Dimensionality and noise in energy selective x-ray imaging.

Robert E Alvarez1.   

Abstract

PURPOSE: To develop and test a method to quantify the effect of dimensionality on the noise in energy selective x-ray imaging.
METHODS: The Cramèr-Rao lower bound (CRLB), a universal lower limit of the covariance of any unbiased estimator, is used to quantify the noise. It is shown that increasing dimensionality always increases, or at best leaves the same, the variance. An analytic formula for the increase in variance in an energy selective x-ray system is derived. The formula is used to gain insight into the dependence of the increase in variance on the properties of the additional basis functions, the measurement noise covariance, and the source spectrum. The formula is also used with computer simulations to quantify the dependence of the additional variance on these factors. Simulated images of an object with three materials are used to demonstrate the trade-off of increased information with dimensionality and noise. The images are computed from energy selective data with a maximum likelihood estimator.
RESULTS: The increase in variance depends most importantly on the dimension and on the properties of the additional basis functions. With the attenuation coefficients of cortical bone, soft tissue, and adipose tissue as the basis functions, the increase in variance of the bone component from two to three dimensions is 1.4 × 10(3). With the soft tissue component, it is 2.7 × 10(4). If the attenuation coefficient of a high atomic number contrast agent is used as the third basis function, there is only a slight increase in the variance from two to three basis functions, 1.03 and 7.4 for the bone and soft tissue components, respectively. The changes in spectrum shape with beam hardening also have a substantial effect. They increase the variance by a factor of approximately 200 for the bone component and 220 for the soft tissue component as the soft tissue object thickness increases from 1 to 30 cm. Decreasing the energy resolution of the detectors increases the variance of the bone component markedly with three dimension processing, approximately a factor of 25 as the resolution decreases from 100 to 3 bins. The increase with two dimension processing for adipose tissue is a factor of two and with the contrast agent as the third material for two or three dimensions is also a factor of two for both components. The simulated images show that a maximum likelihood estimator can be used to process energy selective x-ray data to produce images with noise close to the CRLB.
CONCLUSIONS: The method presented can be used to compute the effects of the object attenuation coefficients and the x-ray system properties on the relationship of dimensionality and noise in energy selective x-ray imaging systems.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 24320442      PMCID: PMC3808483          DOI: 10.1118/1.4824057

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  14 in total

1.  Systematic bias in basis material decomposition applied to quantitative dual-energy x-ray imaging.

Authors:  E L Gingold; B H Hasegawa
Journal:  Med Phys       Date:  1992 Jan-Feb       Impact factor: 4.071

2.  K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors.

Authors:  E Roessl; R Proksa
Journal:  Phys Med Biol       Date:  2007-07-17       Impact factor: 3.609

3.  Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography.

Authors:  J P Schlomka; E Roessl; R Dorscheid; S Dill; G Martens; T Istel; C Bäumer; C Herrmann; R Steadman; G Zeitler; A Livne; R Proksa
Journal:  Phys Med Biol       Date:  2008-07-08       Impact factor: 3.609

4.  Energy-selective reconstructions in X-ray computerized tomography.

Authors:  R E Alvarez; A Macovski
Journal:  Phys Med Biol       Date:  1976-09       Impact factor: 3.609

Review 5.  Detectors for the future of X-ray imaging.

Authors:  M Aslund; E Fredenberg; M Telman; M Danielsson
Journal:  Radiat Prot Dosimetry       Date:  2010-03-05       Impact factor: 0.972

6.  Near optimal energy selective x-ray imaging system performance with simple detectors.

Authors:  Robert E Alvarez
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

7.  Estimator for photon counting energy selective x-ray imaging with multibin pulse height analysis.

Authors:  Robert E Alvarez
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

8.  An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV.

Authors:  J M Boone; J A Seibert
Journal:  Med Phys       Date:  1997-11       Impact factor: 4.071

9.  Modeling the performance of a photon counting x-ray detector for CT: energy response and pulse pileup effects.

Authors:  Katsuyuki Taguchi; Mengxi Zhang; Eric C Frey; Xiaolan Wang; Jan S Iwanczyk; Einar Nygard; Neal E Hartsough; Benjamin M W Tsui; William C Barber
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

10.  Photon Counting Energy Dispersive Detector Arrays for X-ray Imaging.

Authors:  Jan S Iwanczyk; Einar Nygård; Oded Meirav; Jerry Arenson; William C Barber; Neal E Hartsough; Nail Malakhov; Jan C Wessel
Journal:  IEEE Trans Nucl Sci       Date:  2009       Impact factor: 1.679

View more
  13 in total

1.  Material elemental decomposition in dual and multi-energy CT via a sparsity-dictionary approach for proton stopping power ratio calculation.

Authors:  Chenyang Shen; Bin Li; Liyuan Chen; Ming Yang; Yifei Lou; Xun Jia
Journal:  Med Phys       Date:  2018-02-23       Impact factor: 4.071

2.  Signal to noise ratio of energy selective x-ray photon counting systems with pileup.

Authors:  Robert E Alvarez
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

3.  Dual-source multienergy CT with triple or quadruple x-ray beams.

Authors:  Lifeng Yu; Liqiang Ren; Zhoubo Li; Shuai Leng; Cynthia H McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2018-07-24

4.  Noise suppression for dual-energy CT via penalized weighted least-square optimization with similarity-based regularization.

Authors:  Joseph Harms; Tonghe Wang; Michael Petrongolo; Tianye Niu; Lei Zhu
Journal:  Med Phys       Date:  2016-05       Impact factor: 4.071

5.  Spectral prior image constrained compressed sensing (spectral PICCS) for photon-counting computed tomography.

Authors:  Zhicong Yu; Shuai Leng; Zhoubo Li; Cynthia H McCollough
Journal:  Phys Med Biol       Date:  2016-08-23       Impact factor: 3.609

6.  Assessment of vectorial total variation penalties on realistic dual-energy CT data.

Authors:  David S Rigie; Adrian A Sanchez; Patrick J La Rivière
Journal:  Phys Med Biol       Date:  2017-04-21       Impact factor: 3.609

7.  Noise Suppression for Dual-Energy CT Through Entropy Minimization.

Authors:  Michael Petrongolo; Lei Zhu
Journal:  IEEE Trans Med Imaging       Date:  2015-05-01       Impact factor: 10.048

8.  Simultaneous dual-contrast imaging using energy-integrating detector multi-energy CT: An in vivo feasibility study.

Authors:  Zhongxing Zhou; Liqiang Ren; Kishore Rajendran; Felix E Diehn; Joel G Fletcher; Cynthia H McCollough; Lifeng Yu
Journal:  Med Phys       Date:  2022-01-27       Impact factor: 4.071

9.  Spectral Photon Counting CT: Imaging Algorithms and Performance Assessment.

Authors:  Adam S Wang; Norbert J Pelc
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2020-07-07

10.  Human Imaging With Photon Counting-Based Computed Tomography at Clinical Dose Levels: Contrast-to-Noise Ratio and Cadaver Studies.

Authors:  Ralf Gutjahr; Ahmed F Halaweish; Zhicong Yu; Shuai Leng; Lifeng Yu; Zhoubo Li; Steven M Jorgensen; Erik L Ritman; Steffen Kappler; Cynthia H McCollough
Journal:  Invest Radiol       Date:  2016-07       Impact factor: 6.016

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.