Near optimal energy selective x-ray imaging system performance with simple detectors.

Robert E Alvarez1.   

Abstract

PURPOSE: This article describes a method to achieve near optimal performance with low energy resolution detectors. Tapiovaara and Wagner [Phys. Med. Biol. 30, 519-529 (1985)] showed that an energy selective x-ray system using a broad spectrum source can produce images with a larger signal to noise ratio (SNR) than conventional systems using energy integrating or photon counting detectors. They showed that there is an upper limit to the SNR and that it can be achieved by measuring full spectrum information and then using an optimal energy dependent weighting.
METHODS: A performance measure is derived by applying statistical detection theory to an abstract vector space of the line integrals of the basis set coefficients of the two function approximation to the x-ray attenuation coefficient. The approach produces optimal results that utilize all the available energy dependent data. The method can be used with any energy selective detector and is applied not only to detectors using pulse height analysis (PHA) but also to a detector that simultaneously measures the total photon number and integrated energy, as discussed by Roessl et al. [Med. Phys. 34, 959-966 (2007)]. A generalization of this detector that improves the performance is introduced. A method is described to compute images with the optimal SNR using projections in a "whitened" vector space transformed so the noise is uncorrelated and has unit variance in both coordinates. Material canceled images with optimal SNR can also be computed by projections in this space.
RESULTS: The performance measure is validated by showing that it provides the Tapiovaara-Wagner optimal results for a detector with full energy information and also a conventional detector. The performance with different types of detectors is compared to the ideal SNR as a function of x-ray tube voltage and subject thickness. A detector that combines two bin PHA with a simultaneous measurement of integrated photon energy provides near ideal performance across a wide range of operating conditions.
CONCLUSIONS: Low energy resolution detectors can be used in energy selective x-ray imaging systems to produce images with near optimal performance.

Mesh:

Year:  2010        PMID: 20229892      PMCID: PMC2826386          DOI: 10.1118/1.3284538

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  8 in total

1.  Detective quantum efficiency dependence on x-ray energy weighting in mammography.

Authors:  R N Cahn; B Cederström; M Danielsson; A Hall; M Lundqvist; D Nygren
Journal:  Med Phys       Date:  1999-12       Impact factor: 4.071

2.  Properties of preprocessed sinogram data in x-ray computed tomography.

Authors:  Bruce R Whiting; Parinaz Massoumzadeh; Orville A Earl; Joseph A O'Sullivan; Donald L Snyder; Jeffrey F Williamson
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

3.  On the influence of noise correlations in measurement data on basis image noise in dual-energylike x-ray imaging.

Authors:  Ewald Roessl; Andy Ziegler; Roland Proksa
Journal:  Med Phys       Date:  2007-03       Impact factor: 4.071

4.  K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors.

Authors:  E Roessl; R Proksa
Journal:  Phys Med Biol       Date:  2007-07-17       Impact factor: 3.609

5.  An experimental study on the noise properties of x-ray CT sinogram data in Radon space.

Authors:  Jing Wang; Hongbing Lu; Zhengrong Liang; Daria Eremina; Guangxiang Zhang; Su Wang; John Chen; James Manzione
Journal:  Phys Med Biol       Date:  2008-06-03       Impact factor: 3.609

6.  Optimal "image-based" weighting for energy-resolved CT.

Authors:  Taly Gilat Schmidt
Journal:  Med Phys       Date:  2009-07       Impact factor: 4.071

7.  Energy-selective reconstructions in X-ray computerized tomography.

Authors:  R E Alvarez; A Macovski
Journal:  Phys Med Biol       Date:  1976-09       Impact factor: 3.609

8.  Generalized image combinations in dual KVP digital radiography.

Authors:  L A Lehmann; R E Alvarez; A Macovski; W R Brody; N J Pelc; S J Riederer; A L Hall
Journal:  Med Phys       Date:  1981 Sep-Oct       Impact factor: 4.071

  8 in total
  9 in total

1.  Optimization of beam quality for photon-counting spectral computed tomography in head imaging: simulation study.

Authors:  Han Chen; Cheng Xu; Mats Persson; Mats Danielsson
Journal:  J Med Imaging (Bellingham)       Date:  2015-11-06

2.  Dimensionality and noise in energy selective x-ray imaging.

Authors:  Robert E Alvarez
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

3.  Estimator for photon counting energy selective x-ray imaging with multibin pulse height analysis.

Authors:  Robert E Alvarez
Journal:  Med Phys       Date:  2011-05       Impact factor: 4.071

4.  Signal to noise ratio of energy selective x-ray photon counting systems with pileup.

Authors:  Robert E Alvarez
Journal:  Med Phys       Date:  2014-11       Impact factor: 4.071

5.  Estimation of signal and noise for a whole-body research photon-counting CT system.

Authors:  Zhoubo Li; Shuai Leng; Zhicong Yu; Steffen Kappler; Cynthia H McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2017-06-22

6.  A framework for performance characterization of energy-resolving photon-counting detectors.

Authors:  Mats Persson; Paurakh L Rajbhandary; Norbert J Pelc
Journal:  Med Phys       Date:  2018-10-12       Impact factor: 4.071

7.  Determination of linear x-ray attenuation coefficients of pathological brain tissues and use of filters in tissue contrast enhancement in computed tomography.

Authors:  M Erdem Sagsoz; Fazlı Erdogan; Salih Z Erzeneoglu; İhsan Yuce
Journal:  Eurasian J Med       Date:  2010-08

Review 8.  An introduction to photon-counting detector CT (PCD CT) for radiologists.

Authors:  Yuko Nakamura; Toru Higaki; Shota Kondo; Ikuo Kawashita; Isao Takahashi; Kazuo Awai
Journal:  Jpn J Radiol       Date:  2022-10-18       Impact factor: 2.701

9.  Optimizing spectral CT parameters for material classification tasks.

Authors:  D S Rigie; P J La Rivière
Journal:  Phys Med Biol       Date:  2016-05-26       Impact factor: 3.609

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.