Statistical model based iterative reconstruction (MBIR) in clinical CT systems. Part II. Experimental assessment of spatial resolution performance.

Ke Li1, John Garrett2, Yongshuai Ge2, Guang-Hong Chen1.   

Abstract

PURPOSE: Statistical model based iterative reconstruction (MBIR) methods have been introduced to clinical CT systems and are being used in some clinical diagnostic applications. The purpose of this paper is to experimentally assess the unique spatial resolution characteristics of this nonlinear reconstruction method and identify its potential impact on the detectabilities and the associated radiation dose levels for specific imaging tasks.
METHODS: The thoracic section of a pediatric phantom was repeatedly scanned 50 or 100 times using a 64-slice clinical CT scanner at four different dose levels [CTDIvol =4, 8, 12, 16 (mGy)]. Both filtered backprojection (FBP) and MBIR (Veo(®), GE Healthcare, Waukesha, WI) were used for image reconstruction and results were compared with one another. Eight test objects in the phantom with contrast levels ranging from 13 to 1710 HU were used to assess spatial resolution. The axial spatial resolution was quantified with the point spread function (PSF), while the z resolution was quantified with the slice sensitivity profile. Both were measured locally on the test objects and in the image domain. The dependence of spatial resolution on contrast and dose levels was studied. The study also features a systematic investigation of the potential trade-off between spatial resolution and locally defined noise and their joint impact on the overall image quality, which was quantified by the image domain-based channelized Hotelling observer (CHO) detectability index d'.
RESULTS: (1) The axial spatial resolution of MBIR depends on both radiation dose level and image contrast level, whereas it is supposedly independent of these two factors in FBP. The axial spatial resolution of MBIR always improved with an increasing radiation dose level and/or contrast level. (2) The axial spatial resolution of MBIR became equivalent to that of FBP at some transitional contrast level, above which MBIR demonstrated superior spatial resolution than FBP (and vice versa); the value of this transitional contrast highly depended on the dose level. (3) The PSFs of MBIR could be approximated as Gaussian functions with reasonably good accuracy. (4) Thez resolution of MBIR showed similar contrast and dose dependence. (5) Noise standard deviation assessed on the edges of objects demonstrated a trade-off with spatial resolution in MBIR. (5) When both spatial resolution and image noise were considered using the CHO analysis, MBIR led to significant improvement in the overall CT image quality for both high and low contrast detection tasks at both standard and low dose levels.
CONCLUSIONS: Due to the intrinsic nonlinearity of the MBIR method, many well-known CT spatial resolution and noise properties have been modified. In particular, dose dependence and contrast dependence have been introduced to the spatial resolution of CT images by MBIR. The method has also introduced some novel noise-resolution trade-off not seen in traditional CT images. While the benefits of MBIR regarding the overall image quality, as demonstrated in this work, are significant, the optimal use of this method in clinical practice demands a thorough understanding of its unique physical characteristics.

Entities:  

Mesh:

Year:  2014        PMID: 24989389      PMCID: PMC4106476          DOI: 10.1118/1.4884038

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  25 in total

1.  A three-dimensional statistical approach to improved image quality for multislice helical CT.

Authors:  Jean-Baptiste Thibault; Ken D Sauer; Charles A Bouman; Jiang Hsieh
Journal:  Med Phys       Date:  2007-11       Impact factor: 4.071

2.  Volumetric quantification of lung nodules in CT with iterative reconstruction (ASiR and MBIR).

Authors:  Baiyu Chen; Huiman Barnhart; Samuel Richard; Marthony Robins; James Colsher; Ehsan Samei
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

3.  Characterization of statistical prior image constrained compressed sensing. I. Applications to time-resolved contrast-enhanced CT.

Authors:  Pascal Theriault Lauzier; Guang-Hong Chen
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

4.  Characterization of statistical prior image constrained compressed sensing (PICCS): II. Application to dose reduction.

Authors:  Pascal Theriault Lauzier; Guang-Hong Chen
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

5.  Effects of ray profile modeling on resolution recovery in clinical CT.

Authors:  Christian Hofmann; Michael Knaup; Marc Kachelrieß
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

6.  Addition of a channel mechanism to the ideal-observer model.

Authors:  K J Myers; H H Barrett
Journal:  J Opt Soc Am A       Date:  1987-12       Impact factor: 2.129

7.  A simplified approach for modulation transfer function determinations in computed tomography.

Authors:  E L Nickoloff; R Riley
Journal:  Med Phys       Date:  1985 Jul-Aug       Impact factor: 4.071

8.  Human observer detection experiments with mammograms and power-law noise.

Authors:  A E Burgess; F L Jacobson; P F Judy
Journal:  Med Phys       Date:  2001-04       Impact factor: 4.071

9.  Spatial resolution characterization of differential phase contrast CT systems via modulation transfer function (MTF) measurements.

Authors:  Ke Li; Joseph Zambelli; Nicholas Bevins; Yongshuai Ge; Guang-Hong Chen
Journal:  Phys Med Biol       Date:  2013-05-17       Impact factor: 3.609

10.  Channelized hotelling and human observer correlation for lesion detection in hepatic SPECT imaging.

Authors:  H C Gifford; M A King; D J de Vries; E J Soares
Journal:  J Nucl Med       Date:  2000-03       Impact factor: 10.057

View more
  18 in total

1.  Technical Note: Measuring contrast- and noise-dependent spatial resolution of an iterative reconstruction method in CT using ensemble averaging.

Authors:  Lifeng Yu; Thomas J Vrieze; Shuai Leng; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

Review 2.  Regularization strategies in statistical image reconstruction of low-dose x-ray CT: A review.

Authors:  Hao Zhang; Jing Wang; Dong Zeng; Xi Tao; Jianhua Ma
Journal:  Med Phys       Date:  2018-09-10       Impact factor: 4.071

3.  Low contrast detectability and spatial resolution with model-based Iterative reconstructions of MDCT images: a phantom and cadaveric study.

Authors:  Domitille Millon; Alain Vlassenbroek; Aline G Van Maanen; Samantha E Cambier; Emmanuel E Coche
Journal:  Eur Radiol       Date:  2016-06-14       Impact factor: 5.315

4.  Hi-Res scan mode in clinical MDCT systems: Experimental assessment of spatial resolution performance.

Authors:  Juan P Cruz-Bastida; Daniel Gomez-Cardona; Ke Li; Heyi Sun; Jiang Hsieh; Timothy P Szczykutowicz; Guang-Hong Chen
Journal:  Med Phys       Date:  2016-05       Impact factor: 4.071

5.  Spectral prior image constrained compressed sensing (spectral PICCS) for photon-counting computed tomography.

Authors:  Zhicong Yu; Shuai Leng; Zhoubo Li; Cynthia H McCollough
Journal:  Phys Med Biol       Date:  2016-08-23       Impact factor: 3.609

6.  Correlation between a 2D channelized Hotelling observer and human observers in a low-contrast detection task with multislice reading in CT.

Authors:  Lifeng Yu; Baiyu Chen; James M Kofler; Christopher P Favazza; Shuai Leng; Matthew A Kupinski; Cynthia H McCollough
Journal:  Med Phys       Date:  2017-07-13       Impact factor: 4.071

7.  An effective noise reduction method for multi-energy CT images that exploit spatio-spectral features.

Authors:  Zhoubo Li; Shuai Leng; Lifeng Yu; Armando Manduca; Cynthia H McCollough
Journal:  Med Phys       Date:  2017-04-12       Impact factor: 4.071

8.  Statistical model based iterative reconstruction in clinical CT systems. Part III. Task-based kV/mAs optimization for radiation dose reduction.

Authors:  Ke Li; Daniel Gomez-Cardona; Jiang Hsieh; Meghan G Lubner; Perry J Pickhardt; Guang-Hong Chen
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

9.  Can conclusions drawn from phantom-based image noise assessments be generalized to in vivo studies for the nonlinear model-based iterative reconstruction method?

Authors:  Daniel Gomez-Cardona; Ke Li; Jiang Hsieh; Meghan G Lubner; Perry J Pickhardt; Guang-Hong Chen
Journal:  Med Phys       Date:  2016-02       Impact factor: 4.071

10.  Correlation between model observers in uniform background and human observers in patient liver background for a low-contrast detection task in CT.

Authors:  Hao Gong; Lifeng Yu; Shuai Leng; Samantha Dilger; Wei Zhou; Liqiang Ren; Cynthia H McCollough
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2018-03-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.