K J Reynolds1, J T Moyle, L B Gale, M K Sykes, C E Hahn.
Abstract
An in vitro system was developed capable of testing the accuracy and reproducibility of pulse oximeter readings. The pulse oximeter probe receives signals through a pulsating blood cuvette. The development of the design of the cuvette is described. Using the final design (or 'model finger'), a comparison is made between readings from a Datex Satlite pulse oximeter (SpO2) and saturation values obtained by use of a multi-wavelength bench oximeter (SaO2). Linear regression analysis of the data gives SpO2 = 0.88 SaO2 + 11.2 (r = 0.979, p < 0.001).Entities:
Mesh:
Year: 1992 PMID: 1297018 DOI: 10.1007/bf02446795