Assessment of residual error for online cone-beam CT-guided treatment of prostate cancer patients.

Daniel Létourneau1, Alvaro A Martinez, David Lockman, Di Yan, Carlos Vargas, Giovanni Ivaldi, John Wong.   

Abstract

PURPOSE: Kilovoltage cone-beam CT (CBCT) implemented on board a medical accelerator is available for image-guidance applications in our clinic. The objective of this work was to assess the magnitude and stability of the residual setup error associated with CBCT online-guided prostate cancer patient setup. Residual error pertains to the uncertainty in image registration, the limited mechanical accuracy, and the intrafraction motion during imaging and treatment. METHODS AND MATERIALS: The residual error for CBCT online-guided correction was first determined in a phantom study. After online correction, the phantom residual error was determined by comparing megavoltage portal images acquired every 90 degrees to the corresponding digitally reconstructed radiographs. In the clinical study, 8 prostate cancer patients were implanted with three radiopaque markers made of high-winding coils. After positioning the patient using the skin marks, a CBCT scan was acquired and the setup error determined by fusing the coils on the CBCT and planning CT scans. The patient setup was then corrected by moving the couch accordingly. A second CBCT scan was acquired immediately after the correction to evaluate the residual target setup error. Intrafraction motion was evaluated by tracking the coils and the bony landmarks on kilovoltage radiographs acquired every 30 s between the two CBCT scans. Corrections based on soft-tissue registration were evaluated offline by aligning the prostate contours defined on both planning CT and CBCT images.
RESULTS: For ideal rigid phantoms, CBCT image-guided treatment can usually achieve setup accuracy of 1 mm or better. For the patients, after CBCT correction, the target setup error was reduced in almost all cases and was generally within +/-1.5 mm. The image guidance process took 23-35 min, dictated by the computer speed and network configuration. The contribution of the intrafraction motion to the residual setup error was small, with a standard deviation of +/-0.9 mm. The average difference between the setup corrections obtained with coil and soft-tissue registration was greatest in the superoinferior direction and was equal to -1.1 +/- 2.9 mm.
CONCLUSION: On the basis of the residual setup error measurements, the margin required after online CBCT correction for the patients enrolled in this study would be approximatively 3 mm and is considered to be a lower limit owing to the small intrafraction motion observed. The discrepancy between setup corrections derived from registration using coils or soft tissue can be due in part to the lack of complete three-dimensional information with the coils or to the difficulty in prostate delineation and requires further study.

Entities:  

Mesh:

Year:  2005        PMID: 15913917     DOI: 10.1016/j.ijrobp.2005.03.035

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  37 in total

1.  Effect of daily setup errors on individual dose distribution in conventional radiotherapy: an initial study.

Authors:  Akihiro Takemura; Saori Shoji; Sinichi Ueda; Yuichi Kurata; Tomoyasu Kumano; Shigeyuki Takamatsu; Masayuki Suzuki
Journal:  Radiol Phys Technol       Date:  2009-05-01

2.  Prostate intrafraction motion evaluation using kV fluoroscopy during treatment delivery: a feasibility and accuracy study.

Authors:  Justus Adamson; Qiuwen Wu
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

Review 3.  kV cone-beam CT-based IGRT: a clinical review.

Authors:  Judit Boda-Heggemann; Frank Lohr; Frederik Wenz; Michael Flentje; Matthias Guckenberger
Journal:  Strahlenther Onkol       Date:  2011-04-26       Impact factor: 3.621

4.  Multiresolution reconstruction for cone-beam tomography from raw data projections using 3D ridgelets.

Authors:  Francisco Gómez; Cristina Santa Marta; Eduardo Romero
Journal:  J Digit Imaging       Date:  2011-12       Impact factor: 4.056

5.  Metal artifact correction for x-ray computed tomography using kV and selective MV imaging.

Authors:  Meng Wu; Andreas Keil; Dragos Constantin; Josh Star-Lack; Lei Zhu; Rebecca Fahrig
Journal:  Med Phys       Date:  2014-12       Impact factor: 4.071

6.  Comparison of prostate positioning guided by three-dimensional transperineal ultrasound and cone beam CT.

Authors:  Minglun Li; Hendrik Ballhausen; Nina-Sophie Hegemann; Michael Reiner; Stefan Tritschler; Christian Gratzke; Farkhad Manapov; Stefanie Corradini; Ute Ganswindt; Claus Belka
Journal:  Strahlenther Onkol       Date:  2016-12-07       Impact factor: 3.621

Review 7.  Adaptive radiation therapy for prostate cancer.

Authors:  Michel Ghilezan; Di Yan; Alvaro Martinez
Journal:  Semin Radiat Oncol       Date:  2010-04       Impact factor: 5.934

8.  Planning target volume margins for prostate radiotherapy using daily electronic portal imaging and implanted fiducial markers.

Authors:  David Skarsgard; Pat Cadman; Ali El-Gayed; Robert Pearcey; Patricia Tai; Nadeem Pervez; Jackson Wu
Journal:  Radiat Oncol       Date:  2010-06-10       Impact factor: 3.481

9.  Image quality and stability of image-guided radiotherapy (IGRT) devices: A comparative study.

Authors:  Markus Stock; Marlies Pasler; Wolfgang Birkfellner; Peter Homolka; Richard Poetter; Dietmar Georg
Journal:  Radiother Oncol       Date:  2009-08-18       Impact factor: 6.280

10.  Dose perturbations from implanted helical gold markers in proton therapy of prostate cancer.

Authors:  Annelise Giebeler; Jonas Fontenot; Peter Balter; George Ciangaru; Ronald Zhu; Wayne Newhauser
Journal:  J Appl Clin Med Phys       Date:  2009-01-27       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.