Evaluation of the Ohmeda 3700 pulse oximeter: steady-state and transient response characteristics.

D M Kagle, C M Alexander, R S Berko, M Giuffre, J B Gross.   

Abstract

The authors determined the accuracy of the Ohmeda 3700 (version J) pulse oximeter in healthy volunteers rendered hypoxic (SaO2 from 60-98%) by breathing mixtures of O2 in N2. When equipped with an ear probe, the pulse oximeter reading (y) reliably predicted arterial saturation (x) under steady-state conditions (y = 1.05x - 4.66, r = 0.98) as well as when oxygen saturation was rapidly decreasing (y = 1.05x - 6.38, r = 0.96). Conversely, when equipped with a finger probe, the oximeter tended to significantly underestimate steady-state arterial saturation (y = 1.21x - 19.1, r = 0.98, P less than 0.001). In response to this information, the manufacturer modified the oximeter's software (version XJ1), resulting in improved agreement between oximeter readings and arterial values (y = 0.96x + 4.59, r = 0.99). Despite the close correlation between steady-state oximeter readings and arterial saturation, the 99% prediction limits for both the ear and finger probes (version XJ1) were +/- 8%. Finger probe readings did not reliably reflect radial arterial oxygenation during rapid desaturation (y = 0.55x + 45.2, r = 0.78). This may be related to the time required to "arterialize" the blood in the finger; during acute resaturation, we found that the ear- to finger-probe delay was 24.0 +/- 2.3 s (means +/- SE, P less than 0.001).

Entities:  

Mesh:

Year:  1987        PMID: 3826695     DOI: 10.1097/00000542-198703000-00018

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  17 in total

Review 1.  Pulse oximetry.

Authors:  J F Kelleher
Journal:  J Clin Monit       Date:  1989-01

2.  Decreased circulation time in the upper limb reduces the lag time of the finger pulse oximeter response.

Authors:  Z N Ding; K Shibata; K Yamamoto; T Kobayashi; S Murakami
Journal:  Can J Anaesth       Date:  1992-01       Impact factor: 5.063

3.  Ohmeda pulse oximeters.

Authors:  J A Anderson
Journal:  Anesth Prog       Date:  1988-07

Review 4.  Pulse oximetry: analysis of theory, technology, and practice.

Authors:  M W Wukitsch; M T Petterson; D R Tobler; J A Pologe
Journal:  J Clin Monit       Date:  1988-10

Review 5.  Pulse oximetry for hypoxemia: a warning to users and manufacturers.

Authors:  S Fanconi
Journal:  Intensive Care Med       Date:  1989       Impact factor: 17.440

6.  Pulse oximetry: a "which" hunt?

Authors:  M K Sykes
Journal:  J Clin Monit       Date:  1989-04

7.  Evaluation of the Datascope ACCUSAT pulse oximeter in healthy adults.

Authors: 
Journal:  J Clin Monit       Date:  1989-01

8.  Epidural anesthesia affects pulse oximeter readings and response time.

Authors:  Shin Kawana; Hiroaki Watanabe; Akiyoshi Namiki; Naohiro Kokita
Journal:  J Anesth       Date:  1994-06       Impact factor: 2.078

9.  Errors in 14 pulse oximeters during profound hypoxia.

Authors:  J W Severinghaus; K H Naifeh; S O Koh
Journal:  J Clin Monit       Date:  1989-04

10.  The accuracy of the Biox 3700 pulse oximeter in patients receiving vasoactive therapy.

Authors:  J Ibáñez; J Velasco; J M Raurich
Journal:  Intensive Care Med       Date:  1991       Impact factor: 17.440

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.