Corneal thickness measurements with the Topcon SP-2000P specular microscope and an ultrasound pachymeter.

R Bovelle1, S C Kaufman, H W Thompson, H Hamano.   

Abstract

OBJECTIVE: To compare the reproducibility of measurements obtained with a new pachymetry instrument, the Topcon specular microscope (Topcon SP-2000P; Topcon America Corp, Paramus, NJ), with those obtained by ultrasound pachymetry.
METHODS: Corneal thickness was measured in 40 eyes of 40 patients 3 times each with the Topcon SP-2000P and an ultrasound pachymeter (DGH 500, DGH Technology Inc, Exton, Pa) by 2 separate investigators. Comparisons included average thickness as measured by each instrument, average thickness for each instrument as measured by each investigator, and differences in thickness due to corneal abnormalities.
RESULTS: Mean corneal thickness measured by the Topcon instrument was significantly less (32 microm; P<.001) than the mean value obtained with the ultrasound pachymeter. Similarly, mean values obtained with the 2 instruments by the 2 investigators were significantly different (P<.001 and .008 for investigators 1 and 2, respectively), with the Topcon value less than the ultrasound value in both cases. Both instruments detected abnormalities in corneal thickness equally well. However, the measurements obtained with the Topcon instrument by the 2 investigators were more consistent (no significant difference [P=.32]) than those obtained with the ultrasound unit (difference was significant [P=.02]).
CONCLUSIONS: The new noncontact Topcon specular microscope provides measurements of corneal thickness that are somewhat less than those of ultrasound pachymetry, but that seem to be more consistent from one operator to another, possibly as a result of the elimination of observer bias induced by probe placement required by the ultrasound unit. This consistency may be important in the comparison of measurements by different operators over time in patients being followed up after refractive surgery or other therapeutic interventions.

Entities:  

Mesh:

Year:  1999        PMID: 10408449     DOI: 10.1001/archopht.117.7.868

Source DB:  PubMed          Journal:  Arch Ophthalmol        ISSN: 0003-9950


  27 in total

1.  Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis.

Authors:  K Kawana; T Tokunaga; K Miyata; F Okamoto; T Kiuchi; T Oshika
Journal:  Br J Ophthalmol       Date:  2004-04       Impact factor: 4.638

2.  Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry.

Authors:  Hani S Al-Mezaine; Saleh A Al-Amro; Dustan Kangave; Abdulkareem Sadaawy; Taher A Wehaib; Saleh Al-Obeidan
Journal:  Int Ophthalmol       Date:  2007-09-26       Impact factor: 2.031

3.  Measurement of central corneal thickness in health and disease.

Authors:  Imtiaz A Chaudhry
Journal:  Saudi J Ophthalmol       Date:  2009-10

4.  Accutome PachPen handheld ultrasonic pachymeter: intraobserver repeatability and interobserver reproducibility by personnel of different training grades.

Authors:  Mohammadreza Peyman; Lai Yong Tai; Keat Ween Khaw; Choung Min Ng; Maung Maung Win; Visvaraja Subrayan
Journal:  Int Ophthalmol       Date:  2014-09-05       Impact factor: 2.031

5.  Comparison of central corneal thickness and endothelial cell measurements by Scheimpflug camera system and two noncontact specular microscopes.

Authors:  Irmak Karaca; Suzan Guven Yilmaz; Melis Palamar; Halil Ates
Journal:  Int Ophthalmol       Date:  2017-07-03       Impact factor: 2.031

6.  Comparison of specular microscopy and ultrasound pachymetry before and after cataract surgery.

Authors:  Alberto López-Miguel; María Sanchidrián; Itziar Fernández; Alfredo Holgueras; Miguel J Maldonado
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-11-22       Impact factor: 3.117

7.  Comparison of Central Corneal Thickness using Anterior Segment Optical Coherence Tomography Versus Ultrasound Pachymetry.

Authors:  Prasanna Venkatesh Ramesh; Kirti Nath Jha; Krishnagopal Srikanth
Journal:  J Clin Diagn Res       Date:  2017-08-01

8.  Repeatability of central corneal thickness measurements measured with the Topcon SP2000P specular microscope.

Authors:  Kelechi C Ogbuehi; Turki M Almubrad
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-03-15       Impact factor: 3.117

9.  Central corneal thickness in Japanese children.

Authors:  Akiko Hikoya; Miho Sato; Kinnichi Tsuzuki; Yuka Maruyama Koide; Ryo Asaoka; Yoshihiro Hotta
Journal:  Jpn J Ophthalmol       Date:  2009-01-30       Impact factor: 2.447

10.  Comparison of central corneal thickness measurements by Pentacam, noncontact specular microscope, and ultrasound pachymetry in normal and post-LASIK eyes.

Authors:  Saleh Al-Ageel; Abdulrahman M Al-Muammar
Journal:  Saudi J Ophthalmol       Date:  2009-10-24
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.