Metal artifact correction for x-ray computed tomography using kV and selective MV imaging.

Meng Wu1, Andreas Keil2, Dragos Constantin3, Josh Star-Lack3, Lei Zhu4, Rebecca Fahrig5.   

Abstract

PURPOSE: The overall goal of this work is to improve the computed tomography (CT) image quality for patients with metal implants or fillings by completing the missing kilovoltage (kV) projection data with selectively acquired megavoltage (MV) data that do not suffer from photon starvation. When both of these imaging systems, which are available on current radiotherapy devices, are used, metal streak artifacts are avoided, and the soft-tissue contrast is restored, even for regions in which the kV data cannot contribute any information.
METHODS: Three image-reconstruction methods, including two filtered back-projection (FBP)-based analytic methods and one iterative method, for combining kV and MV projection data from the two on-board imaging systems of a radiotherapy device are presented in this work. The analytic reconstruction methods modify the MV data based on the information in the projection or image domains and then patch the data onto the kV projections for a FBP reconstruction. In the iterative reconstruction, the authors used dual-energy (DE) penalized weighted least-squares (PWLS) methods to simultaneously combine the kV/MV data and perform the reconstruction.
RESULTS: The authors compared kV/MV reconstructions to kV-only reconstructions using a dental phantom with fillings and a hip-implant numerical phantom. Simulation results indicated that dual-energy sinogram patch FBP and the modified dual-energy PWLS method can successfully suppress metal streak artifacts and restore information lost due to photon starvation in the kV projections. The root-mean-square errors of soft-tissue patterns obtained using combined kV/MV data are 10-15 Hounsfield units smaller than those of the kV-only images, and the structural similarity index measure also indicates a 5%-10% improvement in the image quality. The added dose from the MV scan is much less than the dose from the kV scan if a high efficiency MV detector is assumed.
CONCLUSIONS: The authors have shown that it is possible to improve the image quality of kV CTs for patients with metal implants or fillings by completing the missing kV projection data with selectively acquired MV data that do not suffer from photon starvation. Numerical simulations demonstrated that dual-energy sinogram patch FBP and a modified kV/MV PWLS method can successfully suppress metal streak artifacts and restore information lost due to photon starvation in kV projections. Combined kV/MV images may permit the improved delineation of structures of interest in CT images for patients with metal implants or fillings.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25471970      PMCID: PMC4290750          DOI: 10.1118/1.4901551

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  55 in total

1.  Megavoltage cone-beam computed tomography using a high-efficiency image receptor.

Authors:  Ed J Seppi; Peter Munro; Stan W Johnsen; Ed G Shapiro; Carlo Tognina; Dan Jones; John M Pavkovich; Chris Webb; Ivan Mollov; Larry D Partain; Rick E Colbeth
Journal:  Int J Radiat Oncol Biol Phys       Date:  2003-03-01       Impact factor: 7.038

2.  Frequency split metal artifact reduction (FSMAR) in computed tomography.

Authors:  Esther Meyer; Rainer Raupach; Michael Lell; Bernhard Schmidt; Marc Kachelrieß
Journal:  Med Phys       Date:  2012-04       Impact factor: 4.071

3.  CT metal artifact reduction method correcting for beam hardening and missing projections.

Authors:  Joost M Verburg; Joao Seco
Journal:  Phys Med Biol       Date:  2012-04-18       Impact factor: 3.609

4.  Magnitude and clinical relevance of translational and rotational patient setup errors: a cone-beam CT study.

Authors:  Matthias Guckenberger; Juergen Meyer; Dirk Vordermark; Kurt Baier; Juergen Wilbert; Michael Flentje
Journal:  Int J Radiat Oncol Biol Phys       Date:  2006-07-01       Impact factor: 7.038

5.  Statistical reconstruction for x-ray computed tomography using energy-integrating detectors.

Authors:  Giovanni M Lasio; Bruce R Whiting; Jeffrey F Williamson
Journal:  Phys Med Biol       Date:  2007-04-02       Impact factor: 3.609

6.  A novel forward projection-based metal artifact reduction method for flat-detector computed tomography.

Authors:  Daniel Prell; Yiannis Kyriakou; Marcel Beister; Willi A Kalender
Journal:  Phys Med Biol       Date:  2009-10-14       Impact factor: 3.609

7.  Prior-based artifact correction (PBAC) in computed tomography.

Authors:  Thorsten Heußer; Marcus Brehm; Ludwig Ritschl; Stefan Sawall; Marc Kachelrieß
Journal:  Med Phys       Date:  2014-02       Impact factor: 4.071

8.  Assessment of residual error for online cone-beam CT-guided treatment of prostate cancer patients.

Authors:  Daniel Létourneau; Alvaro A Martinez; David Lockman; Di Yan; Carlos Vargas; Giovanni Ivaldi; John Wong
Journal:  Int J Radiat Oncol Biol Phys       Date:  2005-07-15       Impact factor: 7.038

Review 9.  Megavoltage cone-beam CT: system description and clinical applications.

Authors:  Olivier Morin; Amy Gillis; Josephine Chen; Michèle Aubin; M Kara Bucci; Mack Roach; Jean Pouliot
Journal:  Med Dosim       Date:  2006       Impact factor: 1.482

10.  Streaking artifacts reduction in four-dimensional cone-beam computed tomography.

Authors:  Shuai Leng; Joseph Zambelli; Ranjini Tolakanahalli; Brian Nett; Peter Munro; Joshua Star-Lack; Bhudatt Paliwal; Guang-Hong Chen
Journal:  Med Phys       Date:  2008-10       Impact factor: 4.071

View more
  8 in total

1.  Synthetic CT generation from CBCT images via deep learning.

Authors:  Liyuan Chen; Xiao Liang; Chenyang Shen; Steve Jiang; Jing Wang
Journal:  Med Phys       Date:  2020-01-13       Impact factor: 4.071

2.  High-kVp Assisted Metal Artifact Reduction for X-ray Computed Tomography.

Authors:  Yan Xi; Yannan Jin; Bruno De Man; Ge Wang
Journal:  IEEE Access       Date:  2016-09-16       Impact factor: 3.367

3.  Quantification of metal artifacts in computed tomography: methodological considerations.

Authors:  Nils Große Hokamp; Brendan Eck; Florian Siedek; Daniel Pinto Dos Santos; Jasmin A Holz; David Maintz; Stefan Haneder
Journal:  Quant Imaging Med Surg       Date:  2020-05

4.  Concurrent kilovoltage CBCT imaging and megavoltage beam delivery: suppression of cross-scatter with 2D antiscatter grids and grid-based scatter sampling.

Authors:  Farhang Bayat; Mohamed Elsayed Eldib; Brian Kavanagh; Moyed Miften; Cem Altunbas
Journal:  Phys Med Biol       Date:  2022-08-09       Impact factor: 4.174

5.  A piecewise-focused high DQE detector for MV imaging.

Authors:  Josh Star-Lack; Daniel Shedlock; Dennis Swahn; Dave Humber; Adam Wang; Hayley Hirsh; George Zentai; Daren Sawkey; Isaac Kruger; Mingshan Sun; Eric Abel; Gary Virshup; Mihye Shin; Rebecca Fahrig
Journal:  Med Phys       Date:  2015-09       Impact factor: 4.071

6.  Multitask 3D CBCT-to-CT translation and organs-at-risk segmentation using physics-based data augmentation.

Authors:  Navdeep Dahiya; Sadegh R Alam; Pengpeng Zhang; Si-Yuan Zhang; Tianfang Li; Anthony Yezzi; Saad Nadeem
Journal:  Med Phys       Date:  2021-08-09       Impact factor: 4.506

7.  Clinical translation of a new flat-panel detector for beam's-eye-view imaging.

Authors:  T C Harris; J Seco; D Ferguson; M Lehmann; P Huber; M Shi; M Jacobson; I Valencia Lozano; M Myronakis; P Baturin; R Fueglistaller; D Morf; R Berbeco
Journal:  Phys Med Biol       Date:  2020-12-07       Impact factor: 4.174

8.  Metal artifact reduction through MVCBCT and kVCT in radiotherapy.

Authors:  Gao Liugang; Sun Hongfei; Ni Xinye; Fang Mingming; Cao Zheng; Lin Tao
Journal:  Sci Rep       Date:  2016-11-21       Impact factor: 4.379

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.