Sercan Akyalcin1, Jeryl D English, Kenneth M Abramovitch, Xiujiang J Rong.
Abstract
OBJECTIVE: To measure surface skin dose from various cone-beam computed tomography (CBCT) scanners using point-dosimeters. MATERIALS &Entities:
Mesh:
Year: 2013 PMID: 24192155 PMCID: PMC3851731 DOI: 10.1186/1746-160X-9-28
Source DB: PubMed Journal: Head Face Med ISSN: 1746-160X Impact factor: 2.151
Scan settings of the CBCT and conventional x-ray imaging units used in the study
| Kodak 9500 | Medium | 86 | 108 | 10.8 sec | 15X9 | 0.2 |
| Kodak 9500 | Large | 120 | 108 | 10.8 sec | 20X18 | 0.3 |
| iCAT Next Generation | Medium | 120 | 18.54 | 8.9 sec | 16X13 | 0.4 |
| iCAT Next Generation | Medium | 120 | 20.27 | 14.7 sec | 16X13 | 0.25 |
| iCAT Next Generation | Large | 120 | 18.54 | 8.9 sec | 23X17 | 0.3 |
| iCAT Next Generation | Large | 120 | 37.07 | 17.8 sec | 23X17 | 0.3 |
| Galileos Comfort | Medium | 85 | 21 | 14 sec | 15x15 | 0.3 |
| Galileos Comfort | Medium | 85 | 42 | 14 sec | 15x15 | 0.3 |
| ProMax panoramic | Standard panoramic 25X30 | 66 | 144 | 16 sec | | |
| ProMax cephalometric | Full cephalogram 30X27 | 68 | 112.2 | 18.7 sec |
Figure 1RS-110 anthropometric head phantom used in this study with the levels where nanoDOTS were located. A- Eye lens B- Parotid gland C- Submandibular gland D- Thyroid gland.
Figure 2nanoDOT OSL dosimeters (Landauer Corp., Glenwood, IL).
Figure 3Example of the axial, coronal and sagittal images obtained with KODAK 9500 scanner using the head anthropomorphic phantom: a- Medium FOV b- Large FOV.
Peak skin and absorbed tissue doses for the investigated scan protocols
| Kodak 9500 | 86 | 108 | 15×9 | 3.58 | 0.42 | 2.88 | 3.04 | 0.32 |
| Kodak 9500 | 120 | 108 | 20×18 | 4.15 | 3.55 | 3.57 | 3.17 | 0.45 |
| iCAT Next Generation | 120 | 18.54 | 16×13 | 1.44 | 0.92 | 1.15 | 1.21 | 0.24 |
| iCAT Next Generation | 120 | 20.27 | 16×13 | 2.51 | 1.60 | 2.05 | 1.79 | 0.25 |
| iCAT Next Generation | 120 | 18.54 | 23×17 | 2.08 | 1.41 | 1.93 | 1.79 | 0.25 |
| iCAT Next Generation | 120 | 37.07 | 23×17 | 2.60 | 1.75 | 2.11 | 2.23 | 0.39 |
| Galileos Comfort | 85 | 21 | 15×15 | 2.49 | 0.58 | 2.45 | 1.93 | 0.36 |
| Galileos Comfort | 85 | 42 | 15×15 | 4.62 | 0.94 | 4.46 | 3.61 | 0.46 |
| ProMax panoramic | 66 | 144 | | 0.26 | 0.05 | 0.24 | 0.14 | 0.04 |
| ProMax cephalometric | 68 | 112.2 | 0.09 | 0.07 | 0.06 | 0.06 | 0.05 |