Estimator for photon counting energy selective x-ray imaging with multibin pulse height analysis.

Robert E Alvarez1.   

Abstract

PURPOSE: This paper describes a noniterative estimator for the energy dependent information from photon counting detectors with multibin pulse height analysis (PHA). The estimator uses the two function decomposition of the attenuation coefficient [R. E. Alvarez and A. Macovski, Phys. Med. Biol. 21, 733-744 (1976)] and its output is the line integrals of the basis set coefficients. The output noise variance and bias is compared to other noniterative estimators and to the Cramèr-Rao lower bound (CRLB).
METHODS: The estimator first computes an initial estimate from a linearized maximum likelihood estimator. The errors in the initial estimates are determined at a set of points from measurements on a calibration phantom. The errors at these known points are interpolated to create two-dimensional look up tables of corrections to the initial estimates. During image acquisition, the linearized maximum likelihood estimate for each data point is used as an input to the correction look up tables, and the final output is the sum of the estimate and the correction. The performance of the estimator is compared to generalizations of the polynomial and rational polynomial estimators for multibin data. The estimators are compared by the mean square error (MSE) and its components, the bias, and the variance of the output. The variance is also compared to the CRLB. The performance is simulated with two to five bins PHA data. The CRLB at a fixed object thickness is also computed as a function of the number of bins.
RESULTS: For two bin data, all the estimators' variances are equal to the CRLB. With three or more bins, only the proposed estimator achieves the CRLB while the others, which were not optimized for noise performance, have much larger output variance. The bias of the proposed estimator is equal to the polynomial estimator for calibration phantoms with 40 or more steps, that is, 1600 combinations of basis materials, but is larger than the rational polynomial bias. In all cases at the photon counts tested, the MSE is essentially equal to the variance, indicating that the bias errors are negligible compared to the variance.
CONCLUSIONS: The estimator provides a noniterative method to compute the energy dependent information from multibin PHA data that achieves the CRLB over a wide range of operating conditions and has low output bias. The estimator can be calibrated based on the measurements of a calibration phantom; so, it does not require measurements of the x-ray energy spectrum or the detector response functions.

Mesh:

Year:  2011        PMID: 21776766      PMCID: PMC3098892          DOI: 10.1118/1.3570658

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors.

Authors:  E Roessl; R Proksa
Journal:  Phys Med Biol       Date:  2007-07-17       Impact factor: 3.609

2.  Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography.

Authors:  J P Schlomka; E Roessl; R Dorscheid; S Dill; G Martens; T Istel; C Bäumer; C Herrmann; R Steadman; G Zeitler; A Livne; R Proksa
Journal:  Phys Med Biol       Date:  2008-07-08       Impact factor: 3.609

3.  An experimental study on the noise properties of x-ray CT sinogram data in Radon space.

Authors:  Jing Wang; Hongbing Lu; Zhengrong Liang; Daria Eremina; Guangxiang Zhang; Su Wang; John Chen; James Manzione
Journal:  Phys Med Biol       Date:  2008-06-03       Impact factor: 3.609

4.  An accurate method for direct dual-energy calibration and decomposition.

Authors:  H N Cardinal; A Fenster
Journal:  Med Phys       Date:  1990 May-Jun       Impact factor: 4.071

5.  Empirical dual energy calibration (EDEC) for cone-beam computed tomography.

Authors:  Philip Stenner; Timo Berkus; Marc Kachelriess
Journal:  Med Phys       Date:  2007-09       Impact factor: 4.071

6.  Energy-selective reconstructions in X-ray computerized tomography.

Authors:  R E Alvarez; A Macovski
Journal:  Phys Med Biol       Date:  1976-09       Impact factor: 3.609

Review 7.  Detectors for the future of X-ray imaging.

Authors:  M Aslund; E Fredenberg; M Telman; M Danielsson
Journal:  Radiat Prot Dosimetry       Date:  2010-03-05       Impact factor: 0.972

8.  Near optimal energy selective x-ray imaging system performance with simple detectors.

Authors:  Robert E Alvarez
Journal:  Med Phys       Date:  2010-02       Impact factor: 4.071

9.  An accurate method for computer-generating tungsten anode x-ray spectra from 30 to 140 kV.

Authors:  J M Boone; J A Seibert
Journal:  Med Phys       Date:  1997-11       Impact factor: 4.071

10.  A fast dual-energy computational method using isotransmission lines and table lookup.

Authors:  K S Chuang; H K Huang
Journal:  Med Phys       Date:  1987 Mar-Apr       Impact factor: 4.071

  10 in total
  18 in total

1.  Spectral response model for a multibin photon-counting spectral computed tomography detector and its applications.

Authors:  Xuejin Liu; Mats Persson; Hans Bornefalk; Staffan Karlsson; Cheng Xu; Mats Danielsson; Ben Huber
Journal:  J Med Imaging (Bellingham)       Date:  2015-09-11

2.  Dimensionality and noise in energy selective x-ray imaging.

Authors:  Robert E Alvarez
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

3.  The effects of extending the spectral information acquired by a photon-counting detector for spectral CT.

Authors:  Taly Gilat Schmidt; Kevin C Zimmerman; Emil Y Sidky
Journal:  Phys Med Biol       Date:  2015-01-23       Impact factor: 3.609

4.  Segmented targeted least squares estimator for material decomposition in multibin photon-counting detectors.

Authors:  Paurakh L Rajbhandary; Scott S Hsieh; Norbert J Pelc
Journal:  J Med Imaging (Bellingham)       Date:  2017-05-18

5.  Experimental comparison of empirical material decomposition methods for spectral CT.

Authors:  Kevin C Zimmerman; Taly Gilat Schmidt
Journal:  Phys Med Biol       Date:  2015-03-27       Impact factor: 3.609

Review 6.  Energy-sensitive photon counting detector-based X-ray computed tomography.

Authors:  Katsuyuki Taguchi
Journal:  Radiol Phys Technol       Date:  2017-01-30

7.  Dual source hybrid spectral micro-CT using an energy-integrating and a photon-counting detector.

Authors:  M D Holbrook; D P Clark; C T Badea
Journal:  Phys Med Biol       Date:  2020-10-21       Impact factor: 3.609

8.  Spectral prior image constrained compressed sensing (spectral PICCS) for photon-counting computed tomography.

Authors:  Zhicong Yu; Shuai Leng; Zhoubo Li; Cynthia H McCollough
Journal:  Phys Med Biol       Date:  2016-08-23       Impact factor: 3.609

9.  A neural network-based method for spectral distortion correction in photon counting x-ray CT.

Authors:  Mengheng Touch; Darin P Clark; William Barber; Cristian T Badea
Journal:  Phys Med Biol       Date:  2016-07-29       Impact factor: 3.609

10.  Image-based Material Decomposition with a General Volume Constraint for Photon-Counting CT.

Authors:  Zhoubo Li; Shuai Leng; Lifeng Yu; Zhicong Yu; Cynthia H McCollough
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2015
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.