M Aslund1, E Fredenberg, M Telman, M Danielsson.
Abstract
In recent decades, developments in detectors for X-ray imaging have improved dose efficiency. This has been accomplished with for example, structured scintillators such as columnar CsI, or with direct detectors where the X rays are converted to electric charge carriers in a semiconductor. Scattered radiation remains a major noise source, and fairly inefficient anti-scatter grids are still a gold standard. Hence, any future development should include improved scatter rejection. In recent years, photon-counting detectors have generated significant interest by several companies as well as academic research groups. This method eliminates electronic noise, which is an advantage in low-dose applications. Moreover, energy-sensitive photon-counting detectors allow for further improvements by optimising the signal-to-quantum-noise ratio, anatomical background subtraction or quantitative analysis of object constituents. This paper reviews state-of-the-art photon-counting detectors, scatter control and their application in diagnostic X-ray medical imaging. In particular, spectral imaging with photon-counting detectors, pitfalls such as charge sharing and high rates and various proposals for mitigation are discussed.Entities:
Mesh:
Year: 2010 PMID: 20207749 DOI: 10.1093/rpd/ncq074
Source DB: PubMed Journal: Radiat Prot Dosimetry ISSN: 0144-8420 Impact factor: 0.972