Comparison of optimised endovaginal vs external array coil T2-weighted and diffusion-weighted imaging techniques for detecting suspected early stage (IA/IB1) uterine cervical cancer.

Kate Downey1, Ayoma D Attygalle2, Veronica A Morgan1, Sharon L Giles1, A MacDonald1, M Davis3, Thomas E J Ind4, John H Shepherd4, Nandita M deSouza5.   

Abstract

OBJECTIVE: To compare sensitivity and specificity of endovaginal versus external-array coil T2-W and T2-W + DWI for detecting and staging small cervical tumours.
METHODS: Optimised endovaginal and external array coil MRI at 3.0-T was done prospectively in 48 consecutive patients with stage Ia/Ib1 cervical cancer. Sensitivity/specificity for detecting tumour and parametrial extension against histopathology for a reading radiologist were determined on coronal T2-W and T2W + DW images. An independent radiologist also scored T2-W images without and with addition of DWI for the external-array and endovaginal coils on separate occasions >2 weeks apart. Cohen's kappa assessed inter- and intra-observer agreement.
RESULTS: Median tumour volume in 19/38 cases positive on subsequent histology was 1.75 cm(3). Sensitivity, specificity, PPV, NPV were: reading radiologist 91.3 %, 89.5 %, 91.3 %, 89.5 %, respectively; independent radiologist T2-W 82.6 %, 73.7 %, 79.1 %, 77.8 % for endovaginal, 73.9 %, 89.5 %, 89.5 %, 73.9 % for external-array coil. Adding DWI improved sensitivity and specificity of endovaginal imaging (78.2 %, 89.5 %); adding DWI to external-array imaging improved specificity (94.7 %) but reduced sensitivity (66.7 %). Inter- and intra-observer agreement on T2-W + DWI was good (kappa = 0.67 and 0.62, respectively).
CONCLUSION: Endovaginal coil T2-W MRI is more sensitive than external-array coil for detecting tumours <2 cm(3); adding DWI improves specificity of endovaginal imaging but reduces sensitivity of external-array imaging. KEY POINTS: • Endovaginal more accurate than external-array T2-W MRI for detecting small cervical cancers. • Addition of DWI improves sensitivity and specificity of endovaginal T2-W imaging. • Addition of DWI substantially reduces sensitivity of external-array T2-W imaging.

Entities:  

Keywords:  Cervical cancer; Diffusion-weighted MRI; Endovaginal coil; External array coil; Fertility

Mesh:

Year:  2015        PMID: 26162579      PMCID: PMC4778155          DOI: 10.1007/s00330-015-3899-5

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


Introduction

Increasingly, because of effective cytological screening programmes, uterine cervical cancer presents at an early stage with small-volume disease. A ‘see-and-treat’ policy at most centres results in referrals for imaging following a diagnostic cone biopsy or large loop excision of the transformation zone (LLETZ) procedure, at which a significant proportion of the disease may be excised. The presence of residual disease is critical to inform further surgical management especially if considering preservation of fertility, so that its identification has become a key requirement from imaging. Cervical amputation (trachelectomy) is an option in these cases, where fertility is not compromised for other reasons [1] but requires at least a 1-cm length of normal cervical canal between the superior tumour margin and the internal os [2, 3]. Therefore, knowledge of residual tumour margins in relation to the normal cervical anatomy [4, 5] is paramount when considering patients for trachelectomy. The superior soft-tissue contrast provided by magnetic resonance imaging (MRI) makes it the imaging technique of choice for demonstrating cervical tumour in relation to normal cervical anatomy [6] in preoperative imaging [7]; standard image acquisition using external phased array receiver coils at field strengths of 1.5-T or 3.0-T results in an in-plane resolution of >1 mm with a 3- to 4-mm slice thickness. At this resolution, tumours of >1 cm have a sensitivity and sensitivity of detection around 60 %, although specificity is high [8, 9], but smaller tumours can be missed because spatial resolution is limited by the signal-to-noise ratio (SNR) available. The spatial resolution of MR images may be improved by the use of receiver coils close to the region of interest because they provide significant increases in SNR [10-12]. The use of an endovaginal receiver coil has previously been shown to improve the detection of tumours of <1.0 cm3 on T2-W imaging [13] and also allows detailed visualisation of the parametrium up to 6 cm from the surface of the internal coil [14]. When combined with diffusion-weighted (DW) contrast, the endovaginal technique further improves sensitivity and specificity of local staging to 92 % and 82 %, respectively [15, 16] because the apparent diffusion coefficient (ADC) of tumour is significantly lower than normal tissue [17]. However, a formal comparison of an optimised T2 + DW protocol for tumour detection using an endovaginal versus an external array-only technique has hitherto not been undertaken because of the time and resource implications of doing both fully optimised protocols in the same patients prior to surgery. The purpose of this work therefore was to undertake such a prospective study and compare the sensitivity and specificity of T2-W and T2-W + DWI obtained using an endovaginal coil with that using an external array coil for detecting small-volume cervical tumours and identifying early parametrial extension.

Methods

Endovaginal coil

A 37-mm ring-design solenoidal receiver coil similar to one previously described for use at 1.5-T [18] was used as part of a coil array with an externally placed flex-L coil (Philips Medical Systems, Best, The Netherlands). The endovaginal coil was designed and built in-house. It was covered in a disposable sheath and required simple manual insertion, without visual or imaging guidance, and manipulated so that the ring encompassed the cervix. The coil was immobilised in an external clamp as previously described. Vaginal air introduced during insertion was aspirated using a Ryles tube to minimise image distortion.

External array

A receive-only SENSE XL Torso coil 16-element phased array was used (Philips Medical Systems). This is a flexible volume coil consisting of anterior and posterior sections, each of eight elements – four upper and four lower. Coil dimensions were 74 mm × 558 mm × 600 mm for each half (upper and lower). Each element had a Z coverage of 22.5 cm.

Sequence optimisation

We initially established the most appropriate imaging protocols for both the endovaginal technique and the external array technique. As high a resolution as possible with the T2-W sequences without compromising image quality was deemed to be a minimum requirement. We matched the voxel size of the external array imaging to the endovaginal DW imaging (Table 1). From our experience in reporting these images, planes sagittal and coronal to the cervix were selected. The endovaginal images had a spatial resolution of 0.35 mm3 (0.42 × 0.42 × 2 mm), while the torso array had a spatial resolution of 3 mm3 acquired (1 × 1 × 3 mm). In order to minimise acquisition time, DWI for comparison was done in the coronal plane only as parametrial extension is best demonstrated in this plane. For the endovaginal technique we used the ZOOM sequence [19] which substantially reduces image distortions that arise due to susceptibility artefacts from the endovaginal coil. This allows a spatial resolution on the DW images of 3.1 mm3 (1.25 × 1.25 × 2 mm). Because of the lower image SNR without the internal coil, it was not possible to derive any meaningful data with the ZOOM-DWI. A conventional echo-planar imaging (EPI) sequence with a pixel resolution equivalent to that of the ZOOM-DWI image was therefore selected (1.25 × 1.25 mm) but a slice thickness of 3 mm was selected to match the slice thickness of the T2-W images and retain SNR, giving a voxel size of 4.7 mm3. Details of the endovaginal coil and external array coil protocols are given in Table 1. SNR measurements were made in a test-object with each coil combination using the optimised sequences and corresponding reconstruction filters, multichannel reconstruction and parallel imaging. A 2.5-cm diameter gel test object (Eurospin, T1 = 500 ms, T2 = 50 ms) surrounded by mineral oil test objects to mimic a body load was imaged with the endovaginal-flex-L coil combination followed by the external array coil, repeating each sequence three times. Mean signal intensity measured from a 200 mm2 region of interest on the central slice within the gel test object from the three repetitions was divided by the mean of the standard deviation of background noise. There was near equivalence of the SNR on T2-W images with each coil combination. There was a 62 % reduction in SNR on the b = 800 s/mm2 ZOOM-DWI images from the endovaginal coil compared to the corresponding T2-W images, but a seven times reduction in SNR on the b = 800 s/mm2 DWI from the external array coil compared to the endovaginal DWI images (Table 1).
Table 1

Sequence parameters used for imaging

ParameterT2-weighted coronal, sagittalDWI coronalT2-weightedDWI
Coil selectionDual: Endovaginal + FLEX-L 2 element phased array bodySingle: EndovaginalSENSE-XL torsoSENSE-XL torso
SequenceTurbo Spin Echo-Multi Shot (TSE factor 10)Spin-Echo Echo Planar Imaging-Single ShotTSE (TSE Factor 13)EPI
TR (ms)2750 (coronal)2500 (sagittal)650021003230
TE (ms)80549053
Flip angle90°90°90°90°
Bandwidth (Hz/Px)217.79.6217.39.6
FOV (mm)100 × 100100 × 100120 × 200110 × 100
Acq pixel size (mm)0.42 × 0.421.25 × 1.251.0 × 1.01.25 × 1.25
Acq matrix2388012088
Recon matrix288224400224
Acq voxel size (mm3)0.35 [0.42 × 0.42 × 2]3.125 [1.25 × 1.25 × 2]3.00 [1 × 1 × 3]4.69 [1.25 × 1.25 × 3]
Slice thickness (mm)2 (0.1 separation)2 (0.1 separation)33
Recon voxel size (mm3)0.24 [0.35 × 0.35 × 2]0.40 [0.45 × 0.45 × 2]0.75 [0.5 × 0.5 × 3]0.72 [0.49 × 0.49 × 3]
Number of slices24 (coronal)22 (sagittal)241616
NSA2142
Included b-values s/mm2 isotropicn/a0, 100, 300, 500, 800n/a0, 100, 300, 500, 800
Acquisition time4 min 18 s (coronal)4 min 0 s (sagittal)4 min 33 s3 min 58 s3 min 50 s
Signal in gel (T1 = 500 ms, T2 = 50 ms) to background noiseFlex-L elements 15 cm apart, gel sited centrally within endovaginal coil ring135Gel sited centrally within endovaginal coil ring51Coil elements 15 cm apart, gel sited centrally130Coil elements 15 cm apart, gel sited centrally7

DWI diffusion-weighted imaging, NSA number of signal averages

Sequence parameters used for imaging DWI diffusion-weighted imaging, NSA number of signal averages

Patients

The study was approved by the local Institutional Review Board. All patients gave informed, written consent. Over a 15-month period (May 2013–Aug 2014), a further follow-on cohort from our previous study [15, 20] of 51 patients were referred for MRI following an abnormal smear test and subsequent histological evidence of invasive cervical cancer. Twenty-two complained of at least one episode of abnormal bleeding unrelated to any preceding surgical procedure. All were successfully prospectively recruited for assessment with endovaginal T2-W MRI followed by high-resolution external phased array MRI at 3.0-T. Three patients were staged as Federation Internationale Gynecology and Obstetrics FIGO 2B at examination under anaesthesia, so that 48 consecutive patients with Stage 1 cervical cancer aged 24–53 years (mean 34.4 ± 8.2 years) were included in this study (Fig. 1). Forty-two patients had a cone or LLETZ biopsy at a median of 4.0 weeks (range 2–12 weeks) prior to MRI in which some or the entire tumour may have been excised; at histology, 26 were squamous cell carcinoma and 16 adenocarcinoma. The other five had punch biopsies only (three squamous cell carcinoma, two adenocarcinomas); of these three had undergone cone or LLETZ biopsies more than 9 months previously.
Fig. 1

Flow chart of patients included in the study and their clinical pathway. MRI magnetic resonance imaging, DW-MRI diffusion-weighted MRI, FIGO Federation Internationale Gynecology and Obstetrics, LLETZ Large loop excision of the transformation zone

Flow chart of patients included in the study and their clinical pathway. MRI magnetic resonance imaging, DW-MRI diffusion-weighted MRI, FIGO Federation Internationale Gynecology and Obstetrics, LLETZ Large loop excision of the transformation zone

Imaging methods

Images were obtained at 3.0-T (Philips Achieva). Hyoscine butyl bromide (Buscopan) 20 mg intramuscularly (IM) was administered to reduce artefact from bowel peristalsis. Endovaginal imaging was performed first. T2-W images were acquired in three orthogonal planes to the cervix and covered it entirely. DWI slice-matched images were acquired in free-breathing coronal to the cervix as parametrial extension is best assessed in this plane. Sequence parameters and acquisition time are given in Table 1. Isotropic apparent diffusion coefficient (ADC) maps based on the monoexponential rate of signal decay with increasing diffusion weighting were generated with the system software (Philips Medical Systems) using all b-values. Following the endovaginal imaging, the coil was removed and, following a 5-minute off-bed break, external array coil imaging was done. T2-W images were acquired in planes sagittal and coronal to the cervix. DW-slice-matched images were obtained coronal to the cervix. As with the endovaginally acquired data, isotropic ADC maps based on the monoexponential rate of signal decay with increasing diffusion weighting were generated with the system software (Philips Medical Systems) using all b-values.

Data analysis

At the time of clinical image reading, the reporting radiologist (20 years’ experience of pelvic MRI), with access to clinical information and all imaging data sets noted the absence or presence of tumour and the absence or presence of tumour invasion into the right and left parametrium. Where tumour was noted, a tumour volume was measured by drawing a region-of-interest around the tumour on every slice on which it appeared and multiplying the total measured tumour area by the slice thickness plus interslice gap. On completion of the patient accrual, an independent radiologist (5 years’ experience of pelvic MRI) without access to the clinical data separately scored first the anonymised coronal and sagittal T2-W external array images as positive or negative for tumour within the cervix; the absence or presence of parametrial extension was also noted as absent or present. This was then repeated with inclusion of the corresponding coronal ADC maps. At a subsequent occasion 2 weeks later to avoid memorisation, the same observer scored the randomized, anonymised coronal and sagittal T2-W endovaginal images without and then with the addition of the corresponding coronal ADC maps. The absence or presence of right and left parametrial extension was also noted.

Histopathological analysis

Tissue was acquired at radical hysterectomy (n = 10), radical vaginal trachelectomy (RVT) (n = 13), cold knife cone (n = 11) or LLETZ procedure (n = 4) in 38 patients. Of the remainder, four went to chemoradiotherapy and were known to have prior positive biopsies and six were deemed to have negative MRI scans with clear margins on the initial cone or LLETZ biopsy so no further intervention at that point was deemed necessary. These latter six cases were therefore excluded from the analysis because of a lack of post-MRI confirmation of the absence of tumour. All specimens were fixed in formalin. For hysterectomy and RVT specimens, a shave of the proximal endocervical resection margin (at or near the internal os) was sliced longitudinally (sagittal plane centrally, coronal plane laterally). The cervix together with the parametrium was sliced into transverse sections up to approximately 1 cm from the ectocervix. The remaining 1 cm of distal cervix together with vaginal cuff was sectioned longitudinally (sagittal plane centrally, coronal plane laterally or sagittally throughout). All slices (each 3–4 mm thick) were processed and embedded in paraffin; 2–3 micron sections were deparaffinised and stained with haematoxylin and eosin. According to the thickness of the slice obtained at macroscopy, each histology slide was separated from the next by 3–4 mm.

Statistical analysis

Sensitivity, specificity, and positive and negative predictive values for detecting the presence of tumour by the reading radiologist were obtained. For the independent radiologist blinded to the clinical information sensitivity, specificity, and positive and negative predictive values of T2-W without and with the addition of DWI were established for detecting the presence of tumour. Data were obtained for the endovaginal technique and separately for the external array-only technique. Similar data were derived for right and left parametrial extension. A Cohen’s kappa to assess agreement between observers for interpreting endovaginal T2-W + DW imaging together was calculated between reading and independent radiologists and between the internal and external coil readings of the independent radiologist on T2-W and separately T2-W + DW imaging.

Results

Thirty-eight patients (ten Stage Ia, 28 Stage 1b) underwent a surgical procedure a median of 27 days (lower quartile 23 days, upper quartile 35 days) after MRI; of these, 19 had residual tumour and 19 had no residual tumour on histology. Four patients who received chemoradiotherapy had tumour on biopsy at staging examination under anaesthesia (Fig. 1). Median tumour volume measured on endovaginal MRI of 23 true positive lesions (Fig. 2) was 1.75 cm3 (lower quartile 0.77 cm3, upper quartile 5.5 cm3).
Fig. 2

True positive T2-W and T2-W + diffusion-weighted imaging (DWI) on endovaginal images missed on external array in a 25-year-old female: Sagittal (a), axial (b), coronal (c) T2-W (FSE TR 2750 ms, TE 80 ms) and ADC map (EPI, TR 6500 ms, TE 54 ms, b = 0, 100, 300, 500, 800 mm2/s, d) with corresponding external array coil T2-W coronal (FSE TR 2100 ms, TE 90 ms, e) and ADC map (EPI TR 3230 ms, TE 53 ms, b = 0, 100, 300, 500, 800 mm2/s, f). The intermediate signal intensity tumour in a (arrow) shows restricted diffusion in d (arrow). The tumour is too small for visualisation in e and f. A 17-mm tumour was confirmed on right parasagittal histopathology (g, arrow)

True positive T2-W and T2-W + diffusion-weighted imaging (DWI) on endovaginal images missed on external array in a 25-year-old female: Sagittal (a), axial (b), coronal (c) T2-W (FSE TR 2750 ms, TE 80 ms) and ADC map (EPI, TR 6500 ms, TE 54 ms, b = 0, 100, 300, 500, 800 mm2/s, d) with corresponding external array coil T2-W coronal (FSE TR 2100 ms, TE 90 ms, e) and ADC map (EPI TR 3230 ms, TE 53 ms, b = 0, 100, 300, 500, 800 mm2/s, f). The intermediate signal intensity tumour in a (arrow) shows restricted diffusion in d (arrow). The tumour is too small for visualisation in e and f. A 17-mm tumour was confirmed on right parasagittal histopathology (g, arrow) Sensitivity and specificity for detecting tumour by the reporting radiologist were 91.3 % and 89.5 %, respectively. There were two false positives (Fig. 3) with tumour volumes estimated as 0.05 cm3 and 1.5 cm3, the latter described as a flat ectocervical plaque. The two false negative lesions (Fig. 4) were 6 mm in maximum dimension on histology. Sensitivity and specificity for detecting tumour by the independent radiologist blinded to the clinical information on endovaginal T2-W alone was 82.6 % and 73.7 %; inclusion of the DWI images showed a small reduction in sensitivity to 78.2 % but improved specificity to 89.5 %. In comparison, the external array T2-W images showed a lower sensitivity for detecting tumour of 73.9 % but a high specificity of 89.5 %. Addition of the DWI images improved specificity further to 94.7 % but at the expense of a substantial loss in sensitivity to 66.7 % (Table 2). The agreement between observers for interpreting endovaginal T2-W + DWI images together was good (Cohen’s kappa = 0.67). Agreement between the independent radiologist assessment of the T2-W images on internal and external techniques was moderate (kappa = 0.49), while the assessment of the T2-W + DW images on internal and external techniques was good (kappa = 0.62).
Fig. 3

False positive on endovaginal T2-W with negative diffusion-weighted imaging (DWI) in a 41-year-old female: Sagittal (a), axial (b), coronal (c) T2-W (FSE TR 2750 ms, TE 80 ms) and ADC map (EPI, TR 6500 ms, TE 54 ms, b = 0, 100, 300, 500, 800 mm2/s, d) with corresponding external array coil T2-W coronal (FSE TR 2100 ms, TE 90 ms, e) and ADC map (EPI TR 3230 ms, TE 53 ms, b = 0, 100, 300, 500, 800 mm2/s, f). The intermediate signal intensity nodule in a, b and c (arrows) did not show restricted diffusion in d (arrow). The tumour is too small for visualisation in e and f. Tumour was absent on histopathology

Fig. 4

False negative on endovaginal T2-W with negative diffusion-weighted imaging (DWI) in a 27-year-old female: Sagittal (a), axial (b), coronal (c) T2-W (FSE TR 2750 ms, TE 80 ms) and ADC map (EPI, TR 6500 ms, TE 54 ms, b = 0, 100, 300, 500, 800 mm2/s, d) with corresponding external array coil T2-W coronal (FSE TR 2100 ms, TE 90 ms, e) and ADC map (EPI TR 3230 ms, TE 53 ms, b = 0, 100, 300, 500, 800 mm2/s, f). There was no evidence of tumour on MRI with either technique, but a 6-mm microinvasive tumour (Stage Ia) was noted on histopathology (g, arrow)

Table 2

Sensitivity, specificity, and positive (PPV) and negative (NPV) predictive values for detecting tumour by a reporting radiologist and an independent radiologist. The comparison of endovaginal vs. external array techniques without and with the addition of diffusion-weighted (DWI) magnetic resonance imaging to the T2-W sequences for the independent radiologist is shown

N = 42Sensitivity%Specificity%PPV%NPV%
Reporting rad. T2W + DWI91.389.591.389.5
T2-WIndependent rad. endovaginal82.673.779.177.8
Independent rad. external array73.989.589.573.9
T2-W + DWIIndependent rad. endovaginal78.289.589.573.9
Independent rad. external array66.794.794.169.2

Rad. radiologist

False positive on endovaginal T2-W with negative diffusion-weighted imaging (DWI) in a 41-year-old female: Sagittal (a), axial (b), coronal (c) T2-W (FSE TR 2750 ms, TE 80 ms) and ADC map (EPI, TR 6500 ms, TE 54 ms, b = 0, 100, 300, 500, 800 mm2/s, d) with corresponding external array coil T2-W coronal (FSE TR 2100 ms, TE 90 ms, e) and ADC map (EPI TR 3230 ms, TE 53 ms, b = 0, 100, 300, 500, 800 mm2/s, f). The intermediate signal intensity nodule in a, b and c (arrows) did not show restricted diffusion in d (arrow). The tumour is too small for visualisation in e and f. Tumour was absent on histopathology False negative on endovaginal T2-W with negative diffusion-weighted imaging (DWI) in a 27-year-old female: Sagittal (a), axial (b), coronal (c) T2-W (FSE TR 2750 ms, TE 80 ms) and ADC map (EPI, TR 6500 ms, TE 54 ms, b = 0, 100, 300, 500, 800 mm2/s, d) with corresponding external array coil T2-W coronal (FSE TR 2100 ms, TE 90 ms, e) and ADC map (EPI TR 3230 ms, TE 53 ms, b = 0, 100, 300, 500, 800 mm2/s, f). There was no evidence of tumour on MRI with either technique, but a 6-mm microinvasive tumour (Stage Ia) was noted on histopathology (g, arrow) Sensitivity, specificity, and positive (PPV) and negative (NPV) predictive values for detecting tumour by a reporting radiologist and an independent radiologist. The comparison of endovaginal vs. external array techniques without and with the addition of diffusion-weighted (DWI) magnetic resonance imaging to the T2-W sequences for the independent radiologist is shown Rad. radiologist In 26 cases who went for trachelectomy or radical hysterectomy, one case was at a different institution and information on the parametrial status was not available. Therefore, 50 parametrial regions were evaluated in 25 cases, assuming that right and left parametrial extension are effectively independent. Tumour within the cervix had been noted on MRI in 17 of these cases by the reporting radiologist and in 17 by the independent radiologist. Tumour within a normal-sized parametrial node was present in one case only on histology. This was missed by the reporting radiologist who marked this parametrial region as negative. The sensitivity, specificity, PPV and NPV for the independent radiologist using the endovaginal compared to the external array technique, and without and with the DWI, are given in Table 3. However, as we had only one case with parametrial extension in our cohort, no firm conclusions can be drawn about the differences in parametrial assessment with and without an endovaginal coil.
Table 3

Sensitivity, specificity, and positive (PPV) and negative (NPV) predictive values for detecting parametrial extension by an independent radiologist. The comparison of endovaginal vs. external array techniques without and with the addition of diffusion-weighted magnetic resonance imaging (DW-MRI) to the T2-W sequences is shown

N = 50 regions (25 patients)Sens%Spec%PPV%NPV%
T2-WIndependent rad. endovaginal100100100100
Independent rad. external array1009533.3100
T2-W + DWIIndependent rad. endovaginal100100100100
Independent rad. external array10097.950100
Sensitivity, specificity, and positive (PPV) and negative (NPV) predictive values for detecting parametrial extension by an independent radiologist. The comparison of endovaginal vs. external array techniques without and with the addition of diffusion-weighted magnetic resonance imaging (DW-MRI) to the T2-W sequences is shown

Discussion

These data clearly demonstrate a higher sensitivity of T2-W endovaginal imaging for detecting small-volume tumours compared to an external array technique, albeit with lower specificity. Importantly, however, the addition of DWI to the endovaginal imaging serves to improve specificity so that it is equivalent to that of the external array T2-W assessment, but the addition of DWI to the external array technique substantially reduces sensitivity of detection of these small lesions because the spatial resolution of DWI with external array imaging is limited by low SNR. The higher sensitivity of detection of small tumours on endovaginal T2-W imaging alone compared to external array coil imaging is to be expected because of the improved spatial resolution with the increased SNR available with a small coil in situ, and the correlation of tumour size with histology has been demonstrated [20]. This has previously been shown not just in endovaginal imaging [18, 21] but also in prostatic imaging [22] where an endorectal technique is now standard in clinical practice. The improved SNR afforded by an endovaginal technique also means that sequences inherently low in SNR can be more usefully deployed. In prostatic imaging the equivalent endorectal imaging has led to the routine acquisition of multiparametric data sets which use a combination of diffusion-weighted dynamic contrast-enhanced [23, 24] and, where available, MR spectroscopic [25] techniques. Although these techniques can be acquired without endorectal imaging, this limits their sensitivity of detection to ‘clinically significant’ prostate cancer. Lesions too small for detection by non-endorectal multiparametric imaging are deemed ‘clinically insignificant’. This is not the case with cervical cancer, where even subcentimetre lesions are highly relevant and demand clinical intervention. The use of DWI in combination with endovaginal T2-W imaging therefore is of crucial importance in management to improve the specificity of detection of these lesions. With the external array coil alone, however, the SNR of the DWI meant that its addition to the external array T2-W imaging was detrimental in the detection of these small tumours and even provided false reassurance that a tumour was absent. Although improvements in external coil technology may well improve coil performance in future and a previous retrospective analysis of trachelectomy patients indicated a high sensitivity and specificity for the pre-operative external coil MRI [26], these improvements are unlikely to equate to the gains in SNR achievable with an internal receiver coil. The higher sensitivity and specificity of the reporting radiologist compared to the independent radiologist in our study may in part be due to increased experience, but is also likely to have been positively influenced by the availability of clinical information at the time of reading, so that the availability of this information is recommended during reading. A major consideration, however, is the image distortion introduced by the placement of an endovaginal probe. This is due to susceptibility in part from coil components and in part from residual air within the coil ring particularly in the region between the ring and the handle. In addition, residual air in the vagina introduced during coil placement further exacerbates the problem. The latter was minimised by sucking out residual air using a Ryles tube prior to imaging. This collapses the vaginal wall around the coil handle and minimises residual vaginal air, although it is not fool-proof and small pockets of air remain, particularly in the vaginal fornices. The coil construction uses MR compatible components with virtually no artefact on prior testing, so it is assumed that their contribution to susceptibility-induced distortions is minimal. Improved coil design and construction to minimise air around components should further reduce these problems. Another approach to improving geometric distortions on the DW images [27] is to replace the echo-planar based sequences with a turbo spin-echo (TSE). Previous attempts at such an approach have resulted in successful implementation of turbo spin-echo DWI for head and neck [28] and endometrial [29] cancers. The resulting images are more robust and suffer less from distortion, but incur a time penalty. Typically an increase of around 50 % in imaging time is expected. There is also an increased tendency to motion artefact and as these images are not single-shot like their echo-planar based counterpart, there is increased image blurring [30]. Optimisation of the DW technique for endovaginal MRI is therefore warranted; evaluation of the advantages and limitations of an EPI versus TSE-based sequences is required but was beyond the scope of the study, not least because of the length of a detailed endovaginal examination followed by a detailed external array examination. The number of b-values to use is also debated [31], but the choice of five b-values in this instance was based on the ability to calculate ADC more accurately, even though quantitative assessments did not form part of the current work. The scan schedule in this small group of patients was intensive and took ~60 min but fortunately would not be required in standard clinical care. It is our normal practise to do a detailed evaluation of the cervix using the endovaginal coil over 25 min and to then remove it and assess the pelvic and abdominal lymph node chains with an external array, which takes another 20 min. Although longer than a standard external array pelvic MRI of 30 min, this has the added value of detailed data prior to planning a fertility-sparing surgical procedure such as a trachelectomy in these young women. Appropriate referral for an endovaginal MRI is largely based on an approximate clinical or histological assessment of tumour volume. Larger tumours, particularly stage 2b and above, are adequately imaged using an external array coil technique [32, 33]. Tumours benefiting from this procedure are predominantly those clinically staged as Ia2 or Ib1, where the tumour may or may not be just palpable or the diagnostic or therapeutic cone biopsy or LLETZ indicates that there is incomplete excision of tumour or tumour extending to margins of the specimen. In these cases, endovaginal MRI is invaluable in depicting the presence of residual tumour and defining its extent in relation to the endocervical canal pre-operatively. As referrals to our study were largely based on these criteria, the tumour median volume in our cohort was 1.75 cm3. The increased detectability of these small tumours with endovaginal MRI is directly related to the improved spatial resolution possible because of the substantial improvement in SNR afforded by the internal coil. Although the numbers of patients are small, the sensitivity and specificity of our endovaginal reads by the reading radiologist are identical to those of previous cohorts studied with this technique [20].

Conclusion

Sensitivity for detecting tumour within the cervix was higher with an endovaginal coil technique than with a conventional external array coil, although specificity was lower. The addition of DWI to the endovaginal T2-W imaging gave a small reduction in sensitivity, but with a substantial improvement in specificity, making this the preferred diagnostic option. Although less sensitive, specificity for tumour detection was high with the external array coil using T2-W imaging; the addition of DWI in this case did not improve specificity, but markedly reduced sensitivity. Therefore DWI is only useful in early stage cervical cancer as an adjunct to T2-W imaging if an endovaginal technique is used. Endovaginal coil MRI would be recommended for assessment of clinical stage 1a and 1b1 cervical cancer.
  33 in total

Review 1.  Design and use of internal receiver coils for magnetic resonance imaging.

Authors:  D J Gilderdale; N M deSouza; G A Coutts; M K Chui; D J Larkman; A D Williams; I R Young
Journal:  Br J Radiol       Date:  1999-12       Impact factor: 3.039

2.  Practical estimate of gradient nonlinearity for implementation of apparent diffusion coefficient bias correction.

Authors:  Dariya I Malkyarenko; Thomas L Chenevert
Journal:  J Magn Reson Imaging       Date:  2014-12       Impact factor: 4.813

3.  Turboprop+: enhanced Turboprop diffusion-weighted imaging with a new phase correction.

Authors:  Chu-Yu Lee; Zhiqiang Li; James G Pipe; Josef P Debbins
Journal:  Magn Reson Med       Date:  2012-09-28       Impact factor: 4.668

4.  The value of ADC, T2 signal intensity, and a combination of both parameters to assess Gleason score and primary Gleason grades in patients with known prostate cancer.

Authors:  Johannes Nowak; Uwe Malzahn; Alexander D J Baur; Uta Reichelt; Tobias Franiel; Bernd Hamm; Tahir Durmus
Journal:  Acta Radiol       Date:  2014-12-12       Impact factor: 1.990

5.  Reduced-FOV excitation decreases susceptibility artifact in diffusion-weighted MRI with endorectal coil for prostate cancer detection.

Authors:  Natalie Korn; John Kurhanewicz; Suchandrima Banerjee; Olga Starobinets; Emine Saritas; Susan Noworolski
Journal:  Magn Reson Imaging       Date:  2014-09-06       Impact factor: 2.546

6.  Comparison of accuracy of intravoxel incoherent motion and apparent diffusion coefficient techniques for predicting malignancy of head and neck tumors using half-Fourier single-shot turbo spin-echo diffusion-weighted imaging.

Authors:  Junichiro Sakamoto; Akiko Imaizumi; Yoshinori Sasaki; Takashi Kamio; Mamoru Wakoh; Mika Otonari-Yamamoto; Tsukasa Sano
Journal:  Magn Reson Imaging       Date:  2014-05-12       Impact factor: 2.546

7.  Comparison of endorectal coil and nonendorectal coil T2W and diffusion-weighted MRI at 3 Tesla for localizing prostate cancer: correlation with whole-mount histopathology.

Authors:  Baris Turkbey; Maria J Merino; Elma Carvajal Gallardo; Vijay Shah; Omer Aras; Marcelino Bernardo; Esther Mena; Dagane Daar; Ardeshir R Rastinehad; W Marston Linehan; Bradford J Wood; Peter A Pinto; Peter L Choyke
Journal:  J Magn Reson Imaging       Date:  2013-11-15       Impact factor: 4.813

8.  Stage IB1 cervical cancer: role of preoperative MR imaging in selection of patients for fertility-sparing radical trachelectomy.

Authors:  Yulia Lakhman; Oguz Akin; Kay J Park; Debra M Sarasohn; Junting Zheng; Debra A Goldman; Michael J Sohn; Chaya S Moskowitz; Yukio Sonoda; Hedvig Hricak; Nadeem R Abu-Rustum
Journal:  Radiology       Date:  2013-06-20       Impact factor: 11.105

9.  Influencing surgical management in patients with carcinoma of the cervix using a T2- and ZOOM-diffusion-weighted endovaginal MRI technique.

Authors:  K Downey; M Jafar; A D Attygalle; S Hazell; V A Morgan; S L Giles; M A Schmidt; T E J Ind; J H Shepherd; N M deSouza
Journal:  Br J Cancer       Date:  2013-07-18       Impact factor: 7.640

10.  Preoperative imaging in patients undergoing trachelectomy for cervical cancer: validation of a combined T2- and diffusion-weighted endovaginal MRI technique at 3.0 T.

Authors:  Katherine Downey; John H Shepherd; Ayoma D Attygalle; Steve Hazell; Veronica A Morgan; Sharon L Giles; Thomas E J Ind; Nandita M Desouza
Journal:  Gynecol Oncol       Date:  2014-02-26       Impact factor: 5.482

View more
  12 in total

1.  Staging, recurrence and follow-up of uterine cervical cancer using MRI: Updated Guidelines of the European Society of Urogenital Radiology after revised FIGO staging 2018.

Authors:  Lucia Manganaro; Yulia Lakhman; Nishat Bharwani; Benedetta Gui; Silvia Gigli; Valeria Vinci; Stefania Rizzo; Aki Kido; Teresa Margarida Cunha; Evis Sala; Andrea Rockall; Rosemarie Forstner; Stephanie Nougaret
Journal:  Eur Radiol       Date:  2021-04-14       Impact factor: 5.315

Review 2.  Fertility-sparing for young patients with gynecologic cancer: How MRI can guide patient selection prior to conservative management.

Authors:  Sinead H McEvoy; Stephanie Nougaret; Nadeem R Abu-Rustum; Hebert Alberto Vargas; Elizabeth A Sadowski; Christine O Menias; Fuki Shitano; Shinya Fujii; Ramon E Sosa; Joanna G Escalon; Evis Sala; Yulia Lakhman
Journal:  Abdom Radiol (NY)       Date:  2017-10

Review 3.  Magnetic resonance imaging for detection of parametrial invasion in cervical cancer: An updated systematic review and meta-analysis of the literature between 2012 and 2016.

Authors:  Sungmin Woo; Chong Hyun Suh; Sang Youn Kim; Jeong Yeon Cho; Seung Hyup Kim
Journal:  Eur Radiol       Date:  2017-07-19       Impact factor: 5.315

Review 4.  Current imaging strategies for the evaluation of uterine cervical cancer.

Authors:  Charis Bourgioti; Konstantinos Chatoupis; Lia Angela Moulopoulos
Journal:  World J Radiol       Date:  2016-04-28

5.  A prospective comparative dosimetric study between diffusion weighted MRI (DWI) & T2-weighted MRI (T2W) for target delineation and planning in cervical cancer brachytherapy.

Authors:  Rishabh Kumar; Geeta S Narayanan; Bhaskar Vishwanthan; Sowmya Narayanan; Sanjeet Mandal
Journal:  Rep Pract Oncol Radiother       Date:  2020-10-28

6.  Cervical Carcinoma: Evaluation Using Diffusion MRI With a Fractional Order Calculus Model and its Correlation With Histopathologic Findings.

Authors:  Xian Shao; Li An; Hui Liu; Hui Feng; Liyun Zheng; Yongming Dai; Bin Yu; Jin Zhang
Journal:  Front Oncol       Date:  2022-04-05       Impact factor: 5.738

7.  An endovaginal MRI array with a forward-looking coil for advanced gynecological cancer brachytherapy procedures: Design and initial results.

Authors:  Akbar Alipour; Akila N Viswanathan; Ronald D Watkins; Hassan Elahi; Wolfgang Loew; Eric Meyer; Marc Morcos; Henry R Halperin; Ehud J Schmidt
Journal:  Med Phys       Date:  2021-10-08       Impact factor: 4.071

8.  High-resolution T2-weighted cervical cancer imaging: a feasibility study on ultra-high-field 7.0-T MRI with an endorectal monopole antenna.

Authors:  Jacob P Hoogendam; Irene M L Kalleveen; Catalina S Arteaga de Castro; Alexander J E Raaijmakers; René H M Verheijen; Maurice A A J van den Bosch; Dennis W J Klomp; Ronald P Zweemer; Wouter B Veldhuis
Journal:  Eur Radiol       Date:  2016-05-31       Impact factor: 5.315

9.  Separation of type and grade in cervical tumours using non-mono-exponential models of diffusion-weighted MRI.

Authors:  Jessica M Winfield; Matthew R Orton; David J Collins; Thomas E J Ind; Ayoma Attygalle; Steve Hazell; Veronica A Morgan; Nandita M deSouza
Journal:  Eur Radiol       Date:  2016-05-24       Impact factor: 5.315

10.  Value of diffusion-weighted imaging for monitoring tissue change during magnetic resonance-guided high-intensity focused ultrasound therapy in bone applications: an ex-vivo study.

Authors:  Sharon L Giles; Jessica M Winfield; David J Collins; Ian Rivens; John Civale; Gail R Ter Haar; Nandita M deSouza
Journal:  Eur Radiol Exp       Date:  2018-05-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.