| Literature DB >> 24274022 |
Haya Matuoq Al Farhan1, Wafa'a Majed Al Otaibi, Hanouf Mohammed Al Razqan, Alanoud Abdullah Al Harqan.
Abstract
BACKGROUND: The aim was to compare the repeatability, reproducibility and inherent precision of ultrasound pachymetry (USP), noncontact specular microscopy (SP-2000P) and the Confoscan 4 confocal microscope (z-ring CS4) in measuring endothelial cell density (ECD), coefficient of variation of cell size (CV), and central corneal thickness (CCT) in normal eyes.Entities:
Mesh:
Year: 2013 PMID: 24274022 PMCID: PMC4222699 DOI: 10.1186/1471-2415-13-73
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
The means and standard deviations for ECD measurements obtained using the SP-2000P and z-ring CS4, intra-observer repeatability, inter-observer reproducibility (z-ring CS4 only), and comparison between SP-2000P vs. z-ring CS4 (mean differences, limits of agreement, paired -test, un-paired test, and p-values)
| Mean ± SD* | 3115.50 ± 279.70 | 3167.50 ± 264.75 | 3192.63 ± 249.42 |
| Intra-observer repeatability (p-value) | 0.22 | 0.99 | 0.13 |
| | Mean difference ± SD* | Limits of agreement | p-value |
| Inter-observer reproducibility with z-ring CS4** | 25.10 ± 128.92 | +304, -354.20 | 0.30 |
| Comparison between SP-2000P vs. z-ring CS4† | -52.10 ± 97.90 | +139.80, -243.90 | 0.46 |
*Standard deviation.
**Paired test.
†Unpaired test.
The means and standard deviations for CV measurements obtained using the SP-2000P and z-ring CS4, intra-observer repeatability, inter-observer reproducibility (z-ring CS4 only), and comparison between SP-2000P vs. z-ring CS4 (mean differences, limits of agreement, paired -test, un-paired test, and p-values)
| Mean ± SD* | 27.12 ± 2.51 | 27.10 ± 2.41 | 27.17 ± 2.25 |
| Intra-observer repeatability (p-value) | 0.75 | 0.34 | 0.98 |
| | Mean difference ± SD* | Limits of agreement | p-value |
| Inter-observer reproducibility with z-ring CS4** | -0.20 ± 1.20 | +3.80, -4.20 | 0.30 |
| Comparison between SP-2000P vs. z-ring CS4† | 0.20 ± 1.45 | +3.00, -2.70 | 0.79 |
*Standard deviation.
**Paired test.
†Un-paired test.
The means and standard deviations for CCT measurements obtained using the SP-2000P and z-ring CS4, intra-observer repeatability, inter-observer reproducibility (z-ring CS4 only), and comparison between SP-2000P vs. z-ring CS4 (mean differences, limits of agreement, paired -test, un-paired test, and p-values)
| Mean ± SD* | 555.11 ± 35.83 | 535.82 ± 41.10 | 552.57 ± 36.83 | 554.97 ± 36.34 |
| Intra-observer repeatability (p-value) | 0.91 | 0.20 | 0.45 | 0.54 |
| | Mean difference ± SD* | Limits of agreement | p-value | |
| Inter-observer reproducibility with z-ring CS4** | -2.40 ± 9.10 | +39.90, -38.70 | 0.16 | |
| USP vs. SP-2000P | 19.30 ± 19.60 | +57.70, -19.20 | | |
| USP vs. z-ring CS4 | 2.50 ± 5.70 | +16.30, -11.20 | ||
| SP-2000P vs. z-ring CS4 | -16.70 ± 18.60 | +23.70, -57.20 | ||
Standard deviation*.
Paired test**.
Overview of recent studies comparing the repeatability and reproducibility of different instruments for measuring CCT, ECD and CV
| | SP-2000P vs. IMAGEnet® | 1.) 8 eyes to evaluate SP-2000P performance | 20–29 | Some subjects were contact lens wearers | No* | |
| 2.) 7 eyes to evaluate reproducibility | ||||||
| 3.) 12 eyes to evaluate repeatability | ||||||
| | SP1000P vs. CS3 | 50 eyes | 21–42 | Neonates and contact lens wearers | No* | |
| USP vs. SP-2000P | | 45 eyes**/62 eyes† | 13–52 | Mild myopia and keratoconics | No* | |
| USP vs. z-ring CS4 | 44 eyes | 22–49 | 22 eyes were post refractive surgery | No* | ||
*In all studies the variation between instruments was statistically significant.
**Mild myopia (minus 1–6 dioptres).
†Patients with keratoconus.