Accuracy of response of six pulse oximeters to profound hypoxia.

J W Severinghaus1, K H Naifeh.   

Abstract

Oxygen saturation, SpO2%, was recorded during rapidly induced 42.5 +/- 7.2-s plateaus of profound hypoxia at 40-70% saturation by 1 or 2 pulse oximeters from each of six manufacturers (NE = Nellcor N100, OH = Ohmeda 3700, NO = Novametrix 500 versions 2.2 and 3.3 (revised instrumentation), CR = Criticare CSI 501 + version .27 and version .28 in 501 & 502 (revised instrumentation), PC = PhysioControl Lifestat 1600, and MQ = Marquest/Minolta PulseOx 7). Usually, one probe of each pair was mounted on the ear, the other on a finger. Semi-recumbent, healthy, normotensive, non-smoking caucasian or asian volunteers (age range 18-64 yr) performed the test six to seven times each. After insertion of a radial artery catheter, subjects hyperventilated 3% CO2, 0-5% O2, balance N2. Saturation ScO2, computed on-line from mass spectrometer end-tidal PO2 and PCO2, was used to manually adjust FIO2 breath by breath to obtain a rapid fall to a hypoxic plateau lasting 30-45s, followed by rapid resaturation. Arterial HbO2% (Radiometer OSM-3) sampled near the end of the plateau averaged 55.5 +/- 7.5%. ScO2% (from the mass spectrometer) and SaO2% (from pH and PO2, by Corning 178) differed from HbO2% by + 0.2 +/- 3.6% and 0.4 +/- 2.8%, respectively. The mean and SD errors of pulse oximeters (vs. HbO2%) were: (table; see text) The plateaus were always long enough to permit instruments to demonstrate a plateau with ear probes, but finger probes sometimes failed to provide plateaus in subjects with peripheral vasoconstriction. Nonetheless, SpO2 read significantly too low with finger probes at 55% mean SaO2.(ABSTRACT TRUNCATED AT 250 WORDS)

Entities:  

Mesh:

Year:  1987        PMID: 3662082     DOI: 10.1097/00000542-198710000-00017

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  48 in total

1.  Non-invasive assessment of shunt and ventilation/perfusion ratio in neonates with pulmonary failure.

Authors:  H L Smith; J G Jones
Journal:  Arch Dis Child Fetal Neonatal Ed       Date:  2001-09       Impact factor: 5.747

2.  A comparison of pulse oximetry and near infrared spectroscopy (NIRS) in the detection of hypoxaemia occurring with pauses in nasal airflow in neonates.

Authors:  S L Watkin; S A Spencer; P W Dimmock; Y A Wickramasinghe; P Rolfe
Journal:  J Clin Monit Comput       Date:  1999-12       Impact factor: 2.502

3.  Reflection and transmission pulse oximetry during compromised peripheral perfusion.

Authors:  H Pälve
Journal:  J Clin Monit       Date:  1992-01

Review 4.  Pulse oximetry.

Authors:  J F Kelleher
Journal:  J Clin Monit       Date:  1989-01

5.  The influence of larger subcutaneous blood vessels on pulse oximetry.

Authors:  Paul D Mannheimer; Neil Michael P O'; Ewald Konecny
Journal:  J Clin Monit Comput       Date:  2004-06       Impact factor: 2.502

6.  Modelling the dynamic ventilatory response to hypoxia in humans.

Authors:  D S Ward; A Dahan; C B Mann
Journal:  Ann Biomed Eng       Date:  1992       Impact factor: 3.934

Review 7.  Pulse oximetry and capnography in intensive and transitional care units.

Authors:  F Bongard; D Sue
Journal:  West J Med       Date:  1992-01

8.  Effect of anemia on pulse oximeter accuracy at low saturation.

Authors:  J W Severinghaus; S O Koh
Journal:  J Clin Monit       Date:  1990-04

9.  A prototype device for standardized calibration of pulse oximeters.

Authors:  C Hornberger; P Knoop; W Nahm; H Matz; E Konecny; H Gehring; R Bonk; H Frankenberger; G Meyfroidt; P Wouters; J Gil-Rodriguez; L Ponz; K Benekos; J Valais; J Avgerinos; A Karoutis; A Ikiades; S Weininger
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

10.  Failure rate of pulse oximetry in the postanesthesia care unit.

Authors:  B S Gillies; K Posner; P Freund; F Cheney
Journal:  J Clin Monit       Date:  1993-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.