Validation of oxygen saturation monitoring in neonates.

Shyang-Yun Pamela K Shiao1, Ching-Nan Ou.   

Abstract

BACKGROUND: Pulse oximetry is commonly used to monitor oxygenation in neonates, but cannot detect variations in hemoglobin. Venous and arterial oxygen saturations are rarely monitored. Few data are available to validate measurements of oxygen saturation in neonates (venous, arterial, or pulse oximetric). Purpose To validate oxygen saturation displayed on clinical monitors against analyses (with correction for fetal hemoglobin) of blood samples from neonates and to present the oxyhemoglobin dissociation curve for neonates.
METHOD: Seventy-eight neonates, 25 to 38 weeks' gestational age, had 660 arterial and 111 venous blood samples collected for analysis.
RESULTS: The mean difference between oxygen saturation and oxyhemoglobin level was 3% (SD 1.0) in arterial blood and 3% (SD 1.1) in venous blood. The mean difference between arterial oxygen saturation displayed on the monitor and oxyhemoglobin in arterial blood samples was 2% (SD 2.0); between venous oxygen saturation displayed on the monitor and oxyhemoglobin in venous blood samples it was 3% (SD 2.1) and between oxygen saturation as determined by pulse oximetry and oxyhemoglobin in arterial blood samples it was 2.5% (SD 3.1). At a Pao(2) of 50 to 75 mm Hg on the oxyhemoglobin dissociation curve, oxyhemoglobin in arterial blood samples was from 92% to 95%; oxygen saturation was from 95% to 98% in arterial blood samples, from 94% to 97% on the monitor, and from 95% to 97% according to pulse oximetry.
CONCLUSIONS: The safety limits for pulse oximeters are higher and narrower in neonates (95%-97%) than in adults, and clinical guidelines for neonates may require modification.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17322018      PMCID: PMC1931484     

Source DB:  PubMed          Journal:  Am J Crit Care        ISSN: 1062-3264            Impact factor:   2.228


  55 in total

1.  Effects of nine hemoglobin variants on five glycohemoglobin methods.

Authors:  W L Roberts; E L Frank; L Moulton; C Papadea; J K Noffsinger; C N Ou
Journal:  Clin Chem       Date:  2000-04       Impact factor: 8.327

2.  Assessing oxygenation in healthy infants.

Authors:  C F Poets
Journal:  J Pediatr       Date:  1999-11       Impact factor: 4.406

3.  Cellular genomic reporter assays for screening and evaluation of inducers of fetal hemoglobin.

Authors:  Jim Vadolas; Hady Wardan; Michael Orford; Robert Williamson; Panayiotis A Ioannou
Journal:  Hum Mol Genet       Date:  2003-11-25       Impact factor: 6.150

4.  Functional versus fractional oxygen saturation readings: bias and agreement using simulated solutions and adult blood.

Authors:  Shyang-Un Pamela Shiao
Journal:  Biol Res Nurs       Date:  2002-04       Impact factor: 2.522

5.  Neonatal pulse oximetry: accuracy and reliability.

Authors:  W W Hay; J M Brockway; M Eyzaguirre
Journal:  Pediatrics       Date:  1989-05       Impact factor: 7.124

6.  Evaluation of pulse oximetry as a continuous monitoring technique in the neonatal intensive care unit.

Authors:  K J Barrington; N N Finer; C A Ryan
Journal:  Crit Care Med       Date:  1988-11       Impact factor: 7.598

7.  Completing the hemodynamic picture: SvO2.

Authors:  K M White
Journal:  Heart Lung       Date:  1985-05       Impact factor: 2.210

8.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

9.  Evaluation of pulse oximetry.

Authors:  M Yelderman; W New
Journal:  Anesthesiology       Date:  1983-10       Impact factor: 7.892

10.  The effect of blood transfusion on the hemoglobin oxygen dissociation curve of very early preterm infants during the first week of life.

Authors:  Virginie De Halleux; Anita Truttmann; Carmen Gagnon; Harry Bard
Journal:  Semin Perinatol       Date:  2002-12       Impact factor: 3.300

View more
  7 in total

1.  A model-based decision support system for critiquing mechanical ventilation treatments.

Authors:  Fleur T Tehrani; Soraya Abbasi
Journal:  J Clin Monit Comput       Date:  2012-04-25       Impact factor: 2.502

2.  Influence of nose and mouth leaks on peripheral oxygen saturation during continuous positive airway pressure in neonates.

Authors:  Hendrik Stefan Fischer; Charles Christoph Roehr; Hans Proquitté; Gerd Schmalisch
Journal:  World J Pediatr       Date:  2013-10-21       Impact factor: 2.764

Review 3.  Accuracy of Pulse Oximetry in the Presence of Fetal Hemoglobin-A Systematic Review.

Authors:  Ena Pritišanac; Berndt Urlesberger; Bernhard Schwaberger; Gerhard Pichler
Journal:  Children (Basel)       Date:  2021-04-30

4.  Early hypoxemia burden is strongly associated with severe intracranial hemorrhage in preterm infants.

Authors:  Zachary A Vesoulis; Rachel L Bank; Doug Lake; Aaron Wallman-Stokes; Rakesh Sahni; J Randall Moorman; Joseph R Isler; Karen D Fairchild; Amit M Mathur
Journal:  J Perinatol       Date:  2018-09-28       Impact factor: 2.521

5.  Racial discrepancy in pulse oximeter accuracy in preterm infants.

Authors:  Zachary Vesoulis; Anna Tims; Hafsa Lodhi; Natasha Lalos; Halana Whitehead
Journal:  J Perinatol       Date:  2021-10-12       Impact factor: 3.225

6.  Monitoring Oxygenation and Gas Exchange in Neonatal Intensive Care Units: Current Practice in the Netherlands.

Authors:  Ratna N G B Tan; Estelle E M Mulder; Enrico Lopriore; Arjan B Te Pas
Journal:  Front Pediatr       Date:  2015-11-03       Impact factor: 3.418

Review 7.  Pulse oximetry: fundamentals and technology update.

Authors:  Meir Nitzan; Ayal Romem; Robert Koppel
Journal:  Med Devices (Auckl)       Date:  2014-07-08
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.