Noise-resolution tradeoffs in x-ray CT imaging: a comparison of penalized alternating minimization and filtered backprojection algorithms.

Joshua D Evans1, David G Politte, Bruce R Whiting, Joseph A O'Sullivan, Jeffrey F Williamson.   

Abstract

PURPOSE: In comparison with conventional filtered backprojection (FBP) algorithms for x-ray computed tomography (CT) image reconstruction, statistical algorithms directly incorporate the random nature of the data and do not assume CT data are linear, noiseless functions of the attenuation line integral. Thus, it has been hypothesized that statistical image reconstruction may support a more favorable tradeoff than FBP between image noise and spatial resolution in dose-limited applications. The purpose of this study is to evaluate the noise-resolution tradeoff for the alternating minimization (AM) algorithm regularized using a nonquadratic penalty function.
METHODS: Idealized monoenergetic CT projection data with Poisson noise were simulated for two phantoms with inserts of varying contrast (7%-238%) and distance from the field-of-view (FOV) center (2-6.5 cm). Images were reconstructed for the simulated projection data by the FBP algorithm and two penalty function parameter values of the penalized AM algorithm. Each algorithm was run with a range of smoothing strengths to allow quantification of the noise-resolution tradeoff curve. Image noise is quantified as the standard deviation in the water background around each contrast insert. Modulation transfer functions (MTFs) were calculated from six-parameter model fits to oversampled edge-spread functions defined by the circular contrast-insert edges as a metric of local resolution. The integral of the MTF up to 0.5 1p/mm was adopted as a single-parameter measure of local spatial resolution.
RESULTS: The penalized AM algorithm noise-resolution tradeoff curve was always more favorable than that of the FBP algorithm. While resolution and noise are found to vary as a function of distance from the FOV center differently for the two algorithms, the ratio of noises when matching the resolution metric is relatively uniform over the image. The ratio of AM-to-FBP image variances, a predictor of dose-reduction potential, was strongly dependent on the shape of the AM's nonquadratic penalty function and was also strongly influenced by the contrast of the insert for which resolution is quantified. Dose-reduction potential, reported here as the fraction (%) of FBP dose necessary for AM to reconstruct an image with comparable noise and resolution, for one penalty parameter value of the AM algorithm was found to vary from 70% to 50% for low-contrast and high-contrast structures, respectively, and from 70% to 10% for the second AM penalty parameter value. However, the second penalty, AM-700, was found to suffer from poor low-contrast resolution when matching the high-contrast resolution metric with FBP.
CONCLUSIONS: The results of this simulation study imply that penalized AM has the potential to reconstruct images with similar noise and resolution using a fraction (10%-70%) of the FBP dose. However, this dose-reduction potential depends strongly on the AM penalty parameter and the contrast magnitude of the structures of interest. In addition, the authors' results imply that the advantage of AM can be maximized by optimizing the nonquadratic penalty function to the specific imaging task of interest. Future work will extend the methods used here to quantify noise and resolution in images reconstructed from real CT data.

Entities:  

Mesh:

Year:  2011        PMID: 21520856      PMCID: PMC3060931          DOI: 10.1118/1.3549757

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  35 in total

1.  AAPM tutorial. CT image detail and noise.

Authors:  P Sprawls
Journal:  Radiographics       Date:  1992-09       Impact factor: 5.333

2.  Penalized-likelihood sinogram smoothing for low-dose CT.

Authors:  Patrick J La Rivière
Journal:  Med Phys       Date:  2005-06       Impact factor: 4.071

3.  Statistical reconstruction for x-ray computed tomography using energy-integrating detectors.

Authors:  Giovanni M Lasio; Bruce R Whiting; Jeffrey F Williamson
Journal:  Phys Med Biol       Date:  2007-04-02       Impact factor: 3.609

4.  Noise and resolution in images reconstructed with FBP and OSC algorithms for CT.

Authors:  A Ziegler; T Köhler; R Proksa
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

Review 5.  Computed tomography--an increasing source of radiation exposure.

Authors:  David J Brenner; Eric J Hall
Journal:  N Engl J Med       Date:  2007-11-29       Impact factor: 91.245

6.  Analysis of Resolution and Noise Properties of Nonquadratically Regularized Image Reconstruction Methods for PET.

Authors:  Sangtae Ahn; Richard M Leahy
Journal:  IEEE Trans Med Imaging       Date:  2008-03       Impact factor: 10.048

Review 7.  Innovations in CT dose reduction strategy: application of the adaptive statistical iterative reconstruction algorithm.

Authors:  Alvin C Silva; Holly J Lawder; Amy Hara; Jennifer Kujak; William Pavlicek
Journal:  AJR Am J Roentgenol       Date:  2010-01       Impact factor: 3.959

8.  The line spread function and modulation transfer function of a computed tomographic scanner.

Authors:  P F Judy
Journal:  Med Phys       Date:  1976 Jul-Aug       Impact factor: 4.071

9.  The effects of scatter in x-ray computed tomography.

Authors:  P M Joseph; R D Spital
Journal:  Med Phys       Date:  1982 Jul-Aug       Impact factor: 4.071

10.  Quadratic regularization design for 2-D CT.

Authors:  Hugo R Shi; Jeffrey A Fessler
Journal:  IEEE Trans Med Imaging       Date:  2008-10-31       Impact factor: 10.048

View more
  18 in total

1.  Image quality and dose optimisation for infant CT using a paediatric phantom.

Authors:  Jack W Lambert; Andrew S Phelps; Jesse L Courtier; Robert G Gould; John D MacKenzie
Journal:  Eur Radiol       Date:  2015-08-26       Impact factor: 5.315

2.  Prospects for in vivo estimation of photon linear attenuation coefficients using postprocessing dual-energy CT imaging on a commercial scanner: comparison of analytic and polyenergetic statistical reconstruction algorithms.

Authors:  Joshua D Evans; Bruce R Whiting; Joseph A O'Sullivan; David G Politte; Paul H Klahr; Yaduo Yu; Jeffrey F Williamson
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

3.  Technical Note: Measuring contrast- and noise-dependent spatial resolution of an iterative reconstruction method in CT using ensemble averaging.

Authors:  Lifeng Yu; Thomas J Vrieze; Shuai Leng; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2015-05       Impact factor: 4.071

4.  Statistical image reconstruction for low-dose CT using nonlocal means-based regularization. Part II: An adaptive approach.

Authors:  Hao Zhang; Jianhua Ma; Jing Wang; Yan Liu; Hao Han; Hongbing Lu; William Moore; Zhengrong Liang
Journal:  Comput Med Imaging Graph       Date:  2015-03-06       Impact factor: 4.790

5.  SR-NLM: a sinogram restoration induced non-local means image filtering for low-dose computed tomography.

Authors:  Zhaoying Bian; Jianhua Ma; Jing Huang; Hua Zhang; Shanzhou Niu; Qianjin Feng; Zhengrong Liang; Wufan Chen
Journal:  Comput Med Imaging Graph       Date:  2013-06-24       Impact factor: 4.790

6.  Soft-tissue imaging with C-arm cone-beam CT using statistical reconstruction.

Authors:  Adam S Wang; J Webster Stayman; Yoshito Otake; Gerhard Kleinszig; Sebastian Vogt; Gary L Gallia; A Jay Khanna; Jeffrey H Siewerdsen
Journal:  Phys Med Biol       Date:  2014-02-07       Impact factor: 3.609

7.  Task-based detectability in CT image reconstruction by filtered backprojection and penalized likelihood estimation.

Authors:  Grace J Gang; J Webster Stayman; Wojciech Zbijewski; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2014-08       Impact factor: 4.071

8.  Known-component 3D image reconstruction for improved intraoperative imaging in spine surgery: A clinical pilot study.

Authors:  Xiaoxuan Zhang; Ali Uneri; J Webster Stayman; Corinna C Zygourakis; Sheng-Fu L Lo; Nicholas Theodore; Jeffrey H Siewerdsen
Journal:  Med Phys       Date:  2019-06-30       Impact factor: 4.071

9.  dPIRPLE: a joint estimation framework for deformable registration and penalized-likelihood CT image reconstruction using prior images.

Authors:  H Dang; A S Wang; Marc S Sussman; J H Siewerdsen; J W Stayman
Journal:  Phys Med Biol       Date:  2014-08-06       Impact factor: 3.609

10.  Evaluation of 3D fluoroscopic image generation from a single planar treatment image on patient data with a modified XCAT phantom.

Authors:  Pankaj Mishra; Ruijiang Li; Sara St James; Raymond H Mak; Christopher L Williams; Yong Yue; Ross I Berbeco; John H Lewis
Journal:  Phys Med Biol       Date:  2013-01-21       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.