Errors in 14 pulse oximeters during profound hypoxia.

J W Severinghaus1, K H Naifeh, S O Koh.   

Abstract

The accuracy of pulse oximeters from fourteen manufacturers was tested during profound brief hypoxic plateaus in 125 subject sets using 50 normal adult volunteers, of whom 29 were studied two to nine times. A data set usually consisted of 10 subjects, and 13 sets were collected between August 1987 and July 1988. In the first 6 sets, six 30-second hypoxic plateaus were obtained per subject at 55 +/- 6% oxyhemoglobin (O2Hb) (range, 40 to 70%). In the last 7 sets, three hypoxic plateaus were obtained at each of four levels, approximately 86, 74, 62, and 50% O2Hb, for the purpose of linear regression analysis. Inspired oxygen was adjusted manually breath by breath in response to arterial oxygen saturation computed on-line from end-tidal oxygen and carbon dioxide tensions. End-plateau arterial blood O2Hb was analyzed by a Radiometer OSM-3 oximeter, and plateau pulse oximeter saturation (SpO2) was read by cursor from a computer record of the analog output. Three to 13 instruments were tested simultaneously by using 1 to 3 duplicate instruments from each of one to seven manufacturers. Variations introduced by manufacturers were tested on subsequent sets in several instruments. An index of error, "ambiguity" (alpha) of oxygen saturation, was defined as the absolute sum of bias and precision (mean and SD of SpO2 - O2Hb) at O2Hb = 55.8 +/- 4.5%, preserving the sign when bias was significant at P less than 0.05. Ambiguity values for finger probes (unless specified) with latest data were: Physio-Control, 3.9 (ear, 3.3); Puritan-Bennett, -4.4; Criticare, 5.8 (forehead, 4.7); Kontron, 5.9 (infant probe) and 6.1 (ear, 5.8; forehead, 7.1); Biochem, -6.0; Datex 6.4 (ear, 6.9; forehead, 6.8); Critikon, 8.4; SiMed, 8.6; Marquest, 9.0; Novametrix, 10.2; Invivo, -12.2 (ear, -14.3); Nellcor, -15.1; Ohmeda, -21.2; and Radiometer, -21.2 (ear, -9.6). Linear regression slopes of 36 instruments from twelve manufacturers generally deviated from 1 in proportion to alpha. The data showed substantial differences in bias and precision between pulse oximeters at low saturations, the most common problems being underestimation of saturation and failing precision.

Entities:  

Mesh:

Substances:

Year:  1989        PMID: 2723709     DOI: 10.1007/BF01617877

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  9 in total

1.  Continuous emergency department monitoring of arterial saturation in adult patients with respiratory distress.

Authors:  J Jones; D Heiselman; L Cannon; R Gradisek
Journal:  Ann Emerg Med       Date:  1988-05       Impact factor: 5.721

2.  Accuracy of pulse oximetry during arterial oxyhemoglobin desaturation in dogs.

Authors:  M J Sendak; A P Harris; R T Donham
Journal:  Anesthesiology       Date:  1988-01       Impact factor: 7.892

3.  Use of pulse oximetry in neonatal anesthesia.

Authors:  K Miyasaka; M Katayama; I Kusakawa; J Ohata; T Kawano; Y Honma
Journal:  J Perinatol       Date:  1987       Impact factor: 2.521

4.  Range of accuracy of two wavelength oximetry.

Authors:  K R Chapman; F L Liu; R M Watson; A S Rebuck
Journal:  Chest       Date:  1986-04       Impact factor: 9.410

5.  Noninvasive detection of profound arterial desaturations using a pulse oximetry device.

Authors:  F G Mihm; B D Halperin
Journal:  Anesthesiology       Date:  1985-01       Impact factor: 7.892

6.  Evaluation of pulse oximetry.

Authors:  M Yelderman; W New
Journal:  Anesthesiology       Date:  1983-10       Impact factor: 7.892

7.  Noninvasive pulse oximetry in children with cyanotic congenital heart disease.

Authors:  R A Boxer; I Gottesfeld; S Singh; M A LaCorte; V A Parnell; P Walker
Journal:  Crit Care Med       Date:  1987-11       Impact factor: 7.598

8.  Accuracy of response of six pulse oximeters to profound hypoxia.

Authors:  J W Severinghaus; K H Naifeh
Journal:  Anesthesiology       Date:  1987-10       Impact factor: 7.892

9.  Evaluation of the Ohmeda 3700 pulse oximeter: steady-state and transient response characteristics.

Authors:  D M Kagle; C M Alexander; R S Berko; M Giuffre; J B Gross
Journal:  Anesthesiology       Date:  1987-03       Impact factor: 7.892

  9 in total
  27 in total

1.  Reflection and transmission pulse oximetry during compromised peripheral perfusion.

Authors:  H Pälve
Journal:  J Clin Monit       Date:  1992-01

2.  Oxygen saturation and breathing patterns in infancy. 1: Full term infants in the second month of life.

Authors:  V A Stebbens; C F Poets; J R Alexander; W A Arrowsmith; D P Southall
Journal:  Arch Dis Child       Date:  1991-05       Impact factor: 3.791

Review 3.  Noninvasive blood gas monitoring: a review for use in the adult critical care unit. Technology Subcommittee of the Working Group on Critical Care, Ontario Ministry of Health.

Authors: 
Journal:  CMAJ       Date:  1992-03-01       Impact factor: 8.262

4.  Validation of oxygen saturation monitoring in neonates.

Authors:  Shyang-Yun Pamela K Shiao; Ching-Nan Ou
Journal:  Am J Crit Care       Date:  2007-03       Impact factor: 2.228

5.  In vitro performance test system for pulse oximeters.

Authors:  K J Reynolds; J T Moyle; L B Gale; M K Sykes; C E Hahn
Journal:  Med Biol Eng Comput       Date:  1992-11       Impact factor: 2.602

6.  A new system to record reliable pulse oximetry data from the Nellcor N-200 and its applications in studies of variability in infant oxygenation.

Authors:  D Sprague; M S Richardson; J W Baish; J S Kemp
Journal:  J Clin Monit       Date:  1996-01

7.  Effect of exposure to 15% oxygen on breathing patterns and oxygen saturation in infants: interventional study.

Authors:  K J Parkins; C F Poets; L M O'Brien; V A Stebbens; D P Southall
Journal:  BMJ       Date:  1998-03-21

Review 8.  Continuous intra-arterial blood gas monitoring.

Authors:  B Venkatesh; S P Hendry
Journal:  Intensive Care Med       Date:  1996-08       Impact factor: 17.440

9.  The utility and safety of hypoxia experiences for rebreather divers.

Authors:  Simon J Mitchell; Hayden M Green; Stacey A Reading; Nicholas Gant
Journal:  Diving Hyperb Med       Date:  2019-06-30       Impact factor: 0.887

10.  Effect of anemia on pulse oximeter accuracy at low saturation.

Authors:  J W Severinghaus; S O Koh
Journal:  J Clin Monit       Date:  1990-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.