for the modulated scan compared to a fixed tube current scan. Overall, the detectability exhibited less variability with phantom size for modulated scans compared to fixed tube current scans.
CONCLUSIONS: Image quality increased with increasing dose and decreasing phantom size. The CT system exhibited nonlinear noise and resolution properties, especially at very low-doses, large phantom sizes, and for low-contrast objects. Objective image quality metrics generally increased with increasing dose and ADMIRE strength, and with decreasing phantom size. The ADMIRE algorithm could offer comparable image quality at reduced doses or improved image quality at the same dose. The use of tube current modulation resulted in more consistent image quality with changing phantom size.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26233220     DOI: 10.1118/1.4923172

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  29 in total

1.  DukeSim: A Realistic, Rapid, and Scanner-Specific Simulation Framework in Computed Tomography.

Authors:  Ehsan Abadi; Brian Harrawood; Shobhit Sharma; Anuj Kapadia; William P Segars; Ehsan Samei
Journal:  IEEE Trans Med Imaging       Date:  2018-12-12       Impact factor: 10.048

2.  Correlation between human detection accuracy and observer model-based image quality metrics in computed tomography.

Authors:  Justin Solomon; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2016-09-22

Review 3.  Intraoperative Imaging and Image Fusion for Venous Interventions.

Authors:  Ponraj Chinnadurai; Jean Bismuth
Journal:  Methodist Debakey Cardiovasc J       Date:  2018 Jul-Sep

4.  Paper-based 3D printing of anthropomorphic CT phantoms: Feasibility of two construction techniques.

Authors:  Paul Jahnke; Stephan Schwarz; Marco Ziegert; Felix Benjamin Schwarz; Bernd Hamm; Michael Scheel
Journal:  Eur Radiol       Date:  2018-08-16       Impact factor: 5.315

5.  Systematic analysis of bias and variability of morphologic features for lung lesions in computed tomography.

Authors:  Jocelyn Hoye; Justin Solomon; Thomas J Sauer; Marthony Robins; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2019-03-26

Review 6.  Using imaging as a biomarker for asthma.

Authors:  Abhaya Trivedi; Chase Hall; Eric A Hoffman; Jason C Woods; David S Gierada; Mario Castro
Journal:  J Allergy Clin Immunol       Date:  2017-01       Impact factor: 10.793

7.  Impact of bowtie filter and object position on the two-dimensional noise power spectrum of a clinical MDCT system.

Authors:  Daniel Gomez-Cardona; Juan Pablo Cruz-Bastida; Ke Li; Adam Budde; Jiang Hsieh; Guang-Hong Chen
Journal:  Med Phys       Date:  2016-08       Impact factor: 4.071

8.  Predicting detection performance with model observers: Fourier domain or spatial domain?

Authors:  Baiyu Chen; Lifeng Yu; Shuai Leng; James Kofler; Christopher Favazza; Thomas Vrieze; Cynthia McCollough
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2016-03-30

9.  Quality evaluation of image-based iterative reconstruction for CT: Comparison with hybrid iterative reconstruction.

Authors:  Hiroki Kawashima; Katsuhiro Ichikawa; Kosuke Matsubara; Hiroji Nagata; Tadanori Takata; Satoshi Kobayashi
Journal:  J Appl Clin Med Phys       Date:  2019-05-02       Impact factor: 2.102

10.  Detection of Colorectal Hepatic Metastases Is Superior at Standard Radiation Dose CT versus Reduced Dose CT.

Authors:  Corey T Jensen; Nicolaus A Wagner-Bartak; Lan N Vu; Xinming Liu; Bharat Raval; David Martinez; Wei Wei; Yuan Cheng; Ehsan Samei; Shiva Gupta
Journal:  Radiology       Date:  2018-11-27       Impact factor: 11.105

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.